Induztr
Engineering Digest Document Viewer Related Models

Unichiller 025w OLÉ Technical Data

Contents

  1. Overview p. 1
    Product name and cover page of the datasheet
  2. Spec p. 2
    Technical data according to DIN 12876

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with a differential pressure between 1 and 6 bar and a G1/2 male connection.

To ensure the site cooling water system can provide adequate pressure and the correct fitting before installation.

Source
“min. cooling water differential pressure 1 bar”
Operation

The unit is designed for cooling water intake between 5°C and 20°C; lower temperatures require insulating the cooling water pipes to prevent condensation.

To avoid condensation and ensure reliable operation when specifying the cooling water supply temperature.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops and cooling water flow rate may increase as the cooling water temperature rises, so performance must be derated accordingly.

To correctly size the cooling water supply and predict actual cooling performance at site conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Selection

The plastic pump is not resistant to acetone and polyglycols, so alternative fluids or a more resistant pump option must be considered.

To avoid pump damage and select a compatible heat transfer fluid or upgraded pump material.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

To prevent freezing and ensure proper heat transfer fluid selection for low-temperature operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

To verify site power quality meets the allowed voltage and frequency tolerances before installation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

To ensure the correct plug type is ordered for the destination country and avoid installation delays.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The refrigerant R-449A has a Global Warming Potential of 1397 and a CO2 equivalent of 0.6 t, which matters for refrigerant handling and environmental compliance.

To assess environmental impact and comply with refrigerant regulations during installation and maintenance.

Source
“Global Warming Potential (GWP) 1397”
Operation

The unit has an autostart function for power failure, allowing automatic restart after power is restored.

To decide whether unattended restart after power loss is acceptable for the process.

Source
“Autostart function for power failure”
Selection

An optional Pt100 sensor connection can display (not control) process temperature and is only available factory fitted at additional cost.

To plan for process temperature monitoring and include the factory-fitted option in the order.

Source
“Option: Pt100 sensor connection #10519 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge)”
Installation

The unit has a protection degree of IP20 and must be installed in an environment with ambient temperature between 5°C and 40°C.

To ensure the installation location meets environmental and electrical protection requirements.

Source
“min. ambient temperature 5 °C”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable cooling water must be used to avoid corrosion.

To select compatible cooling water quality and prevent corrosion or fouling in the water circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”