The chiller is rated for an operating temperature range of -20 to 85 °C, so the process must stay within these limits.
Defines the process temperature window the unit can support.
Source
“Operating temperature range -20...85 °C”
| Parameter | Value |
|---|---|
| Order-No. | VDH4100300 |
| Operating temperature range | -20...85 °C |
| temperature set point / display | colour LED Touchscreen |
| Internal temperature sensor | Pt100 |
| Temperature stability at -10°C | 0,5 K |
| Interface digital | RS232 |
| Safety classification | I / NFL |
| Heating power | 3 kW |
| Cooling power at ambient temperature 20°C | |
| at 15°C | 8 kW |
| at 10°C | 5 kW |
| at 0°C | 4,2 kW |
| at -10°C | 3,5 kW |
| at -20°C | 2 kW |
| Refrigeration machine | air-cooled, CFC- and HCFC-free |
| Refrigerant (ASHRAE, GHS) | R-449A (A1, H280) |
| Refrigerant quantity | 4 kg |
| CO2 equivalent | 5,59 t |
| Circulation pump: max. delivery | 70 l/min |
| max. delivery pressure | 3,7 bar |
| Pump connection | G3/4 male |
| min. filling capacity | 100 l |
| max. filling capacity | 110 l |
| Overall dimensions WxDxH ** | 680x730x1520 mm |
| Net weight | 210 kg |
| sound pressure level +/- 4 dB(A) | 63 dB(A) |
| Power supply (3 Phase) | 400V 3~ 50Hz |
| max. current (3 Phase) | 22 A |
| Fuse (3 phase) | 3x32A |
| Degree of Protection | IP52 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 32 °C |
| from Serial-No.: | 1.0/24 |
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
The chiller is rated for an operating temperature range of -20 to 85 °C, so the process must stay within these limits.
Defines the process temperature window the unit can support.
“Operating temperature range -20...85 °C”
Cooling capacity is specified at 20 °C ambient and drops if ambient temperature rises, so capacity must be derated for warmer rooms.
Prevents undersizing when the installation room is warmer than 20 °C.
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
The unit requires a 400 V three-phase 50 Hz supply with 22 A max current and 3x32 A fusing.
Determines electrical supply, cable and protection sizing at the installation site.
“Power supply (3 Phase) 400V 3~ 50Hz max. current (3 Phase) 22 A Fuse (3 phase) 3x32A”
Three-phase devices drawing less than 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation scope and ordering of the mating plug.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Voltage and frequency tolerances are ±5% and ±2% but must not combine in the unfavorable direction, so supply quality must be checked.
Ensures the site power quality is compatible and avoids invalid tolerance combinations.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
The unit is IP52 and must be installed where ambient temperature stays between 5 °C and 32 °C.
Sets environmental limits and protection level for the installation location.
“Degree of Protection IP52 min. ambient temperature 5 °C max. ambient temperature 32 °C”
The refrigerant is R-449A with 4 kg charge and 5.59 t CO2 equivalent, which matters for refrigerant handling and leak reporting.
Needed for refrigerant safety classification and CO2-equivalent reporting obligations.
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Refrigerant quantity 4 kg CO2 equivalent 5,59 t”
The pump connection is G3/4 male and the tank holds 100–110 l, so piping and fluid volume must be planned accordingly.
Determines connection fittings and required heat transfer fluid volume.
“Pump connection G3/4 male min. filling capacity 100 l max. filling capacity 110 l”
Flow rate is adjusted via a manual bypass valve on the back of the chiller, so rear access is needed for balancing.
Requires rear clearance for commissioning and flow adjustment.
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
The recommended thermofluid is a 50:50 water–monoethylene glycol mixture.
Guides correct fluid selection for the operating temperature range.
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
The equipment is Class A, Group 1 under EN55011, so it may cause radio interference in residential environments.
Affects EMC suitability of the installation location and possible mitigation.
“Classification (disturbance) to EN55011: Class A, Group 1”
The unit complies with US-SNAP for industrial process refrigeration and with applicable EU laws.
Confirms regulatory suitability for industrial process refrigeration applications.
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”
Space requirements must be respected per the operating conditions referenced online.
Ensures adequate clearance for air-cooled condenser airflow and service access.
“** Please respect space requirements. See operating conditions at www.huber-online.com”