The unit is designed exclusively for external open operation and has no expansion tank, so the application must provide fluid filling and volume compensation.
Determines whether the unit can be used with a closed loop or requires an external open bath/reservoir with volume compensation.
Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation
Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so the customer must provide the plug and connection.
Affects electrical installation planning and the bill of materials for the power connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are only valid if they do not run in opposite directions, so a -10% voltage combined with +3% frequency is not permitted.
Guides site power quality assessment before connecting the 400 V three-phase supply.
Source
“Example: -10% voltage and +3% frequency -> not allowed !”
Compliance
The equipment is EMC class A group 1, which typically requires industrial environment installation and may need filtering in residential areas.
Affects EMC compliance planning and whether the installation location is suitable.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
Leakage current from external temperature control must be considered, so protective earthing and RCD selection need attention.
Important for electrical safety design and selection of residual current devices.
Source
“Attention: leakage current externe Temperierung”
Operation
The HotBox has no active refrigeration, so reaching 60 °C minimum depends on ambient temperature being below 60 °C and on application heat loss.
Critical for selecting the unit when sub-ambient or 60 °C control is required.
Source
“These devices do not have an active refrigeration system.”
Performance
To reach the 60 °C minimum temperature, the application heat loss must be between 0.1 kW and 2.8 kW depending on the HB model.
Defines the allowable thermal load on the application for the unit to achieve its minimum temperature.
Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the HB model, to reach the minimum temperature of 60°C.”
Installation
Space requirements around the unit must be respected according to the operating conditions published online.
Needed for layout planning and ensuring adequate ventilation and access.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Selection
The pump connection is M38x1.5 male, so mating hoses and fittings must match this thread.
Determines the connection hardware required for the external circuit.
Source
“Pump connection M38x1,5 male”
Operation
The unit has a 48 kW heating power, which defines the maximum heating capacity available for the application.
Used to size the application and estimate heat-up times.
Source
“Heating power 48 kW”
Operation
The circulation pump delivers up to 200 l/min at up to 5.5 bar, which sets the hydraulic capability of the loop.
Needed to check whether the external circuit pressure drop is compatible with the pump.
Source
“max. delivery 200 l/min”
Installation
The unit requires a 400 V 3~ 50 Hz three-phase supply.
Determines the electrical supply infrastructure required at the installation site.