Unichiller 110T-H Technical Data

Contents

  1. Overview p. 1
    Introduces the Unichiller 110T-H chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 6 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Technical data according to DIN 12876 — Unichiller 110T-H (Order-No.: 3027.0079.01)

ParameterValue
Operating temperature range-10...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C11 kW
Cooling power at 0°C6 kW
Cooling power at -10°C2,7 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:C3
max. delivery96 l/min
max. delivery pressure5,6 bar
Delivery at 0,5 bar91 l/min
Delivery at 1,0 bar86 l/min
Delivery at 1,5 bar81 l/min
Delivery at 2,0 bar75 l/min
Delivery at 2,5 bar69 l/min
Delivery at 3,0 bar63 l/min
Delivery at 3,5 bar56 l/min
Delivery at 4,0 bar48 l/min
Delivery at 4,5 bar40 l/min
Delivery at 5,0 bar29 l/min
Pump connectionG1 1/4 male
min. filling capacity8,36 l
Volume of expansion7 l
Overall dimensions WxDxH **600x790x1614 mm
Net weight232 kg
sound pressure level +/- 4 dB(A)74 dB(A)

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -10 to 100 °C, so it must be matched to the process temperature window.

Defines whether the unit can cover the required process temperatures before selection.

Source
“Operating temperature range -10...100 °C”
Performance

Cooling capacity drops sharply with lower temperature (11 kW at 15 °C, 6 kW at 0 °C, 2.7 kW at -10 °C), so sizing must use the capacity at the actual set point.

Prevents undersizing when the process runs cold rather than at ambient.

Source
“at 15°C 11 kW at 0°C 6 kW at -10°C 2,7 kW”
Installation

The unit requires a 3-phase 400 V 50 Hz supply, so site power must be verified before installation.

Determines electrical infrastructure and whether a suitable three-phase feed is available.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the plug and connection.

Affects electrical installation scope and procurement of the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are linked: ±5% voltage only with ±2% frequency, and opposite-sign deviations are not allowed.

Guides power quality assessment at the installation site to avoid invalid supply conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

Cooling capacity is specified at 20 °C room temperature and may fall if ambient temperature rises, so the room must be kept within limits.

Affects thermal design and ventilation of the installation room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The permitted ambient temperature range is 5 to 40 °C, so the installation location must stay within these limits.

Defines site conditions required for reliable operation and warranty compliance.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit is IP20, so it must be installed in a dry, protected location without water ingress risk.

Restricts placement in wet or dusty environments and informs enclosure planning.

Source
“Degree of Protection IP20”
Installation

The pump connection is G1 1/4 male, so hoses and fittings must match this thread.

Ensures correct hydraulic connection hardware is ordered.

Source
“Pump connection G1 1/4 male”
Performance

Pump delivery falls with rising pressure (96 l/min max, 29 l/min at 5.0 bar), so the system pressure drop must be checked against the pump curve.

Prevents insufficient flow in high-pressure-drop external circuits.

Source
“max. delivery 96 l/min max. delivery pressure 5,6 bar”
Safety

The refrigerant is R-449A with GHS classification H280 and GWP 1397, so refrigerant handling and leak response must follow safety rules.

Informs safety planning, labeling, and environmental compliance for the refrigerant charge.

Source
“R-449A (A1, H280)”
Safety

The unit has adjustable overtemperature protection per DIN 12876, which must be set for the application.

Critical for protecting temperature-sensitive process fluids and equipment.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Compliance

The equipment is EMC Class A Group 1, so it may cause radio interference in residential areas and is intended for industrial use.

Affects site selection and EMC planning, especially near sensitive electronics.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The expansion tank is atmospheric open and has a cover, so the tank must be kept covered and checked for level.

Open tanks are prone to evaporation and contamination, requiring periodic level checks.

Source
“atmospheric open expansion tank”