Unichiller 075-H Technical Data

Contents

  1. Overview p. 1
    Introduces the Unichiller 075-H model and its purpose
  2. Spec pp. 1-2
    Technical data and performance parameters according to DIN 12876
  3. p. 2 table, 9 rows
  4. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)13 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S504197

Technical data according to DIN 12876

ParameterUnichiller 075-H
Order-No.3040.0032.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C
at 100°C7,5 kW
at 20°C7,5 kW
at 0°C6,1 kW
at -10°C4 kW
at -20°C2,4 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,03 kg
CO2 equivalent1,4 t
Circulation pump:E1
max. delivery48 l/min
max. delivery pressure3,4 bar
Delivery at 0,3 bar46 l/min
Delivery at 0,5 bar44 l/min
Delivery at 1,0 bar40 l/min
Delivery at 1,5 bar33 l/min
Delivery at 2,0 bar27 l/min
Delivery at 2,5 bar20 l/min
Delivery at 3,0 bar11 l/min
Pump connectionG1 1/4 male
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1165 mm
Net weight298 kg

** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window.

Defines whether the unit can cover the required process temperatures before selection.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity drops from 7.5 kW at 20 °C to 2.4 kW at -20 °C, so low-temperature duty needs a larger unit or derating check.

Prevents undersizing when the process runs at the cold end of the range.

Source
“at -20°C 2,4 kW”
Installation

The unit requires a 400 V three-phase supply with 13 A max current and 3x16 A fusing.

Determines electrical supply, cable and protection sizing at the installation site.

Source
“Power supply (3 Phase) 400V 3~ 50Hz max. current (3 Phase) 13 A Fuse (3 phase) 3x16A”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the site must provide a suitable connection.

Avoids surprises at commissioning because no country plug is supplied.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The pump connection is G1 1/4 male, so hoses and adapters must match this thread.

Ensures correct hydraulic connection and avoids leaks or adapter mismatch.

Source
“Pump connection G1 1/4 male”
Installation

The minimum filling capacity is 18 l and expansion volume is 48 l, which sets the minimum loop volume and expansion allowance.

Needed to size the external circuit and expansion tank correctly.

Source
“min. filling capacity 18 l Volume of expansion 48 l”
Installation

The unit is air-cooled and requires ambient temperature between 5 °C and 40 °C.

Room conditions must be checked because cooling capacity falls as ambient rises.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is valid at 20 °C room temperature and may drop if ambient temperature rises.

Prevents overestimating capacity in a warm plant room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Safety

The refrigerant is R-449A with GHS H280, so leak handling and ventilation must account for a pressurised gas hazard.

Affects safety assessment and refrigerant handling procedures.

Source
“R-449A (A1, H280)”
Compliance

The unit is classified Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require EMC precautions in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Voltage and frequency tolerances are ±5 % and ±2 % but not simultaneously in the unfavourable combination.

Site power quality must be checked to avoid invalid supply conditions.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The unit is intended for externally closed applications, which constrains the hydraulic circuit design.

Prevents use in open-loop or unsuitable external circuits.

Source
“For externally closed applications.”
Maintenance

The refrigerant charge is 1.03 kg with CO2 equivalent 1.4 t, which matters for leak checks and refrigerant records.

Supports maintenance planning and regulatory refrigerant accounting.

Source
“Refrigerant quantity 1,03 kg CO2 equivalent 1,4 t”