Unichiller 100Tw-H Technical Datasheet

Contents

  1. Overview p. 1
    Product introduction and general description of the chiller unit
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters per DIN 12876
  3. p. 2 table, 18 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow 0°C510 l/h
Consumption at water 15°C, flow -10°C375 l/h
Consumption at water 15°C, flow -20°C276 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity6,5 l
Volume of expansion5,5 l
Overall dimensions WxDxH **600x600x1450 mm
Net weight235 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)15 A
Fuse (3 phase)3x20A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S457081

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 100Tw-H (Order-No.: 3017.0039.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C10 kW
Cooling power at 0°C10 kW
Cooling power at -10°C6,5 kW
Cooling power at -20°C3 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,75 kg
CO2 equivalent2,4 t
Circulation pumpC3
max. delivery96 l/min
max. delivery pressure5,6 bar
Delivery at 0,5 bar92 l/min
Delivery at 1,0 bar86 l/min
Delivery at 1,5 bar81 l/min
Delivery at 2,0 bar75 l/min
Delivery at 2,5 bar68 l/min
Delivery at 3,0 bar62 l/min
Delivery at 3,5 bar54 l/min
Delivery at 4,0 bar46 l/min
Delivery at 4,5 bar37 l/min
Delivery at 5,0 bar24 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 15°C510 l/h

Order-No.: 3017.0039.01

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 3x20 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable cross-section and protection device needed before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

The unit is water-cooled and needs a cooling water supply with at least 1 bar differential pressure and no more than 6 bar.

Cooling water pressure must be verified on site to guarantee rated cooling performance and avoid damage.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation occurs.

Out-of-range cooling water temperature reduces cooling power and can cause condensation and corrosion.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops from 10 kW at 0 °C to 3 kW at -20 °C, so low-temperature duty must be checked against the load.

Ensures the selected chiller can actually remove the process heat at the lowest required operating temperature.

Source
“at -20°C 3 kW”
Performance

The circulation pump delivers 96 l/min at zero head but only 24 l/min at 5.0 bar, so system pressure drop must be matched to the pump curve.

Prevents undersizing of flow in circuits with high pressure loss and confirms achievable flow at the application's back pressure.

Source
“max. delivery 96 l/min”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 1.75 kg corresponding to 2.4 t CO2 equivalent.

Affects refrigerant handling, leak reporting obligations and room ventilation requirements.

Source
“R-449A (A1, H280)”
Operation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.

Defines the acceptable installation environment and excludes wet or dusty locations without additional protection.

Source
“max. ambient temperature 40 °C”
Installation

The chiller needs a minimum filling capacity of 6.5 l and has 5.5 l expansion volume, which sets the heat transfer fluid charge.

Determines the volume of heat transfer fluid to prepare and the expansion tank sizing for the loop.

Source
“min. filling capacity 6,5 l”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality selection to avoid corrosion and fouling of the wetted materials.

Source
“Please use suitable cooling water.”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combined deviations outside this are not allowed.

Protects the compressor and electronics from supply quality issues at the installation site.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.

Avoids delays at commissioning because the customer must supply the mains connector.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit emits a sound pressure level of 65 dB(A) ±4 dB(A), which must be considered for the installation location.

Relevant for workplace noise limits and placement near occupied areas.

Source
“sound pressure level +/- 4 dB(A) 65 dB(A)”
Selection

The chiller has a safety classification of III / FL according to DIN 12876 and adjustable overtemperature protection.

Confirms the safety class and protective function required for flammable or sensitive bath fluids.

Source
“With adjustable overtemperature protection according to DIN 12876.”