Unichiller 022-H OLÉ Technical Data Sheet

Contents

  1. Overview p. 1
    Product name and cover page of the datasheet
  2. Spec p. 2
    Technical data according to DIN 12876

Engineering Facts

12 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -10 to 100 °C, so the process must stay within these limits.

Defines the application window and excludes processes requiring lower or higher temperatures.

Source
“Operating temperature range -10...100 °C”
Performance

Cooling capacity drops as the setpoint falls, from 2.2 kW at 15 °C to 1 kW at -10 °C, so heat load must be checked at the actual working temperature.

Sizing on nominal capacity alone can lead to insufficient cooling at low temperatures.

Source
“Cooling power at 15°C 2,2 kW at 0°C 1,6 kW at -10°C 1 kW”
Installation

The unit requires a 220–240 V single-phase 50 Hz supply with up to 15 A current and a 16 A fuse.

Determines electrical infrastructure, cable sizing and breaker rating at the installation site.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 15 A Fuse 16 A”
Installation

The chiller is air-cooled and must be installed where ambient temperature stays between 5 °C and 40 °C.

Ambient conditions affect cooling performance and reliable operation of the air-cooled refrigeration machine.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity data are valid at 20 °C room temperature and may fall if ambient temperature rises.

Prevents overestimating available cooling capacity in warm plant rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, and the circuit contains 0.59 kg with 0.8 t CO2 equivalent.

Affects refrigerant handling, leak response and environmental reporting obligations.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Global Warming Potential (GWP) 1397 Refrigerant quantity 0,59 kg CO2 equivalent 0,8 t”
Safety

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump specified.

Prevents chemical attack and premature pump failure when selecting the heat transfer fluid.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

Avoids freezing and damage when operating near or below 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Guides assessment of site power quality before connecting the chiller.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Compliance

The equipment is EMC classified as Class A, Group 1 to EN55011, which has implications for use in residential environments.

Class A equipment may require additional EMC measures in domestic or light-industrial locations.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Ensures the correct plug type is supplied for the destination country.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Space requirements around the unit must be respected according to the operating conditions.

Air-cooled units need adequate clearance for heat rejection and service access.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”