Product introduction and general description of the Unistat 410w unit
Specpp. 1-2
Technical performance data and operating parameters per DIN 12876
p. 2table, 22 rows
p. 1table, 28 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 0,4 bar
42 l/min
Delivery at 0,6 bar
33 l/min
Delivery at 0,8 bar
20 l/min
Pump connection
M24x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
280 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
1,5 l
Filling capacity expansion tank
2,8 l
Overall dimensions WxDxH **
426x360x631 mm
Net weight
68 kg
sound pressure level +/- 4 dB(A)
61 dB(A)
Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current
16 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S577520
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unistat 410w (Order-No.: 1066.0013.01)
Parameter
Value
Operating temperature range
-45...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
3 kW
Cooling power with Thermooil at 250°C
1,3 kW
Cooling power with Thermooil at 200°C
1,5 kW
Cooling power with Thermooil at 100°C
2,5 kW
Cooling power with Thermooil at 20°C
2,5 kW
Cooling power with Ethanol at 0°C
1,5 kW
Cooling power with Ethanol at -20°C
0,8 kW
Cooling power with Ethanol at -40°C
0,17 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
Refrigerant quantity
0,098 kg
CO2 equivalent
0,0 t
UN-number
UN 3358
Circulation pump: max. delivery
56 l/min
Circulation pump: max. delivery pressure
0,9 bar
Circulation pump: Delivery at 0,2 bar
50 l/min
Таблица транспонирована: характеристики в строках, значения в колонке Value. Cooling power сгруппирована по средам (Thermooil / Ethanol) и температурам, как напечатано.
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is water-cooled and requires a cooling water supply, so a suitable water circuit must be available at the installation site.
Determines site requirements and rules out air-cooled installation.