Product name and general description of the chiller unit
Specpp. 1-2
Technical data and performance parameters according to DIN 12876
p. 2table, 3 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S608927
Technical data according to DIN 12876 — Unichiller 015-H (Order-No.: 3051.0044.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at ambient temperature +20°C
at 15°C
1,5 kW
at 0°C
1 kW
at -10°C
0,7 kW
at -20°C
0,3 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,17 kg
CO2 equivalent
0,0 t
Circulation pump:
Immersion pump
max. delivery
29 l/min
max. delivery pressure
1 bar
Delivery at 0,5 bar
21 l/min
Pump connection
G3/4 male
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Overall dimensions WxDxH **
420x487x579 mm
Net weight
55 kg
sound pressure level +/- 4 dB(A)
63 dB(A)
Power supply requirement
220-240V 1~/2~ 50Hz
max. current
14,5 A
Fuse
16 A
Degree of Protection
IP20
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window.
Defines whether the unit can cover the required process temperatures before selection.
Source
“Operating temperature range
-20...100 °C”
Selection
Cooling capacity falls with ambient temperature, so the required heat load must be checked against the capacity curve at the actual site ambient.
Prevents undersizing when the chiller is installed in a warm room.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit requires a 220-240 V single/two-phase 50 Hz supply with a 16 A fuse and draws up to 14.5 A, so the electrical supply must be sized accordingly.
Determines the electrical circuit, protection and cabling needed at the installation site.
Source
“Power supply requirement
220-240V 1~/2~ 50Hz
max. current
14,5 A
Fuse
16 A”
Installation
Voltage and frequency tolerances are linked, and a -5% voltage combined with +2% frequency is explicitly not allowed.
Prevents installation on a supply where combined voltage/frequency deviation exceeds the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The chiller is air-cooled and requires respect of space requirements, so sufficient clearance for air flow must be provided.
Air-cooled units need adequate clearance to avoid reduced cooling performance or overheating.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects site safety classification and handling requirements.
Flammable refrigerant classification influences installation location, ventilation and safety planning.
Source
“Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used without a more resistant pump option.
Prevents chemical attack and pump failure when selecting the heat transfer fluid.
Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection the manufacturer recommends mixing water with glysantine or ethylene glycol rather than using pure water.
Guides correct fluid preparation to avoid freezing and damage at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct plug type is ordered for the destination country and avoids rework on site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance
The equipment is classified as EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments rather than residential ones.
Affects EMC planning and suitability for the installation environment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Performance
The circulation pump delivers a maximum of 29 l/min at 1 bar and 21 l/min at 0.5 bar, which sets the achievable flow in the external circuit.
Used to check whether the pump can overcome the pressure drop of the external loop.
Source
“max. delivery
29 l/min
max. delivery pressure
1 bar
Delivery at 0,5 bar
21 l/min”
Operation
The minimum filling capacity is 3.8 l with 1.7 l expansion volume, so the external loop and fluid charge must be planned within these limits.
Determines the required heat transfer fluid volume and expansion allowance.
Source
“min. filling capacity
3,8 l
Volume of expansion
1,7 l”
Installation
The pump connection is G3/4 male, so mating hoses and fittings must match this thread.
Ensures correct hose couplings are available for commissioning.
Source
“Pump connection
G3/4 male”
Operation
The unit is rated IP20, so it must be protected from water ingress and is not suitable for wet or washdown environments.
Constrains the installation location with respect to moisture and dust.