Product name and general introduction to the chiller
Specp. 2
Technical data according to DIN 12876
p. 1table, 27 rows
Unichiller GL 250w — Technical data according to DIN 12876
Parameter
Value
Order-No.
5509.0005.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 20°C
25 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)
Global Warming Potential (GWP)
1
UN-number
UN 2857
max. delivery
100 l/min
max. delivery pressure
3,5 bar
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
Volume of expansion
30 l
Overall dimensions WxDxH **
730x804x1738 mm
Power supply (3 Phase)
460V 3~ 60Hz
Pressure equipment category
II
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
1.0/26
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1 bar and a maximum supply pressure of 6 bar, so the site water system must be sized accordingly.
Ensures the cooling water circuit can deliver adequate flow and pressure without overpressurizing the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is water-cooled and must be connected to cooling water with G3/4 male fittings, which dictates the piping interface at the installation site.
Needed to prepare matching hoses and fittings before delivery and to avoid installation delays.
Source
“Cooling water connection G3/4 male”
Operation
The chiller is designed for cooling water intake between 5°C and 20°C, and lower water temperatures require insulating the cooling water pipes inside the unit.
Operating outside this range or without insulation can cause condensation and performance or reliability issues.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity is rated at 20°C room temperature and 15°C cooling water inlet with 1 bar differential pressure, so actual capacity will differ if site conditions deviate.
Prevents overestimating available cooling power when site water or ambient temperatures are higher.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Performance
Higher cooling water temperatures reduce cooling power and may require increased cooling water flow rate, so the water supply must be evaluated for worst-case conditions.
Affects selection margin and ensures the cooling water supply can handle increased flow demand.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used to avoid corrosion or damage.
Guides water quality requirements and material compatibility checks before connecting the unit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Ensures the electrical supply at the site stays within the allowed envelope to avoid invalid warranty or malfunction.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
External branch circuit protection according to UL 489 is required, so the electrical installation must include a suitable UL 489 breaker.
Needed for code-compliant and safe electrical connection of the chiller.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase devices with current consumption less than 63A are delivered with a cable but without a plug, so the site must provide a suitable connection point.
Avoids assuming a plug is included and allows proper electrical termination planning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The unit uses refrigerant R-744 (A1, H280) with GWP 1 and UN number UN 2857, which affects transport, handling and refrigerant safety classification.
Important for safety assessments, transport documentation and refrigerant handling procedures.
Source
“R-744 (A1, H280)”
Selection
The chiller has IP20 protection and must be installed in an environment between 5°C and 40°C ambient temperature.
Confirms the unit is for indoor, dry locations and that ambient conditions must be controlled.
Source
“Degree of Protection IP20”
Installation
The overall dimensions are 730x804x1738 mm and space requirements must be respected, so the installation location must accommodate the unit and clearance.
Prevents installation in insufficient space that could obstruct operation or maintenance.