Installation
The unit requires a three-phase 400 V supply fused at 3x20 A, so the electrical installation must be sized for a 20 A per-phase load.
Determines the site power supply, cable cross-section and breaker rating needed before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The unit is rated IP20, so it must be installed in a dry indoor location protected from water and dust ingress.
Affects placement and enclosure requirements at the installation site.
Source
“Degree of Protection IP20”
Installation
Ambient temperature at the installation site must stay between 5 °C and 40 °C for correct operation.
Room conditions must be verified during site selection and ventilation planning.
Source
“max. ambient temperature 40 °C”
Installation
Cooling water must be supplied at 1 bar minimum differential pressure and no more than 6 bar, with a G3/4 male connection.
Defines the cooling water supply system and fittings required for commissioning.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water intake temperature must be between 5 °C and 20 °C, and pipes must be insulated if it can drop below the dew point.
Prevents condensation damage and ensures rated cooling performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
The stated cooling power of 10 kW is only valid at 20 °C room temperature, 15 °C cooling water and 1 bar differential pressure.
Prevents overestimating cooling capacity when site conditions differ from the reference values.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Selection
This model has no active refrigeration, so reaching 65 °C depends on adequate heat loss from the application or on water cooling.
Critical for deciding whether the unit suits the process or an external cooling source is needed.
Source
“These devices do not have an active refrigeration system.”
Operation
The pump delivers 96 l/min at zero head but only 20 l/min at 3.0 bar, so the system pressure drop must be checked against the pump curve.
Ensures the circulation loop resistance does not starve flow at the required rate.
Source
“Delivery at 3,0 bar 20 l/min”
Installation
Pump connections are M30x1.5 male with Y-shaped bore per DIN 3863, requiring matching ball-socket hoses and sleeve nuts per DIN 3870.
Specifies the correct hose and fitting standard to avoid leaks or mismatched connections.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be sourced on site.
Affects electrical commissioning planning and procurement.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
Voltage and frequency tolerances of ±10% apply only if they do not run in opposite directions, e.g. -10% voltage with +3% frequency is not allowed.
Defines acceptable grid quality for reliable operation of the electronics and pump.
Source
“-10% voltage and +3% frequency -> not allowed !”
Safety
The unit is safety class III / FL with adjustable overtemperature protection per DIN 12876.
Relevant for risk assessment and compliance with laboratory safety requirements.
Source
“Safety classification III / FL”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and treatment to avoid corrosion and fouling of the heat exchanger.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation
The expansion tank is not thermoregulated and must be locked for open bath operation.
Affects setup and safe operation depending on whether a closed or open bath is used.
Source
“Expansion tank (not thermoregulated) for closed systems, lockable for open baths.”