Product name and general description of the CC-K25 unit
Specpp. 1-2
Technical data and performance parameters per DIN 12876
p. 2table, 12 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
Net weight
41 kg
sound pressure level +/- 4 dB(A)
50 dB(A)
Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current refrigerated bath
2 A
max. current immersion thermostat
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S395642
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-30...200 °C
Temperature stability at 70°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at 20°C
0,4 kW
Cooling power at 0°C
0,35 kW
Cooling power at -10°C
0,27 kW
Cooling power at -20°C
0,16 kW
Cooling power at -30°C
0,065 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,072 kg
CO2 equivalent
0,0 t
UN-number
UN 3358
Pressure / Suction pump max. delivery
27 l/min
max. delivery pressure
0,7 bar
max. delivery (suction)
22 l/min
max. delivery pressure (suction)
0,4 bar
Pump connenction (optional)
M16x1 male
Bath volume
23,5 l
Width bath opening WxD
290x329 mm
Bath depth
200 mm
Height of bath opening
450 mm
Overall dimensions WxDxH **
350x555x615 mm
** см. сноску в оригинале (на странице не приведена)
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is classified as safety class III / FL per DIN 12876, which determines the required overtemperature protection and suitability for flammable liquids.
Safety class and FL designation dictate whether the bath can be used with flammable heat transfer fluids and what protective measures are mandatory.
Source
“Safety classification
III / FL”
Selection
The refrigeration circuit uses R-290 (propane, A3, H220) with a charge of only 0.072 kg, so the installation location must satisfy flammable-refrigerant requirements.
A3 refrigerant affects room ventilation, ignition-source restrictions and service access during installation and maintenance.
Source
“Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)”
Installation
The power supply must be 220-240 V single- or two-phase at 50/60 Hz, and the branch circuit must be protected with a fuse between 10 A and 16 A.
Defines the electrical supply and upstream protection needed before the unit can be connected.
Source
“Power supply requirement
220-240V 1~/2~ 50/60Hz”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is explicitly not allowed.
Prevents commissioning on a supply where the combined deviation would exceed the motor's permissible operating envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The enclosure is rated IP20, so the unit must be installed in a dry location protected from water ingress.
IP20 rules out washdown or wet environments and constrains where the bath can be sited.
Source
“Degree of Protection
IP20”
Installation
Ambient temperature at the installation site must stay between 5 °C and 40 °C.
Outside this range cooling capacity drops and reliable operation is not guaranteed.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
Cooling capacity falls from 0.4 kW at 20 °C to 0.065 kW at -30 °C, so the achievable heat load depends strongly on the working temperature.
Sizing the bath for a process requires checking the cooling capacity at the actual setpoint, not at 20 °C.
Source
“at 20°C
0,4 kW
at 0°C
0,35 kW
at -10°C
0,27 kW
at -20°C
0,16 kW
at -30°C
0,065 kW”
Performance
Cooling capacity is specified at 20 °C room temperature and may drop if ambient temperature rises.
Prevents overestimating available cooling in a warm laboratory and guides ventilation planning.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation
The pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction.
Determines whether the pump can supply an external circuit or cascade at the required flow and head.