Product name and general description of the chiller unit
Specpp. 1-2
Technical data, performance figures and operating parameters per DIN 12876
p. 2table, 9 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Net weight
61 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
220-240V 1~ 50Hz
max. current
15 A
Fuse
16 A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.2/24
Technical data according to DIN 12876 — Unichiller 022w-H (Order-No.: 3010.0122.01)
Parameter
Value
Operating temperature range
-10...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at ambient temperature +20°C at 15°C
2,2 kW
Cooling power at ambient temperature +20°C at 0°C
1,6 kW
Cooling power at ambient temperature +20°C at -10°C
1 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,43 kg
CO2 equivalent
0,6 t
UN-number
UN 2857
Circulation pump
Immersion pump
max. delivery
29 l/min
max. delivery pressure
1 bar
Delivery at 0,5 bar
21 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
95 l/h
Consumption at water 15°C, flow 0°C
96 l/h
Consumption at water 15°C, flow -10°C
77 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Overall dimensions WxDxH **
420x487x579 mm
** см. сноску в оригинале (на странице не раскрыта)
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water inlet temperature between 5°C and 20°C, and pipes must be insulated if water temperature can drop below the condensation point.
Defines the acceptable cooling water supply conditions and insulation requirement to prevent condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops and cooling water flow rate may increase as cooling water temperature rises, so sizing must account for actual site water temperature.
Prevents undersizing the cooling water supply and overestimating cooling capacity at elevated water temperatures.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility checks to avoid corrosion or damage.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Ensures the electrical supply stays within the allowed combined voltage and frequency envelope.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The plastic pump is not resistant to acetone and polyglycols, so these media must be avoided or a more resistant plastic option requested.
Prevents chemical attack on the pump when selecting heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Informs heat transfer fluid preparation for low-temperature operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Avoids delivery of the wrong plug type and ensures correct electrical connection.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
Three-phase devices below 63A are supplied with cable but without plug, and above 63A without cable or plug, so site wiring must be planned accordingly.
Clarifies scope of electrical supply and need for on-site plug or wiring.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance
The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1.
Determines EMC environment suitability and potential restrictions in residential areas.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Space requirements must be respected according to the operating conditions referenced by the manufacturer.
Ensures adequate clearance for operation and maintenance.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Safety
The unit has adjustable overtemperature protection according to DIN 12876.
Provides a safety function that must be configured for the application.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection
The refrigeration machine is water-cooled and uses CFC- and HCFC-free refrigerant R-449A with GWP 1397.
Affects cooling water availability and environmental/regulatory considerations.
Source
“Refrigeration machine water-cooled, CFC- and HCFC-free”