Product introduction and general description of the unit
Specp. 2
Technical data according to DIN 12876
p. 2table, 23 rows
p. 1table, 26 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 1,0 bar
179 l/min
Delivery at 2,0 bar
154 l/min
Delivery at 3,0 bar
124 l/min
Delivery at 4,0 bar
87 l/min
Delivery at 5,0 bar
42 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
200 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
2100 l/h
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
12 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
918x963x1771 mm
Net weight
700 kg
Power supply requirement
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unimotive GL 20w-XT (Order-No.: 5033.0002.01)
Parameter
Value
Operating temperature range
-45...150 °C
Temperature stability at -10°C
0,05 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
I / NFL
Heating power
24 kW
Cooling power with Water/ Glysantin 40/60 at 20°C
28 kW
Cooling power with Water/ Glysantin 40/60 at 0°C
25 kW
at -20°C
18,6 kW
at -30°C
11 kW
at -40°C
6,8 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)
Global Warming Potential (GWP)
1
Refrigerant quantity
3,3 kg
CO2 equivalent
0,0 t
Circulation pump:
MK pump
max. delivery
201 l/min
max. delivery pressure
5,3 bar
Значения приведены дословно; строки 'Cooling power with Water/ Glysantin 40/60' и последующие 'at ...' относятся к охлаждающей мощности при соответствующих температурах.
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is designed for water-ethylene glycol mixtures with corrosion protection down to -45°C, so it suits automotive temperature simulation and material testing rather than plain water duty.
Confirms the intended heat transfer fluid and low-temperature capability before specifying the unit for a given application.
Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation
Cooling water must be supplied between 5°C and 20°C, and if it drops below the condensation point the cooling water pipes inside the unit must be insulated.
Defines the site cooling water conditions and the need for pipe insulation to avoid condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The cooling water circuit requires a differential pressure between 1 and 4 bar and must not exceed 6 bar static pressure.
Sets the facility cooling water supply requirements for correct and safe operation.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit needs a 400 V three-phase 50 Hz supply and, being above 63 A, is delivered without cable and without plug.
Affects electrical installation planning, cabling and on-site connection responsibility.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Prevents specifying the unit on a supply where the combined tolerance window is exceeded.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit is IP20 and must be operated between 5°C and 40°C ambient, so it needs a protected indoor location with adequate space.
Constrains the installation room and ventilation/space planning.
Source
“max. ambient temperature 40 °C”
Performance
Cooling capacity falls with lower temperature, from 28 kW at 20°C to 6.8 kW at -40°C, so the duty point must be checked against the actual process temperature.
Prevents oversizing assumptions based only on the 20°C figure when the process runs cold.
Source
“at -40°C 6,8 kW”
Performance
Cooling capacity figures are valid only at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, and higher cooling water temperature reduces capacity.
Requires derating of the published cooling capacity for actual site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Performance
The circulation pump delivers 201 l/min at zero head but only 42 l/min at 5.0 bar, so the operating point must be matched to the external circuit resistance.
Ensures the pump curve is checked against the pressure drop of the external application.
Source
“Delivery at 5,0 bar 42 l/min”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 200 mm²/s, which limits fluid choice at low temperatures.
Prevents selecting a fluid that becomes too viscous at the low end of the temperature range.
Source
“max. permissible kin. viscosity 200 mm²/s”
Safety
The refrigerant is R-744 (CO2) with GWP 1 and 3.3 kg charge, giving 0.0 t CO2 equivalent, which simplifies refrigerant compliance and leak reporting.
Relevant for refrigerant regulations, risk assessment and environmental reporting.
Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280)”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification I/NFL, which must be set for the actual fluid and application.
Defines the safety device that must be configured correctly during commissioning.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation
The pump connection is M38x1.5 male with bore shape Y (60°) per DIN 3863, requiring matching ball socket and sleeve nut fittings.
Ensures correct hose and fitting selection for leak-free connection.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and treatment to avoid corrosion and fouling of the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”