Unistat P905 Temperature Control System Datasheet

Contents

  1. Overview p. 1
    Product introduction and general description of Unistat P905
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 21 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
max. delivery65 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar60 l/min
Delivery at 1,0 bar54 l/min
Delivery at 1,5 bar47 l/min
Delivery at 2,0 bar39 l/min
Delivery at 2,5 bar29 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity3,2 l
Filling capacity expansion tank5,3 l
Overall dimensions WxDxH **540x654x1500 mm
Net weight272 kg
Power supply factory configured (3 Phase)460V 3~ 60Hz
max. current (3 Phase)19 A
Fuse (3 phase)3x20A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature35 °C
from Serial-No.:S573024

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat P905 Technical data according to DIN 12876

CharacteristicValue
Order-No.1054.0011.01
Operating temperature range-90...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C3,6 kW
Cooling power with Thermooil at 200°C3,6 kW
Cooling power with Thermooil at 100°C3,6 kW
Cooling power with Ethanol at 0°C3,6 kW
Cooling power with Ethanol at -20°C3,5 kW
Cooling power with Ethanol at -40°C3,5 kW
Cooling power with Ethanol at -60°C2 kW
Cooling power with Ethanol at -80°C0,4 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,5 kg
CO2-Äquvivalent5,4 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage0,75 kg
CO2-Äquvivalent 2. Stufe11,1 t
Circulation pump:MK pump

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a refrigerated heating circulator with an air-cooled refrigeration machine, suitable for both externally closed and externally open applications.

Confirms the equipment type and that it can serve open external loops, which affects loop design and accessories.

Source
“Refrigerated Heating Circulator with air-cooled cooling machine and optical level indicator.”
Selection

The 'P' model pump is intended for high-pressure-drop applications such as flow-through chemistry or semiconductor industry.

Determines whether this pump variant matches your pressure-drop requirements before ordering.

Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation

The pump connection is M30x1.5 male, so mating hoses or fittings must match this thread.

Prevents ordering incompatible hoses or adapters during installation.

Source
“Pump connection M30x1,5 male”
Installation

Piping and flexible tempering hoses must use ball sockets per DIN 3863 and sleeve nuts per DIN 3870.

Specifies the exact fitting standard required for leak-free connections.

Source
“pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation

The three-phase unit is supplied with cable but without a plug, so a suitable plug must be field-installed.

Affects electrical installation planning and bill of materials.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

External branch circuit protection according to UL 489 is required.

Mandates specific upstream protection hardware in the electrical panel.

Source
“External branch circuit protection according to UL 489 required.”
Safety

Leakage current exceeds 3.5 mA, which may require additional protective measures such as a higher-rated RCD.

Impacts residual-current device selection and electrical safety compliance.

Source
“Attention: leakage current > 3,5mA”
Operation

Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises.

Requires derating or ventilation planning in warm rooms to meet process cooling needs.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The unit must be operated within an ambient temperature range of 5 °C to 35 °C.

Defines the allowable site conditions for reliable operation.

Source
“min. ambient temperature 5 °C”
Operation

Voltage and frequency tolerances are ±5% and ±2% respectively, but not simultaneously in opposite directions.

Guides power quality assessment at the installation site.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Compliance

The equipment is classified as Class A, Group 1 for electromagnetic disturbance per EN55011.

Class A equipment may require additional EMC measures in residential or light-industrial environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The refrigerant is R-452A with a GWP of 2141 and a charge of 2.5 kg, giving a CO2 equivalent of 5.4 t.

Relevant for refrigerant handling, leak reporting thresholds, and environmental compliance.

Source
“Refrigerant quantity 2,5 kg”
Maintenance

The second-stage refrigerant is R-23 with a GWP of 14800 and a charge of 0.75 kg, giving 11.1 t CO2 equivalent.

Affects service procedures and environmental reporting for the cascade circuit.

Source
“Refrigerant quantity 2nd stage 0,75 kg”
Selection

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

Constrains fluid selection, especially at low temperatures where viscosity rises.

Source
“max. permissible kin. viscosity 50 mm²/s”