Minichiller 800-H OLÉ

Contents

  1. Spec p. 2
    Technical specifications according to DIN 12876 for the Minichiller 800-H OLÉ
  2. p. 2 table, 9 rows

Technical data according to DIN 12876

ParameterValue
sound pressure level +/- 4 dB(A)56 dB(A)
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current12 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S622649

1.2/26

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is rated for a -20 to 100 °C operating range, so it must be matched to the process temperature window before selection.

Defines whether the chiller can cover the required process temperatures at all.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity drops from 0.8 kW at 15 °C to 0.3 kW at -20 °C, so the heat load must be checked at the lowest operating temperature.

Sizing on nominal capacity alone risks insufficient cooling at low temperatures.

Source
“Cooling power at 15°C 0,8 kW at 0°C 0,6 kW at -10°C 0,45 kW at -20°C 0,3 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source assessment.

Flammable refrigerant classification drives installation safety requirements.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation

The pump connections are M16x1 male, so mating hoses and fittings must be ordered to this thread.

Prevents mismatch of process hoses and adapters during installation.

Source
“Pump connection M16x1 male”
Operation

Whisper mode reduces sound pressure to 51 dB(A) but also cuts pump delivery to 14 l/min and 0.2 bar, so quiet operation trades off hydraulic performance.

Users must verify that reduced flow and pressure still meet the process loop demand.

Source
“Pump data for whisper mode: Delivery rate: 14 l/min Delivery pressure: 0.2 bar”
Installation

The unit requires a 208-240 V single/two-phase 50/60 Hz supply with max current 12 A and a 16 A fuse.

Electrical supply and protection must be provisioned before commissioning.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz max. current 12 A Fuse 16 A”
Installation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C room temperature.

Installation location and ventilation directly affect achievable cooling performance.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Compliance

Voltage and frequency tolerances are ±5% and ±2% but must not be combined in the unfavorable direction, so supply quality must be checked.

Out-of-tolerance supply can prevent warranty-compliant operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Selection

The plastic pump is not resistant to acetone and polyglycols, so these media require an alternative pump material.

Media compatibility must be confirmed before selecting the standard pump version.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Missing plug specification delays electrical installation on site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The OLÉ controller includes an autostart function after power failure, which may restart the process unattended.

Automatic restart behavior must be considered in the process risk assessment.

Source
“Autostart function for power failure”
Maintenance

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Prevents freezing and damage in the cooling circuit at sub-zero operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”