Product name and general description of the chiller unit
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-10...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
5,5 kW
Cooling power at 0°C
3 kW
Cooling power at -10°C
1,5 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,95 kg
CO2 equivalent
1,3 t
Circulation pump
C3
max. delivery
100 l/min
max. delivery pressure
5,6 bar
Delivery at 0,3 bar
97 l/min
Delivery at 1,0 bar
90 l/min
Delivery at 1,5 bar
84 l/min
Delivery at 2,0 bar
77 l/min
Delivery at 2,5 bar
71 l/min
Delivery at 3,0 bar
64 l/min
Delivery at 3,5 bar
57 l/min
Delivery at 4,0 bar
48 l/min
Delivery at 4,5 bar
39 l/min
Delivery at 5,0 bar
27 l/min
Delivery at 5,5 bar
8 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
330 l/h
Order-No.: 3015.0079.01; Unichiller 055Tw-H
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a differential pressure of at least 0.6 bar and no more than 6 bar.
The site cooling water system must be able to deliver this pressure window, otherwise the unit will not operate correctly or may be damaged.
Source
“min. cooling water differential pressure
0,6 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake temperatures between 5 °C and 20 °C, and pipes must be insulated if the water can fall below the condensation point.
Cooling water outside this range changes cooling capacity and can cause condensation, so the facility water supply must be checked before installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
The published cooling capacities are only valid at 15 °C cooling water, so the actual duty must be derated if the site water is warmer.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Water quality and treatment must be compatible with these materials to avoid corrosion or fouling of the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
The three-phase unit is supplied with a cable but without a plug, so the customer must provide a suitable connection.
The electrical installation must include a suitable 3-phase plug or isolator, since none is supplied with the unit.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are ±5 % and ±2 % respectively, but a -5 % voltage combined with a +2 % frequency is not allowed.
The site power quality must be checked against this combined tolerance rule, not just against each tolerance separately.
Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance
The unit is EMC classified as Class A, Group 1 to EN55011, which is intended for industrial environments.
Class A equipment may cause interference in residential areas, so installation location and EMC environment must be considered.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The refrigerant is R-449A, classified A1 with GHS hazard H280, and the charge is 0.95 kg with a CO2 equivalent of 1.3 t.
Refrigerant safety class, hazard labelling and CO2 equivalent affect handling, leak detection and F-gas reporting obligations.
Source
“R-449A (A1, H280)”
Operation
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
The overtemperature cut-out must be set correctly for the application to protect the process and the equipment.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection
The circulation pump delivers 100 l/min maximum, but flow falls to 8 l/min at 5.5 bar, so the system pressure drop must be matched to the pump curve.
External circuit pressure drop determines the actual flow, so the pump curve must be checked against the application resistance.