Product introduction and general description of the Unistat P610w
Specp. 2
Technical data according to DIN 12876 standard
p. 2table, 20 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 2,5 bar
36 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
400 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
6,5 l
Filling capacity expansion tank
26 l
Overall dimensions WxDxH **
630x704x1520 mm
Net weight
358 kg
sound pressure level +/- 4 dB(A)
62 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
16 A
Fuse
3x20A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S452288
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Unistat P610w
Order-No.
1052.0001.01
Operating temperature range
-60...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 200°C
7 kW
Cooling power with Thermooil at 100°C
7 kW
Cooling power with Ethanol at 0°C
7 kW
Cooling power with Ethanol at -20°C
6,4 kW
Cooling power with Ethanol at -40°C
2,6 kW
Cooling power with Ethanol at -60°C
0,05 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
2,55 kg
CO2 equivalent
5,5 t
Circulation pump: MK pump max. delivery
82 l/min
Circulation pump: MK pump max. delivery pressure
3 bar
Delivery at 0,5 bar
75 l/min
Delivery at 1,0 bar
68 l/min
Delivery at 1,5 bar
59 l/min
Delivery at 2,0 bar
49 l/min
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 16 A maximum current and 3x20 A fusing, so the electrical installation must be sized accordingly.
Determines the site power supply, cable sizing and breaker rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz max. current 16 A Fuse 3x20A”
Installation
The cooling water circuit must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and the pipes must be insulated if condensation can occur.
Defines the cooling water utility requirements and condensation protection needed at the installation site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Prevents corrosion or damage by selecting suitable cooling water quality and materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
The pump connection is M30x1.5 male and the cooling water connection is G1/2 male, so matching hoses and fittings must be provided.
Ensures correct mechanical interface for process and cooling water piping during installation.
Source
“Pump connection M30x1,5 male”
Installation
The unit weighs 358 kg and measures 630x704x1520 mm, so floor loading and access space must be planned.
Needed for transport, rigging and allocating sufficient installation space.
Source
“Overall dimensions WxDxH ** 630x704x1520 mm Net weight 358 kg”
Operation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be used in wet or condensing environments.
Defines the permissible operating environment and protection against ingress.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation
The cooling capacity drops sharply at low temperatures, from 2.6 kW at -40 °C to only 0.05 kW at -60 °C, so the -60 °C limit is only usable with very small loads.
Critical for selecting the unit when low-temperature operation with real heat loads is required.
Source
“at -40°C 2,6 kW at -60°C 0,05 kW”
Operation
The pump delivers 82 l/min maximum and 36 l/min at 2.5 bar, so the external circuit pressure drop must be checked against the pump curve.
Ensures the circulation pump can overcome the pressure drop of the external application.
The leakage current exceeds 3.5 mA, so a suitable protective earth and residual current device must be provided.
Important for electrical safety and selection of the RCD and earthing arrangement.
Source
“Attention: leakage current > 3,5mA”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.
Prevents operation outside the permitted supply quality and avoids invalid tolerance combinations.
Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance
The equipment is EMC class A, group 1, so it may cause radio interference in residential areas and is intended for industrial environments.
Determines EMC compatibility and suitability for the intended industrial installation site.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Three-phase devices below 63 A are supplied with a cable but without a plug, so the customer must provide the final connection.
Clarifies the scope of supply and the need to arrange a suitable plug or direct connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit has a water-cooled refrigeration machine and requires 400 l/h cooling water at 15 °C with 0 °C flow, so the cooling water supply must be sized for this flow.
Needed to verify the available cooling water flow rate at the installation site.