Product introduction and general description of the CS 125-H unit
Specp. 2
Technical specifications according to DIN 12876 standard
p. 1table, 33 rows
Technical data according to DIN 12876 — CS 125-H by Van der Heijden (VDH4100343)
Parameter
Value
Order-No.
CS 125-H by Van der Heijden
Operating temperature range
-20...85 °C
temperature set point / display
colour LED Touchscreen
Internal temperature sensor
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
RS232
Safety classification
I / NFL
Heating power
6 kW
Cooling power at ambient temperature 20°C
.
at 15°C
17 kW
at 10°C
12,5 kW
at 0°C
12 kW
at -10°C
9 kW
at -20°C
6 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Refrigerant quantity
6 kg
CO2 equivalent
8,38 t
Circulation pump: max. delivery
70 l/min
max. delivery pressure
3,7 bar
Pump connection
G3/4 male
min. filling capacity
225 l
max. filling capacity
245 l
Overall dimensions WxDxH **
1400x1000x1800 mm
Net weight
400 kg
sound pressure level +/- 4 dB(A)
75 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
30 A
Fuse (3 phase)
3x32A
Degree of Protection
IP52
min. ambient temperature
5 °C
max. ambient temperature
32 °C
from Serial-No.:
1.0/24
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller's cooling capacity drops as ambient temperature rises, so it must be sized against the actual room temperature at the installation site.
Cooling capacity is rated at 20°C ambient; higher site temperatures reduce available capacity and can undersize the unit.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit requires a 400 V three-phase supply with a 3x32 A fuse and draws up to 30 A, so the electrical supply and protection must be sized accordingly.
Determines the feeder, breaker/fuse rating and three-phase supply required at the installation point.
Source
“Fuse (3 phase)
3x32A”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.
Affects electrical installation planning and what the customer must supply on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances must be respected simultaneously, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Prevents connecting the chiller to a supply whose combined voltage/frequency deviation exceeds the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The chiller must be operated within an ambient temperature window of 5°C to 32°C.
Defines the permissible room conditions for reliable operation and must be checked at the installation location.
Source
“min. ambient temperature
5 °C
max. ambient temperature
32 °C”
Installation
The unit is air-cooled and requires space for the condenser airflow, so the space requirements around the chiller must be respected.
Insufficient clearance around an air-cooled condenser will degrade cooling performance and can cause overheating.
Source
“Please respect space requirements.”
Operation
The flow rate is adjusted via a manual bypass valve located on the back of the chiller.
Operators need access to the rear of the unit to set the flow, which affects installation layout and commissioning.
Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation
The recommended heat transfer fluid is a 50:50 mixture of water and monoethylene glycol.
Determines the fluid to be prepared and filled for correct operation and freeze protection.
Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Operation
The tank filling capacity must be between 225 l and 245 l.
Defines the required fluid volume for commissioning and the limits for safe operation.
Source
“min. filling capacity
225 l
max. filling capacity
245 l”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.
Informs safety assessment, labeling and handling requirements for the refrigerant charge.
Source
“R-449A (A1, H280)”
Safety
The unit has safety classification I / NFL according to DIN 12876.
Defines the safety category relevant to the application and the permissible operating conditions.
Source
“Safety classification
I / NFL”
Compliance
The equipment emits as Class A, Group 1 under EN55011, which has implications for electromagnetic compatibility in the installation environment.
Class A equipment may require EMC measures and is relevant for compliance in the installation environment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Compliance
The equipment is compliant with US-SNAP and applicable EU laws, with US-SNAP end-use being industrial process refrigeration.
Confirms regulatory compliance for the intended industrial process refrigeration application.
Source
“This equipment is compliant to US-SNAP and all applicable EU laws.”
Performance
The circulation pump delivers up to 70 l/min at a maximum delivery pressure of 3.7 bar.
Used to check whether the pump can meet the flow and pressure requirements of the connected circuit.