Product name and general description of the Unistat T330 HT unit
Specp. 2
Technical data according to DIN 12876 for the unit
p. 1table, 29 rows
Unistat T330 HT — Technical data according to DIN 12876
Characteristic
Value
Order-No.
1004.0055.01
Operating temperature range
65... 300 °C
Temperature stability at 70°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
External sensor
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
24 kW
Cooling power at 300°C
18 kW
Cooling power at 200°C
10 kW
Cooling power at 100°C
3,5 kW
Circulation pump:
MK pump
max. delivery
96 l/min
max. delivery pressure
3,5 bar
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
7 l
Filling capacity expansion tank
24 l
Overall dimensions WxDxH **
540x749x1700 mm
Power supply (3 Phase)
400V 3~ 50Hz
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
1.0/25
** Overall dimensions WxDxH — размеры указаны приблизительно и могут быть изменены.
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit covers a working temperature range of 65 to 300 °C, so it is only suitable for high-temperature closed-loop or open-bath applications, not for sub-ambient work.
Defines the process window and rules out low-temperature duties during equipment selection.
Source
“Operating temperature range
65... 300 °C”
Installation
The unit requires a 400 V three-phase 50 Hz supply, so the electrical infrastructure must be verified before ordering.
Determines the site power connection and whether a suitable three-phase feed is available.
Source
“Power supply (3 Phase)
400V 3~ 50Hz”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.
Affects installation planning and the electrical bill of materials at the site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The pump connection is an M30x1.5 male thread, so mating hoses or adapters must match this thread.
Prevents mismatch of process hoses and adapters during piping installation.
Source
“Pump connection
M30x1,5 male”
Installation
The pump port bore follows DIN 3863 shape Y (60°) and requires ball-socket fittings with DIN 3870 sleeve nuts, so standard fittings may not seal correctly.
Ensures correct fitting selection and leak-free connections on site.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN
3863, sleeve nut according to DIN 3870.”
Installation
The enclosure is IP20, so the unit must be installed in a dry indoor location protected from water and dust ingress.
Constrains the installation environment and rules out washdown or outdoor siting.
Source
“Degree of Protection
IP20”
Installation
Ambient temperature must stay between 5 °C and 40 °C, so the room climate must be checked before commissioning.
Out-of-range ambient conditions can derate performance or trip the unit.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The unit needs a minimum fill of 7 l and has a 24 l expansion tank, so the loop volume and fluid inventory must be planned accordingly.
Determines the required thermofluid quantity and expansion volume for the installation.
Source
“min. filling capacity
7 l
Filling capacity expansion tank
24 l”
Safety
Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this higher leakage.
Affects electrical safety design and the choice of residual-current protection.
Source
“Attention: leakage current > 3,5mA”
Safety
The unit is classified III / FL according to DIN 12876 and has adjustable overtemperature protection, so it can be applied in flammable-liquid duties only within that classification.
Governs whether the unit may be used with flammable fluids and what safety measures apply.
Source
“Safety classification
III / FL”
Operation
The pump delivers up to 96 l/min at 3.5 bar, so the external circuit pressure drop must be checked against this curve.
Ensures the circulation pump can overcome the resistance of the connected process loop.