Unichiller 200T Technical Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller 200T chiller unit and its main features
  2. Spec p. 2
    Lists technical data according to DIN 12876 standard
  3. p. 1 table, 34 rows

Technical data according to DIN 12876 — Unichiller 200T (Order-No.: 3028.0148.01)

CharacteristicValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C20 kW
Cooling power at 0°C10 kW
Cooling power at -10°C5 kW
Cooling power at -20°C3 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x1582x1902 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller's cooling capacity drops as ambient temperature rises, so size the unit for the actual room temperature rather than the 20 °C catalogue values.

Prevents undersizing when the installation room runs warmer than 20 °C.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires an external branch circuit protection device per UL 489, which must be provided by the installer.

Affects electrical panel design and compliance of the installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase units drawing less than 63 A are shipped with a cable but without a plug, so a suitable plug must be sourced on site.

Avoids surprises at commissioning when no mating plug is available.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances must be met simultaneously, and certain combinations such as -5% voltage with +2% frequency are not permitted.

Guides verification of the site power quality before connecting the chiller.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit is rated IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

Determines suitable installation environment and enclosure requirements.

Source
“Degree of Protection IP20”
Operation

The chiller is intended only for externally closed applications, which constrains the hydraulic circuit design.

Prevents use in open-loop or vented process circuits that would void the design intent.

Source
“For externally closed applications.”
Operation

Ambient temperature must stay between 5 °C and 40 °C for correct operation.

Defines the allowable room conditions for reliable operation.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.

Informs site safety assessment and service procedures for the refrigeration circuit.

Source
“R-449A (A1, H280)”
Compliance

The equipment emits Class A, Group 1 disturbances per EN55011, which typically requires industrial-environment EMC measures.

Affects EMC planning and suitability for residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Compliance

The unit complies with US-SNAP for industrial process refrigeration end-use, which matters for US installations.

Confirms regulatory acceptability for the intended US application.

Source
“The US-SNAP end-use for this equipment is the industrial process refrigeration.”
Maintenance

The expansion tank is atmospherically open and has a cover included, so level checks and venting are part of routine upkeep.

Signals maintenance tasks and the need to keep the tank cover in place.

Source
“atmospheric open expansion tank”
Selection

The pump delivers 202 l/min at 1.1 bar but only 77 l/min at 4.0 bar, so system pressure drop must be matched to the pump curve.

Ensures the selected pump curve covers the actual external circuit resistance.

Source
“Delivery at 4,0 bar 77 l/min”
Installation

Overall dimensions are 905x1582x1902 mm and space requirements must be respected, so plan adequate clearance around the unit.

Prevents installation in spaces too tight for service access and airflow.

Source
“** Please respect space requirements.”