The unit is rated for a working temperature range of -20 to 200 °C, so it can cover both sub-ambient and high-temperature bath duties.
Defines whether the bath can handle the required process temperatures before selection.
Source
“Operating temperature range -20...200 °C”
Performance
Cooling capacity falls from 0.25 kW at 20 °C to only 0.05 kW at -20 °C, so low-temperature heat loads must be sized against the reduced capacity.
Prevents oversizing the cooling load at the cold end of the operating range.
Source
“at -20°C 0,05 kW”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source restrictions.
Flammable refrigerant classification drives installation location and safety assessment.
Source
“R-290 (A3, H220)”
Installation
The power supply requirement is 220-240 V single or two-phase at 50/60 Hz, so the site must provide the correct mains configuration.
Mismatch with site power will prevent commissioning or require a transformer.
Source
“220-240V 1~/2~ 50/60Hz”
Installation
The minimum fuse rating is 10 A and the maximum current of the refrigerated bath is 1.5 A, so the branch circuit must be protected at no less than 10 A.
Ensures the electrical supply and protection device are correctly sized.
Source
“min. Fuse 10A”
Installation
The unit is rated IP20, so it must be installed in a dry indoor location protected from water ingress.
IP20 limits where the bath can be placed and rules out washdown or wet environments.
Source
“Degree of Protection IP20”
Operation
The allowable ambient temperature range is 5 to 40 °C, and cooling capacity may drop if ambient temperature rises.
High ambient temperature reduces cooling performance and must be considered in room layout.
Source
“max. ambient temperature 40 °C”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant plastic pump specified.
Prevents chemical attack and premature pump failure from incompatible bath fluids.
Source
“The plastic pump is not resistant against acetone and polyglycols”
Operation
For freeze protection the manufacturer recommends mixing water with glysantine or ethylene glycol rather than using pure water.
Avoids freezing and damage when operating near or below 0 °C.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.
Defines the acceptable mains quality for reliable operation and warranty compliance.
Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, so it is intended for industrial rather than residential environments.
Class A equipment may cause interference in residential areas and affects EMC planning.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
The correct plug type must be stated on the order to match the site socket.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Maintenance
The optional Pt100 sensor connection is only available factory fitted, so it must be decided at the time of purchase.
Retrofitting is not possible, so the option must be included in the initial order.
Source
“only available factory fitted, additional charge”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety class III/FL, providing fail-safe protection.
Confirms the safety concept required for unattended or high-temperature operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”