The chiller is intended only for externally closed applications, so the user circuit must be closed.
Determines whether the unit can be used with an open bath or only with a closed external loop.
Source
“For externally closed applications.”
| Parameter | Value |
|---|---|
| min. filling capacity | 5 l |
| Volume of expansion | 7,5 l |
| Overall dimensions WxDxH ** | 640x937x1606 mm |
| Net weight | 283 kg |
| sound pressure level +/- 4 dB(A) | 74 dB(A) |
| Power supply (3 Phase) | 400V 3~ 50Hz |
| max. current (3 Phase) | 12 A |
| Fuse (3 phase) | 3x16A |
| Pressure equipment category | I |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | 1.1/26 |
** Please respect space requirements. See operating conditions at www.huber-online.com
| Parameter | Value |
|---|---|
| Order-No. | 3015.0095.01 |
| Operating temperature range | -20...100 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Temperature stability at -10°C | 0,2 K |
| Safety classification | III / FL |
| Heating power | 2 kW |
| Cooling power at 15°C | 5,5 kW |
| Cooling power at 0°C | 2,3 kW |
| Cooling power at -10°C | 0,8 kW |
| Cooling power at -20°C | 0,2 kW |
| Refrigeration machine | air-cooled, natural refrigerant |
| Refrigerant (ASHRAE, GHS) | R-290 (A3, H220) |
| Global Warming Potential (GWP) | 0,02 |
| Refrigerant quantity | 0,95 kg |
| CO2 equivalent | 0,0 t |
| UN-number | UN 3358 |
| Circulation pump: | C3 |
| max. delivery | 100 l/min |
| max. delivery pressure | 5,6 bar |
| Delivery at 0,3 bar | 97 l/min |
| Delivery at 1,0 bar | 90 l/min |
| Delivery at 1,5 bar | 84 l/min |
| Delivery at 2,0 bar | 77 l/min |
| Delivery at 2,5 bar | 71 l/min |
| Delivery at 3,0 bar | 64 l/min |
| Delivery at 3,5 bar | 57 l/min |
| Delivery at 4,0 bar | 48 l/min |
| Delivery at 4,5 bar | 39 l/min |
| Delivery at 5,0 bar | 27 l/min |
| Delivery at 5,5 bar | 8 l/min |
| Pump connection | G1 1/4 male |
The chiller is intended only for externally closed applications, so the user circuit must be closed.
Determines whether the unit can be used with an open bath or only with a closed external loop.
“For externally closed applications.”
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects installation and safety planning.
Flammable refrigerant classification requires appropriate site safety and handling measures.
“R-290 (A3, H220)”
The unit is air-cooled, so sufficient air flow and ambient temperature limits must be respected.
Air-cooled condensers need adequate ventilation and ambient conditions for rated performance.
“air-cooled, natural”
Cooling capacity drops strongly at lower temperatures, from 5.5 kW at 15 °C to 0.2 kW at -20 °C.
The process load must be checked against the actual cooling capacity at the required operating temperature.
“at 15°C 5,5 kW at 0°C 2,3 kW at -10°C 0,8 kW at -20°C 0,2 kW”
The unit requires a three-phase 400 V 3~ 50 Hz supply with a 3x16 A fuse.
Electrical supply and protection must be provisioned before installation.
“400V 3~ 50Hz”
Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.
The installer must plan and provide the correct plug and connection.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
The allowable ambient temperature range is 5 °C to 40 °C, and cooling capacity may fall above 20 °C.
Site ambient conditions must be within limits and performance derating considered.
“max. ambient temperature 40 °C”
Output data are valid at 20 °C room temperature, and higher ambient temperature may reduce cooling capacity.
Selection calculations should account for actual ambient temperature at the installation site.
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Voltage and frequency tolerances are ±5% and ±2% respectively, but not all combinations are allowed.
The power quality at the site must be checked to avoid invalid supply combinations.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
The equipment is EMC classified as Class A, Group 1, which has implications for use in residential environments.
Class A equipment may require additional EMC measures in residential or sensitive environments.
“Classification (disturbance) to EN55011: Class A, Group 1”
The unit has IP20 protection, so it must be protected from water and dust ingress.
The installation location must avoid moisture and conductive dust.
“Degree of Protection IP20”
The minimum filling capacity is 5 l and expansion volume is 7.5 l, which must be considered when filling the loop.
Insufficient fill volume or expansion space can cause operational problems.
“min. filling capacity 5 l”
The unit is heavy at 283 kg and large at 640x937x1606 mm, requiring adequate floor space and handling equipment.
Transport, floor loading and clearance must be planned before installation.
“640x937x1606 mm”