Induztr
Engineering Digest Document Viewer Related Models

HB120 Heating Bath Datasheet

Contents

  1. Overview p. 1
    Product introduction and general description of HB120 heating bath
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 27 rows

Technical data according to DIN 12876

ParameterValue
Order-No.HB120
Operating temperature range60...250 °C
Resolution of display0,01 K
Temperature stability at 70°C0,1 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationII / FL
Heating power at 460V12 kW
Heating power at 400V12 kW
Heating power at 380V10,5 kW
Circulation pump:MK pump
max. delivery100 l/min
max. delivery pressure3,5 bar
Pump connectionM30x1,5 male
Overall dimensions WxDxH479x539x695 mm
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
Net weight88 kg
Power supply (3 Phase)380-460V 3~ 50/60Hz
max. current (3 Phase)20 A
Fuse (3 phase)3x25A
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S575934

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is designed exclusively for external open operation and has no expansion tank, so the application must provide filling and volume compensation.

Determines whether the unit can be used with a closed loop or requires an external open bath/reservoir with volume compensation.

Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation

Three-phase devices below 63 A are supplied with a cable but without a plug, so the installer must provide a suitable plug and connection.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The supply voltage tolerance is ±10% only if the frequency tolerance does not act in the opposite direction, so supply quality must be checked before installation.

Prevents selecting or installing the unit on a supply where combined voltage and frequency deviations exceed the permitted envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety

Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.

Affects electrical safety design, earth connection and residual-current device selection.

Source
“Attention: leakage current > 3,5mA”
Installation

The pump connection is an M30x1.5 male thread, so mating hoses or adapters must match this thread.

Ensures correct selection of hoses, unions and adapters for the external circuit.

Source
“Pump connection M30x1,5 male”
Operation

The minimum achievable temperature of 60 °C depends on ambient temperature being below 60 °C and on the heat loss of the application.

Prevents overestimating cooling capability when the unit has no active refrigeration.

Source
“The ability to reach the min. temperature 60°C, depends on how far the ambient temperature is below 60°C, and the heat loss of the application.”
Selection

The application heat loss must be between 0.1 kW and 2.8 kW depending on the model to reach the 60 °C minimum temperature.

Allows checking whether the connected application load is compatible with reaching the minimum temperature.

Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the HB model, to reach the minimum temperature of 60°C.”
Installation

The unit requires a three-phase 380–460 V 50/60 Hz supply with a maximum current of 20 A and 3x25 A fusing.

Defines the electrical supply, cable and protection requirements for installation.

Source
“Power supply (3 Phase) 380-460V 3~ 50/60Hz”
Operation

The ambient temperature must remain between 5 °C and 40 °C for correct operation.

Sets the allowable site conditions and affects placement and room ventilation.

Source
“min. ambient temperature 5 °C”
Compliance

The equipment is classified as Class A, Group 1 for electromagnetic disturbance, which has implications for use in residential environments.

Informs EMC compliance planning and possible restrictions in residential or light-industrial locations.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Requires commissioning personnel to configure the safety limit correctly for the thermal fluid and application.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance

The circulation pump delivers up to 100 l/min and 3.5 bar, which must be sufficient for the external circuit pressure drop.

Used to verify hydraulic compatibility of the external application with the pump capability.

Source
“max. delivery 100 l/min”