Minichiller 1200w-H OLÉ

Contents

  1. Overview p. 1
    Product name and cover page for the Minichiller unit
  2. Spec p. 2
    Technical data table according to DIN 12876
  3. p. 2 table, 16 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -20°C24 l/h
Cooling water connectionG1/2 male
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
min. filling capacity2,8 l
expansion tank2,2 l
Overall dimensions WxDxH **280x490x424 mm
Net weight37 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current12 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S562918

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires cooling water with a minimum differential pressure of 3 bar and a maximum pressure of 6 bar, so the facility water supply must be sized accordingly.

Insufficient differential pressure will prevent the water-cooled refrigeration unit from operating correctly.

Source
“min. cooling water differential pressure 3 bar”
Installation

The unit is designed for cooling water intake between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the condensation point.

Out-of-range cooling water temperature affects cooling capacity and can cause condensation and corrosion.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase, so the water supply must be sized with margin.

Sizing the cooling water circuit only for nominal conditions can lead to insufficient cooling in practice.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects installation location and ignition-source restrictions.

Flammable refrigerant classification requires appropriate safety measures and room ventilation considerations.

Source
“R-290 (A3, H220)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must not be used as heat transfer fluids.

Using incompatible media will damage the pump and void reliable operation.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Proper antifreeze selection prevents freezing and damage in the cooling circuit at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.

The electrical supply must be verified against this combined tolerance to avoid invalid operating conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Failing to specify the plug type at order time can delay commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance

The equipment emits Class A, Group 1 disturbances per EN55011, so it is intended for industrial electromagnetic environments.

Class A equipment may not be suitable for residential or light-industrial EMC environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The unit has IP20 protection, so it must be installed in a dry location protected from water ingress.

IP20 does not protect against water, limiting where the chiller can be safely installed.

Source
“Degree of Protection IP20”
Operation

The ambient temperature must remain between 5°C and 40°C for specified operation.

Exceeding the ambient limit reduces cooling performance and may trip protection.

Source
“max. ambient temperature 40 °C”
Selection

The optional Pt100 sensor connection is for display only, not for control, and is available only factory-fitted at additional cost.

Engineers expecting external process temperature control must select a different configuration.

Source
“Option: Pt100 sensor connection #10519 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge)”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or degrade wetted materials in the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”