Product introduction and key features of the chiller
Specp. 2
Technical data according to DIN 12876
p. 2table, 15 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Consumption at water 15°C, flow -20°C
36 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Overall dimensions WxDxH **
350x496x622 mm
Net weight
57 kg
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
14 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S586152
** Please respect space requirements. See operating conditions at www.huber-online.com
Unichiller P007w-H — Technical data according to DIN 12876
Parameter
Value
Order-No.
3012.0265.01
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at ambient temperature +20°C at 15°C
0,7 kW
Cooling power at ambient temperature +20°C at 0°C
0,55 kW
Cooling power at ambient temperature +20°C at -10°C
0,4 kW
Cooling power at ambient temperature +20°C at -20°C
0,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,2 kg
CO2 equivalent
0,3 t
Circulation pump
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
66 l/h
Consumption at water 15°C, flow 0°C
54 l/h
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires cooling water inlet temperature between 5°C and 20°C, and below the condensation point the cooling water pipes inside the unit must be insulated.
Defines the permissible cooling water supply temperature range and insulation requirement for correct installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The cooling water circuit must be supplied with at least 1 bar differential pressure and must never exceed 6 bar.
Determines the required cooling water supply pressure and the maximum allowable pressure for safe connection.
Source
“min. cooling water differential pressure”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides selection of cooling water quality and avoids corrosion or damage to wetted materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant plastic pump ordered.
Prevents chemical attack on the pump when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Specifies the recommended antifreeze approach for low-temperature operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Compliance
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.
Defines the acceptable mains supply quality for reliable operation and warranty compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
Ensures the correct mains plug is ordered for the installation site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Selection
The unit is classified as EN55011 Class A Group 1 for electromagnetic disturbance, which affects its suitability in residential environments.
Class A equipment may not be suitable for residential use and must be considered during site selection.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection
The refrigeration machine is water-cooled and uses CFC- and HCFC-free refrigerant R-449A with GWP 1397 and 0.2 kg charge.
Informs refrigerant type, environmental impact and the need for a cooling water supply.
Source
“Refrigeration machine
water-cooled, CFC- and
HCFC-free”
Performance
Cooling power drops from 0.7 kW at +20°C ambient to 0.2 kW at -20°C, so low-temperature duty must be sized accordingly.
Allows correct sizing of cooling capacity across the operating temperature range.
Source
“Cooling power at ambient temperature +20°C”
Installation
The unit requires a 208-240V 1~/2~ 50/60Hz supply with maximum current 14 A and a 16 A fuse.
Defines electrical supply and protection requirements for installation.
Source
“Power supply requirement
208-240V 1~/2~ 50/60Hz”
Installation
The unit has IP20 protection and must be operated between 5°C and 40°C ambient temperature.
Sets environmental limits and ingress protection for the installation location.
Source
“Degree of Protection
IP20”
Maintenance
The unit has a minimum filling capacity of 3.8 l and an expansion volume of 1.7 l, which must be respected when filling the bath.
Ensures correct fluid fill level and expansion allowance for safe operation.
Source
“min. filling capacity
3,8 l”
Selection
The pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with delivery falling to 6 l/min at 2.0 bar.
Allows matching pump performance to the pressure drop of the external application.