The operating temperature range depends on the cooling method, so the choice of water cooling or refrigerator must match the required process temperature.
Determines whether the bath can reach the required low temperatures and which cooling option must be specified.
Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply with a minimum 10 A fuse, which must be provided at the installation site.
Needed to size the branch circuit and protective device correctly before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
External branch circuit protection according to UL 489 is required, so a suitable UL-listed breaker must be included in the installation.
Affects the electrical installation design and compliance with UL requirements.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are interdependent, so a −10% voltage combined with +3% frequency is not permitted.
Prevents installation on a supply where combined voltage and frequency deviations exceed the allowed envelope.
Source
“Example: -10% voltage and +3% frequency -> not allowed !”
Installation
Single- and two-phase devices are delivered with a country-specific plug that must be specified at ordering.
Ensures the correct mains plug is supplied and avoids rework at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The ambient temperature must stay between 5 °C and 40 °C for correct operation.
Defines the permissible room conditions where the circulator may be installed and operated.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation
The bath must be filled with at least 2.5 l, which sets the minimum heat transfer fluid charge.
Prevents operation with insufficient fluid and ensures proper pump and heating behaviour.
Source
“min. filling capacity 2,5 l”
Operation
Polyglycol must not be used as heat transfer fluid when the Huber pump adapter is fitted.
Avoids selecting an incompatible fluid that could damage the pump adapter or system.
Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Safety
The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.
Provides the required safety cut-out and must be configured correctly during commissioning.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety
The safety classification is III / FL according to DIN 12876, which defines the protection concept for flammable liquids.
Determines whether the bath is suitable for the intended safety category and fluid.
Source
“Safety classification III / FL”
Performance
The pressure pump delivers up to 27 l/min at 0.7 bar, which sets the maximum external circuit flow capability.
Used to check whether the pump can supply the required flow and pressure in the external loop.