The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase feed.
Determines electrical infrastructure and whether a plug or hard wiring is needed.
Source
“Power supply requirement 400V 3~ 50Hz”
| Parameter | Value |
|---|---|
| min. filling capacity | 93 l |
| Filling capacity expansion tank | 230 l |
| Overall dimensions WxDxH ** | 4600x2600x2250 mm |
| Power supply requirement | 400V 3~ 50Hz |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | 1.0/24 |
** Please respect space requirements. See operating conditions at www.huber-online.com
| Characteristic | Value |
|---|---|
| Operating temperature range | -60...200 °C |
| Temperature stability at -10°C | 0,01 K |
| temperature set point / display | 5,7" colour Touchscreen |
| Resolution of display | 0,01 K |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| digital input | ECS ONE |
| digital output | POKO ONE |
| Alarm message | optic, acoustic, relay |
| Safety classification | III / FL |
| Heating power | 96 kW |
| Cooling power with Thermooil at 200°C | 130 kW |
| Cooling power with Thermooil at 100°C | 130 kW |
| Cooling power with Ethanol at 0°C | 130 kW |
| Cooling power with Ethanol at -20°C | 80 kW |
| Cooling power with Ethanol at -40°C | 59 kW |
| Cooling power with Ethanol at -60°C | 15 kW |
| Refrigeration machine | water-cooled, natural refrigerant |
| Refrigerant (ASHRAE, GHS) | R-1270 (A3, H220) |
| Global Warming Potential (GWP) | 0 |
| UN-number | UN 3358 |
| Circulation pump: MK pump max. delivery | 600 l/min |
| Circulation pump: MK pump max. delivery pressure | 4 bar |
| Pump connection | DN50 PN16 (DIN EN 1092-1) |
| max. permissible kin. viscosity | 50 mm²/s |
| Cooling water connection | DN65 PN16 |
| min. cooling water differential pressure | 3 bar |
| max. cooling water pressure | 6 bar |
Значения приведены дословно по тексту страницы; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' объединены с соответствующими температурными точками.
The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase feed.
Determines electrical infrastructure and whether a plug or hard wiring is needed.
“Power supply requirement 400V 3~ 50Hz”
Three-phase devices below 63 A are supplied with cable but without a plug, so the installer must provide the final connection.
Affects planning of the electrical connection and required site hardware.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cooling water must be supplied at 3–6 bar differential pressure, so the site water system must meet this range.
Insufficient or excessive cooling water pressure can prevent proper operation or damage the unit.
“min. cooling water differential pressure 3 bar”
Cooling water inlet temperature must be between 5 °C and 20 °C, and pipes must be insulated if below the dew point.
Defines the permissible cooling water supply conditions and condensation protection measures.
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 3 bar differential pressure, so actual capacity depends on site conditions.
Allows correct interpretation of published cooling capacity and prediction of derating on site.
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 3 bar differential pressure between cooling water inlet and -outlet.”
Higher cooling water temperature reduces cooling power and may require higher cooling water flow, so capacity must be reassessed.
Prevents oversizing assumptions and ensures the cooling water circuit can handle increased flow.
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides selection of cooling water quality and avoids corrosion or material damage.
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Ensures the site power quality is within the unit's permissible operating envelope.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
The refrigerant is R-1270 (A3, H220) with GWP 0 and UN number UN 3358, so refrigerant safety and transport classifications apply.
Affects safety classification, handling requirements and transport documentation.
“R-1270 (A3, H220)”
The pump connection is DN50 PN16 and the cooling water connection is DN65 PN16, so matching flanges and pipework are required.
Determines the size and standard of mating flanges and hoses for installation.
“Pump connection DN50 PN16 (DIN EN 1092-1)”
Pump connections use bore shape Y (60°) per DIN 3863 and ball socket/sleeve nut per DIN 3863/3870, so hoses must match this standard.
Ensures correct fitting of pipework and flexible tempering hoses without leaks.
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Ambient temperature must be between 5 °C and 40 °C, so the installation location must be controlled within this range.
Out-of-range ambient temperature can impair performance or cause faults.
“min. ambient temperature 5 °C”
The unit has IP20 protection, so it must be installed in a dry indoor environment protected from water ingress.
Prevents installation in wet or dusty locations that could damage the equipment.
“Degree of Protection IP20”
The unit has an electronic level indicator and automatic switch-over/capacity adaptation, so level and function should be monitored during operation.
Supports routine checks and early detection of fluid level or switching issues.
“Refrigerated Heating Circulator with water-cooled cooling machine and electronical level indicator.”