Product introduction and main features of KISS 112A
Specp. 2
Technical data according to DIN 12876
p. 1table, 37 rows
Technical data according to DIN 12876
Parameter
KISS 112A
Order-No.
2052.0003.98
Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connenction (optional)
M16x1 male
Bath volume
12 l
min. filling capacity
7,3 l
Height of bath opening
165 mm
Width bath opening WxD/ bath depth
275x161/ 150 mm
Overall dimensions WxDxH **
333x360x335 mm
Net weight
8 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
401722
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The usable temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.
Determines whether the bath can reach the required low temperature with the available cooling utility.
Source
“Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so the branch circuit must be sized accordingly.
Needed to confirm site power availability and breaker sizing before installation.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A”
Installation
An external branch circuit protection device according to UL 489 is mandatory.
Affects the electrical panel design and compliance of the installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10% only if they do not oppose each other, e.g. -10% voltage with +3% frequency is not allowed.
Prevents selecting a site with unstable supply that would violate the allowed operating envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct plug type is supplied for the destination country.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant plastic option ordered.
Prevents premature pump failure when choosing the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than using pure water.
Guides fluid selection to avoid freezing damage in low-temperature operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety
The unit is safety class III / FL with adjustable overtemperature protection according to DIN 12876.
Confirms the safety concept and the need to set the overtemperature limit during commissioning.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety
The autostart function restarts the unit automatically after a power failure, which must be considered in the risk assessment.
Unattended restart after mains interruption can be hazardous and must be evaluated for the process.
Source
“Autostart function for power failure”
Installation
The enclosure is rated IP20, so it must be installed in a dry location protected from water ingress.
Restricts the installation environment and prevents placing the unit where splashing or condensation occurs.
Source
“Degree of Protection
IP20”
Operation
Ambient temperature must stay between 5 °C and 40 °C for specified performance.
Defines the allowable room conditions at the installation site.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
The pressure/suction pump delivers up to 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction.
Used to check whether the external circuit hydraulic resistance is compatible with the pump.