Unichiller 260T Technical Data Sheet

Contents

  1. Overview p. 1
    Product name and general description of the Unichiller 260T
  2. Spec p. 2
    Technical data according to DIN 12876 for the chiller
  3. p. 1 table, 34 rows

Technical data according to DIN 12876 — Unichiller 260T (Order-No.: 3058.0003.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C26 kW
Cooling power at 0°C26 kW
Cooling power at -10°C16 kW
Cooling power at -20°C10 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:D3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x2172x1900 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH — как напечатано в исходной таблице.

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller's cooling capacity drops as ambient temperature rises, so selection must account for actual room conditions.

Cooling capacity ratings are valid only at 20°C room temperature; higher ambient reduces available capacity.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

Three-phase units under 63A are supplied with cable but no plug, so the installer must provide a suitable plug.

Determines what electrical termination hardware must be procured and installed on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

External branch circuit protection per UL 489 is mandatory and must be specified from the electrical schematics.

The unit does not include branch protection; the site electrical design must provide it.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are interdependent: -5% voltage with +2% frequency is not allowed.

Supply quality must be checked against the combined tolerance rule, not voltage alone.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit requires minimum clearance and specific operating conditions that must be respected during siting.

Air-cooled refrigeration needs adequate airflow clearance; ignoring this degrades capacity and reliability.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.

A1 classification and H280 (pressurized gas) affect service procedures and room ventilation requirements.

Source
“R-449A (A1, H280)”
Compliance

The equipment emits Class A, Group 1 disturbances per EN55011, which constrains its use in residential environments.

Class A equipment is intended for industrial environments and may cause interference in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The unit is rated IP20, so it must be installed in a dry, protected location away from water ingress.

IP20 offers no water protection; installation location must avoid splashes and condensation.

Source
“Degree of Protection IP20”
Operation

The ambient operating window is 5°C to 40°C, limiting where and when the chiller can run reliably.

Operation outside this range risks poor performance or equipment damage.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The pump delivers 210 l/min at zero head but only 31 l/min at 4.5 bar, so system pressure drop must be matched to the pump curve.

External circuit pressure drop determines actual flow; oversizing the loop starves flow.

Source
“max. delivery 210 l/min max. delivery pressure 4,7 bar”
Installation

The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread.

Thread mismatch prevents connection; correct couplings must be ordered.

Source
“Pump connection G1 1/4 male”
Operation

The expansion vessel volume is 39 l with a minimum filling capacity of 14 l, defining the minimum system fill.

Underfilling below 14 l or exceeding expansion volume causes poor temperature control or overflow.

Source
“min. filling capacity 14 l Volume of expansion 39 l”
Selection

Temperature stability is 0.5 K at -10°C, which sets the achievable control accuracy for the process.

Processes requiring tighter than ±0.5 K need a different model or control upgrade.

Source
“Temperature stability at -10°C 0,5 K”
Operation

The unit is designed for externally closed applications, so an open external loop is not permitted.

Using an open loop risks evaporation, contamination, and loss of temperature control.

Source
“For externally closed applications.”