Selection
The chiller uses flammable refrigerant R-290 (A3, H220), so the installation site must be assessed for flammable-refrigerant safety and ventilation.
A3 flammability classification affects site selection, ventilation, and ignition-source control during installation.
Source
“R-290 (A3, H220)”
Installation
The unit requires a 220-240 V single-phase 50 Hz supply with up to 16 A current and a 20 A fuse.
Determines the electrical circuit, breaker/fuse rating, and single-phase supply needed at the installation point.
Source
“Power supply requirement 220-240V 1~ 50Hz”
Installation
Voltage and frequency tolerances are coupled: ±5% voltage only with simultaneous ±2% frequency, and certain combinations are not allowed.
Prevents commissioning on a supply whose combined voltage/frequency deviation exceeds the allowed envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The chiller is air-cooled and needs free space around it; space requirements must be respected.
Air-cooled condensers need clearance for airflow, otherwise cooling capacity drops and the unit may overheat.
Source
“Please respect space requirements.”
Performance
Cooling capacity is rated at 20 °C room temperature and will drop if ambient temperature rises.
Sizing must account for the actual ambient at the installation site, not just the 20 °C catalogue values.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Operation
The unit is only rated for ambient temperatures between 5 °C and 40 °C.
Operating outside this ambient range risks overload or shutdown; the room must stay within limits.
Source
“max. ambient temperature 40 °C”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used as heat-transfer media.
Prevents chemical attack and pump failure; choose a compatible fluid or request the more resistant plastic option.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
Protects the evaporator and circuit from freezing when operating near or below 0 °C.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Single/two-phase units are delivered with a power cable and country-specific plug that must be specified at ordering.
The plug type must be declared on the order, otherwise the unit may arrive with the wrong connector for the site.
Source
“with power cable and country-specific plug (please specify when ordering)”
Compliance
The equipment is EMC Class A, Group 1, which is intended for industrial environments and may cause interference in residential areas.
Class A equipment is not intended for residential use; plan EMC measures if installed near sensitive electronics.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has adjustable overtemperature protection per DIN 12876 and safety classification III / FL.
The overtemperature cut-out must be set correctly for the fluid and process to protect the bath and samples.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance
The refrigerant charge is only 0.25 kg of R-290 with GWP 0.02, so refrigerant handling and leak checks are simple but flammability rules still apply.
Small charge limits leak consequences, but A3 flammability still governs service and room requirements.
Source
“Refrigerant quantity 0,25 kg”