Grande Fleur-eo

Contents

  1. Overview p. 1
    Product introduction and general description of the unit
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 16 rows
  4. p. 1 table, 28 rows

Technical data according to DIN 12876

ParameterValue
Global Warming Potential (GWP)0,02
Refrigerant quantity0,08 kg
CO2-Äquvivalent0,0 t
UN-numberUN 3358
Circulation pump:
max. delivery47 l/min
max. delivery pressure0,9 bar
Pump connectionM24x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity1,5 l
Overall dimensions WxDxH **295x530x570 mm
Power supply requirement208-240V 1~/2~ 50/60Hz
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/25

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicGrande Fleur-eo
Order-No.1041.0018.01
Operating temperature range-40...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power at 240V2 kW
Heating power at 230V1,8 kW
Heating power at 220V1,6 kW
Heating power at 208V1,5 kW
Heating power convertible1,5 - 2 kW
Cooling power with Thermooil at 200°C0,6 kW
Cooling power with Thermooil at 150°C0,6 kW
Cooling power with Thermooil at 100°C0,6 kW
Cooling power with Thermooil at 50°C0,6 kW
Cooling power with Thermooil at 20°C0,6 kW
Cooling power with Ethanol at 0°C0,6 kW
Cooling power with Ethanol at -20°C0,35 kW
Cooling power with Ethanol at -30°C0,2 kW
Cooling power with Ethanol at -40°C0,04 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit is designed only for external open operation and has no expansion tank, so the application loop must accommodate fluid filling and thermal expansion.

Determines whether an external reservoir or expansion vessel must be added to the installation.

Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation

Fluid filling and compensation for temperature-dependent volume changes must be provided through the externally connected application.

Requires the customer loop to be sized and arranged to accept fill and expansion volume.

Source
“The filling of the thermofluid, and allowance for the temperature dependent volume changes is only possible via an externally connected application.”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects site classification and handling.

Flammable refrigerant classification drives ventilation, ignition-source and safety requirements at the installation site.

Source
“R-290 (A3, H220)”
Installation

The unit is water-cooled and requires cooling water intake between 5 °C and 20 °C.

Cooling water supply temperature must be verified before installation; higher temperatures reduce cooling capacity.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.

Prevents condensation and corrosion on internal cooling water piping.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable cooling water quality is required.

Guides cooling water quality selection to avoid corrosion or fouling of the circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Power supply quality must be checked against these limits before commissioning.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Compliance

The example states that -5% voltage with +2% frequency is not allowed, while -5% voltage with -2% frequency is allowed.

Clarifies that voltage and frequency deviations must be evaluated together, not independently.

Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance

The equipment is EMC classified as Class A, Group 1 according to EN55011.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

The correct plug type must be ordered to match the site power outlet.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Three-phase devices below 63 A are supplied with cable but without plug.

Site electrical termination must be planned separately for three-phase units.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Pump connections use bore shape Y (60°) per DIN 3863, with ball socket per DIN 3863 and sleeve nut per DIN 3870.

Ensures compatible hoses and fittings are selected for the pump connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s.

Limits the choice of heat transfer fluid, especially at low temperatures.

Source
“max. permissible kin. viscosity 50 mm²/s”
Operation

The minimum filling capacity is 1.5 l.

The external application and piping must provide at least this fluid volume for safe operation.

Source
“min. filling capacity 1,5 l”