Product name and general description of the chiller unit
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 27 rows
Unichiller GL 500w — Technical data according to DIN 12876
Characteristic
Value
Order-No.
5512.0001.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 20°C
50 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)
Global Warming Potential (GWP)
1
UN-number
UN 2857
max. delivery
100 l/min
max. delivery pressure
3,5 bar
Pump connection
G1 1/4 male
Cooling water connection
G1 1/4 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
Volume of expansion
30 l
Overall dimensions WxDxH **
730x804x1738 mm
Power supply (3 Phase)
400V 3~ 50Hz
Pressure equipment category
II
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
1.0/26
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1 bar between inlet and outlet.
Ensures the cooling water circuit can provide adequate flow; insufficient differential pressure will impair cooling performance.
Source
“min. cooling water differential pressure”
Installation
The maximum permissible cooling water pressure is 6 bar.
Prevents overpressure damage to the water-cooled refrigeration circuit and connected piping.
Source
“max. cooling water pressure”
Installation
The unit must be operated with cooling water intake temperature between 5°C and 20°C.
Defines the allowable cooling water supply temperature range for proper operation and to avoid condensation issues.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.
Prevents condensation and potential water damage or corrosion when using cold cooling water.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling power drops as cooling water temperature increases, and the required cooling water flow rate may increase.
Affects sizing of the cooling water supply and expected performance under varying site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.
Guides water quality requirements to avoid corrosion or damage to the mixed-material cooling water circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Compliance
Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but combinations such as -5% voltage and +2% frequency are not allowed.
Ensures the electrical supply stays within safe operating limits and avoids invalid tolerance combinations.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices with current consumption less than 63A are delivered with a cable but without a plug.
The installer must provide and connect a suitable plug or direct connection for this three-phase unit.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The unit uses refrigerant R-744 (A1, H280), which is classified with GHS hazard H280.
Informs about refrigerant safety classification and hazard communication for handling and site safety.
Source
“R-744 (A1, H280)”
Installation
The unit has protection degree IP20 and must be installed in an environment with ambient temperature between 5°C and 40°C.
IP20 means no water protection and limited finger protection; ambient limits affect location and ventilation.
Source
“Degree of Protection
IP20”
Selection
The chiller is intended for externally closed applications.
Confirms the hydraulic configuration required; open-loop or unsealed external circuits are not suitable.
Source
“For externally closed applications.”
Installation
The pump and cooling water connections are G1 1/4 male.
Determines the mating fittings and hoses needed for installation.
Source
“Pump connection
G1 1/4 male”
Operation
The temperature stability at -10°C is 0.2 K.
Indicates achievable temperature control accuracy for process requirements.
Source
“Temperature stability at -10°C
0,2 K”
Compliance
The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1.
Class A equipment may require additional EMC precautions in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”