Grande Fleur w-eo

Contents

  1. Overview p. 1
    Product name and cover page of the datasheet
  2. Spec p. 2
    Technical specifications according to DIN 12876 standard
  3. p. 2 table, 11 rows

Technical data according to DIN 12876

ParameterValue
Cooling water connectionG1/2 male
min. cooling water differential pressure2 bar
max. cooling water pressure6 bar
min. filling capacity1,5 l
Overall dimensions WxDxH **295x530x570 mm
Net weight55 kg
Power supply requirement110-120V 1~ 60Hz
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit is designed only for external open operation and has no expansion tank, so the thermal fluid and its temperature-dependent volume changes must be handled by an externally connected application.

Determines whether an external expansion vessel and open bath/loop must be provided during installation.

Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation

Cooling water must be supplied at 2–6 bar and between 5 °C and 20 °C, and pipes must be insulated if the water temperature can fall below the dew point.

Sets the cooling-water supply requirements and condensation/insulation measures for installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity is rated at 20 °C room temperature and 2 bar cooling-water differential pressure, so lower water pressure or higher water temperature will reduce achievable cooling.

Needed to verify that site cooling-water conditions match the published performance data.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet °C and 2 bar differential pressure between cooling water inlet and -”
Installation

The cooling-water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides cooling-water quality and material compatibility checks before connecting the unit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.

Must be checked against the site power supply to avoid invalid supply combinations.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Compliance

The equipment is EMC classified as Class A, Group 1 to EN55011, which affects installation in residential or sensitive environments.

Class A equipment may require additional EMC precautions depending on the installation location.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

The correct plug type must be stated on the order to match the site power outlet.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Pump connections use bore shape Y (60°) per DIN 3863 with ball socket and sleeve nut per DIN 3870, so matching hoses and fittings are required.

Ensures correct mating fittings and hoses are procured for the M24x1.5 pump connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Safety

The refrigerant is R-290 (A3, H220) with GWP 0.02, so flammable-refrigerant safety measures apply.

A3 flammability classification affects room ventilation, ignition-source control and service access planning.

Source
“R-290 (A3, H220)”
Operation

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

The overtemperature cut-out must be set correctly for the thermal fluid and application to protect the process.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The maximum permissible kinematic viscosity of the thermal fluid is 50 mm²/s, which limits fluid selection at low temperatures.

The chosen heat-transfer fluid must stay below this viscosity limit over the operating temperature range.

Source
“max. permissible kin. viscosity”
Selection

The minimum filling capacity is 1.5 l, so the external application volume must be sized accordingly.

The external loop must provide at least this volume for safe circulation.

Source
“min. filling capacity”
Installation

The unit requires 5–40 °C ambient temperature and IP20 protection, so it must be installed in a dry, protected location within that range.

Site ambient conditions and enclosure protection must match the installation location.

Source
“min. ambient temperature”