Product introduction and general description of the Unistat 520w unit
Specpp. 1-2
Technical data and performance specifications according to DIN 12876
p. 2table, 20 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
492 l/h
Consumption at water 15°C, flow 0°C
492 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
4,9 l
Filling capacity expansion tank
16 l
Overall dimensions WxDxH **
540x604x1332 mm
Net weight
210 kg
sound pressure level +/- 4 dB(A)
61 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
14 A
Fuse
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S388606
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unistat 520w (Order-No.: 1072.0001.01)
Parameter
Value
Operating temperature range
-55...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
6 kW
Cooling power with Thermooil at 200°C
6 kW
Cooling power with Thermooil at 100°C
6 kW
Cooling power with Thermooil at 20°C
6 kW
Cooling power with Ethanol at 0°C
6 kW
Cooling power with Ethanol at -20°C
4,2 kW
Cooling power with Ethanol at -40°C
1,5 kW
Cooling power with Ethanol at -50°C
0,65 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,44 kg
CO2 equivalent
3,1 t
Circulation pump: MK pump max. delivery
79 l/min
max. delivery pressure
1,5 bar
Delivery at 0,25 bar
72 l/min
Delivery at 0,5 bar
63 l/min
Delivery at 0,75 bar
54 l/min
Delivery at 1,0 bar
43 l/min
Delivery at 1,25 bar
29 l/min
Значения приведены дословно; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' в PDF напечатаны как группы с подстроками температур.
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device rating before installation.
Source
“Power supply requirement 400V 3~ 50Hz max. current 14 A Fuse 3x16A”
Installation
The cooling water circuit needs at least 1 bar differential pressure and must not exceed 6 bar, with G1/2 male connections.
Defines the required cooling water supply pressure and connection size for proper operation.
Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation
Cooling water inlet temperature must be between 5 °C and 20 °C, and pipes must be insulated if it can drop below the dew point.
Prevents condensation and ensures rated cooling performance is available at the site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops significantly at low temperatures with ethanol, from 6 kW at 0 °C to only 0.65 kW at -50 °C.
Needed to check whether the unit can remove the process heat at the lowest required temperature.
Source
“at -50°C 0,65 kW”
Performance
The circulation pump delivers 79 l/min maximum and 43 l/min at 1.0 bar, which limits achievable flow in the external circuit.
Used to verify that the external loop pressure drop can be overcome at the required flow.