CS 125 Immersion Circulator Datasheet

Contents

  1. Overview p. 1
    Product introduction and main features of CS 125
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 32 rows

CS 125 by Van der Heijden — Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...15 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C0,5 K
Interface digitalRS232
Safety classificationI / NFL
Cooling power at ambient temperature 20°C
at 15°C17 kW
at 10°C12,5 kW
at 0°C12 kW
at -10°C9 kW
at -20°C6 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity6 kg
CO2 equivalent8,38 t
Circulation pump:
max. delivery70 l/min
max. delivery pressure3,7 bar
Pump connectionG3/4 male
min. filling capacity225 l
max. filling capacity245 l
Overall dimensions WxDxH **1400x1000x1800 mm
Net weight394 kg
sound pressure level +/- 4 dB(A)75 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)17 A
Fuse (3 phase)3x20A
Degree of ProtectionIP52
min. ambient temperature5 °C
max. ambient temperature32 °C
from Serial-No.:1.0/24

** Overall dimensions WxDxH; Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller's cooling capacity drops as ambient temperature rises, so selection must be based on the actual site ambient temperature.

Cooling capacity is rated at 20°C ambient; higher ambient reduces available capacity, affecting sizing.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so the installer must provide a suitable plug.

Affects electrical installation planning and procurement of the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are interdependent: a -5% voltage deviation with +2% frequency is not allowed.

Site power quality must satisfy the combined tolerance rule or the unit may not operate reliably.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit requires a 3-phase 400 V 50 Hz supply protected by 3x20 A fuses and draws up to 17 A.

Determines electrical supply, protection device rating and cable sizing for installation.

Source
“Fuse (3 phase) 3x20A”
Installation

The chiller must be installed where ambient temperature stays between 5°C and 32°C.

Operating outside this ambient range risks performance loss or equipment damage.

Source
“min. ambient temperature 5 °C max. ambient temperature 32 °C”
Installation

Space requirements around the unit must be respected, so the installation layout must leave adequate clearance.

Air-cooled condenser needs clearance for airflow; ignoring it degrades cooling and may cause overheating.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Safety

The refrigerant R-449A is classified A1 but carries GHS hazard H280 (pressurized gas), so handling and labeling must account for pressure hazard.

Safety classification and hazard statement inform risk assessment and refrigerant handling procedures.

Source
“R-449A (A1, H280)”
Safety

The unit has a CO2 equivalent of 8.38 t from 6 kg of R-449A, which matters for F-gas reporting and leak-check obligations.

Determines regulatory leak-check frequency and environmental reporting requirements.

Source
“CO2 equivalent 8,38 t”
Operation

The flow rate is adjusted via a manual bypass valve on the back of the chiller, so access to the rear is needed for balancing.

Installation must leave rear access for flow adjustment during commissioning and operation.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation

The recommended thermofluid is a 50:50 water-monoethylene glycol mixture, which sets the fluid to use for the -20…15°C range.

Using the correct fluid ensures freeze protection and proper heat transfer within the operating range.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Performance

The circulation pump delivers up to 70 l/min at 3.7 bar, which constrains the external circuit pressure drop that can be served.

External loop resistance must be checked against pump capability to ensure required flow.

Source
“max. delivery pressure 3,7 bar”
Compliance

The equipment is EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments, not residential.

Class A equipment may cause interference in residential areas; installation environment must be industrial.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Compliance

The unit is compliant with US-SNAP for industrial process refrigeration end-use, which supports its use in that application.

Confirms refrigerant acceptability for the intended industrial process refrigeration application.

Source
“The US-SNAP end-use for this equipment is the industrial process refrigeration.”