Unimotive GL 07w — Technical Data Sheet

Contents

  1. Overview p. 1
    Product introduction and general description of the unit
  2. Spec pp. 1-2
    Technical data and performance parameters per DIN 12876
  3. p. 2 table, 15 rows
  4. p. 1 table, 26 rows

Technical data according to DIN 12876

ParameterValue
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight511 kg
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unimotive GL 07w (Order-No.: 5031.0001.01)

CharacteristicValue
Operating temperature range-45...95 °C
Temperature stability at -10°C0,05 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Water/ Glysantin 40/60 at 20°C12 kW
Cooling power with Water/ Glysantin 40/60 at 0°C10 kW
at -20°C7 kW
at -30°C4,5 kW
at -40°C2,5 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
Circulation pump:MK pump
max. delivery150 l/min
max. delivery pressure5,3 bar
Pump connectionM38x1,5 male
max. permissible kin. viscosity200 mm²/s

Cooling power values at 0°C, -20°C, -30°C, -40°C относятся к строке 'Cooling power with Water/ Glysantin 40/60'.

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is designed for water-ethylene glycol mixtures with corrosion protection down to -45°C, so it suits automotive temperature simulation and material testing.

Confirms the intended heat transfer fluid and low-temperature capability before specifying the unit for a given application.

Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation

Cooling water must be supplied at 5–20°C inlet with 1–4 bar differential pressure and no more than 6 bar, which constrains the facility water supply design.

The site cooling water loop must meet these limits or cooling performance and safe operation are compromised.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature drops below the condensation point, the cooling water pipes inside the unit must be insulated to prevent condensation.

Prevents condensation and corrosion/electrical risk when cold cooling water is used.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops as cooling water temperature rises, and cooling water flow rate may need to increase, so performance must be derated accordingly.

Prevents overestimating available cooling capacity at higher site water temperatures.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to avoid corrosion or fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Defines acceptable power supply quality for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.

Affects electrical installation design and protective device selection.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices under 63 A are supplied with cable but without plug, so a suitable plug must be provided on site.

Ensures the electrical connection hardware is planned before commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Pump connections use bore shape Y (60°) per DIN 3863 with ball socket and sleeve nut per DIN 3870, so matching hoses and fittings are required.

Prevents mismatch of hydraulic connections during installation.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for safe operation.

Safety function must be configured correctly for the application and fluid.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection

The optional Flow Control Cube enables precise flow rate measurement and control, which may be needed for accurate testing.

Determines whether the optional flow control accessory should be specified.

Source
“The optionally available "Flow Control Cube" enables precise flow rate measurement and control.”
Installation

Space requirements must be respected, with overall dimensions 730x804x1738 mm and 511 kg net weight.

Ensures adequate floor space and structural support for the unit.

Source
“Please respect space requirements.”