Installation
The unit requires a 400 V three-phase supply with a 63 A fuse, so the site electrical infrastructure must be sized accordingly.
Determines the electrical supply, cable and protection device needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit needs at least 1 bar differential pressure and must not exceed 6 bar, so the building water supply must be checked.
Insufficient or excessive cooling water pressure will prevent rated cooling performance or damage the circuit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is water-cooled and must be fed with cooling water between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.
Cooling water temperature directly affects achievable cooling power and condensation risk.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops sharply at low temperatures, from 11 kW at -20 °C to only 1 kW at -90 °C, so the process load at the cold end must be small.
Prevents oversizing expectations for deep-cold process cooling duty.
Source
“at -90°C 1 kW”
Installation
The pump connections are M38x1.5 male with DIN 3863 ball-socket geometry, so matching hoses and fittings must be ordered.
Ensures compatible pipework and avoids leaks or adapters at commissioning.
Source
“Pump connection M38x1,5 male”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid choice at low temperatures.
Wrong fluid viscosity can overload the pump and reduce circulation.
Source
“max. permissible kin. viscosity 50 mm²/s”
Safety
The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, so the installation area must be assessed for flammable refrigerant safety.
Flammable refrigerant classification affects room ventilation, ignition sources and safety compliance.
Source
“R-1270 (A3, H220)”
Safety
The unit has safety classification III / FL and adjustable overtemperature protection according to DIN 12876.
Confirms suitability for flammable fluids and the required protective measures.
Source
“Safety classification III / FL”
Installation
The unit weighs 893 kg and measures 954x1504x1622 mm, so floor loading and access route must be planned.
Affects rigging, floor strength and doorway clearance during installation.
Source
“Net weight 893 kg”
Operation
The ambient temperature must stay between 5 °C and 40 °C for correct operation.
Out-of-range ambient conditions can degrade performance and trip alarms.
Source
“min. ambient temperature 5 °C”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Site power quality must be verified to avoid invalid supply conditions and warranty issues.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.
Prevents commissioning delays due to missing power connection hardware.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit has a 29 l expansion tank and a minimum filling capacity of 12 l, which sets the minimum fluid charge.
Ensures enough heat transfer fluid is available for safe circulation and expansion.
Source
“min. filling capacity 12 l”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Incompatible water chemistry can cause corrosion and fouling of the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”