The chiller is intended only for externally closed applications, so the user circuit must be closed.
Prevents selecting this unit for open-loop or tank-venting duty where it would be unsuitable.
Source
“For externally closed applications.”
| Parameter | Value |
|---|---|
| Operating temperature range | -20...100 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Temperature stability at -10°C | 0,2 K |
| Safety classification | III / FL |
| Heating power | 1,6 kW |
| Cooling power at 15°C | 2 kW |
| Cooling power at 0°C | 2 kW |
| Cooling power at -10°C | 1,5 kW |
| Cooling power at -20°C | 0,8 kW |
| Refrigeration machine | air-cooled, CFC- and HCFC-free |
| Refrigerant (ASHRAE, GHS) | R-449A (A1, H280) |
| Global Warming Potential (GWP) | 1397 |
| Circulation pump: | B |
| max. delivery | 30 l/min |
| max. delivery pressure | 3,8 bar |
| Delivery at 0,12 bar | 29 l/min |
| Delivery at 0,5 bar | 25 l/min |
| Delivery at 1,0 bar | 21 l/min |
| Delivery at 1,5 bar | 17 l/min |
| Delivery at 2,0 bar | 13,5 l/min |
| Delivery at 2,5 bar | 10 l/min |
| Delivery at 3,0 bar | 6,5 l/min |
| Pump connection | G3/4 male |
| min. filling capacity | 2,5 l |
| Volume of expansion | 2 l |
| Overall dimensions WxDxH ** | 450x510x1230 mm |
| Net weight | 141 kg |
| Power supply requirement | 208V 2~ 60Hz |
| max. current | 14 A |
| Fuse | 2x16 A |
** Overall dimensions WxDxH — как напечатано в исходном тексте.
The chiller is intended only for externally closed applications, so the user circuit must be closed.
Prevents selecting this unit for open-loop or tank-venting duty where it would be unsuitable.
“For externally closed applications.”
The unit requires a 208 V two-phase 60 Hz supply with 14 A maximum current and 2x16 A fusing.
Electrical supply and protection must be provisioned before installation.
“Power supply requirement 208V 2~ 60Hz max. current 14 A Fuse 2x16 A”
Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.
Avoids assuming a ready-to-plug power connection and plans for local plug installation.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are explicitly not allowed.
Supply quality must be checked because some voltage/frequency combinations invalidate the tolerance.
“-5% voltage and +2% frequency -> not allowed!”
Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises.
Cooling performance must be derated for warmer installation rooms.
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
The unit is rated IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.
Defines environmental limits and protection against ingress for the installation site.
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
The refrigerant is R-449A with GHS classification H280 and GWP 1397, so refrigerant handling and leak response must follow safety rules.
Affects safety planning, refrigerant labeling, and environmental compliance.
“R-449A (A1, H280)”
The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.
Overtemperature protection is a required safety function for the external circuit.
“With adjustable overtemperature protection according to DIN 12876.”
The pump connection is G3/4 male, so mating hoses and couplings must match this thread.
Prevents mismatch of process hoses and fittings during installation.
“Pump connection G3/4 male”
The minimum filling capacity is 2.5 l and expansion volume is 2 l, which sets the minimum heat-transfer fluid charge.
Ensures enough fluid volume and expansion space for safe operation.
“min. filling capacity 2,5 l Volume of expansion 2 l”
Cooling power falls from 2 kW at 0 °C to 0.8 kW at -20 °C, so low-temperature duty has reduced capacity.
Selection must account for reduced cooling capacity at the lowest operating temperatures.
“at -20°C 0,8 kW”
Pump delivery drops with increasing pressure, from 29 l/min at 0.12 bar to 6.5 l/min at 3.0 bar.
External circuit pressure drop must be checked against the pump curve.
“Delivery at 3,0 bar 6,5 l/min”
The equipment is EMC Class A Group 1, which is industrial disturbance classification and may require EMC precautions.
Class A equipment may need additional EMC measures in residential or sensitive environments.
“Classification (disturbance) to EN55011: Class A, Group 1”
Space requirements around the unit must be respected according to the operating conditions.
Air-cooled units need clearance for airflow and service access.
“Please respect space requirements. See operating conditions at www.huber-online.com”