Unistat TR402 Temperature Control System Datasheet

Contents

  1. Overview p. 1
    Product introduction and model identification for Unistat TR402
  2. Spec p. 2
    Technical data per DIN 12876 including performance and dimensions
  3. p. 1 table, 33 rows

Technical data according to DIN 12876

CharacteristicValue
Operating temperature range80...425 °C
Resolution of display0,01 K
Temperature stability at 70°C0,05 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracy2-point-calibration
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power at 240V3 kW
Heating power at 230V3 kW
Heating power at 220V2,7 kW
Heating power at 208V2,4 kW
Heating power at 200V2,2 kW
Circulation pump: max. delivery31 l/min
max. delivery pressure1 bar
Pump connenction flowM24x1,5
Pump connection returnM24x1,5
Cooling water connectionG1/2 male
min. filling capacity3 l
Volume of expansion2,9 l
Overall dimensions WxDxH288x332x893 mm
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
Net weight49 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current16 A
Fuse16 A
min. ambient temperature5 °C
max. ambient temperature40 °C

Order-No.: 1084.0002.01; Unistat TR402

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit covers an operating temperature range of 80 to 425 °C, so it must be matched to the process temperature window.

Defines whether the thermostat can serve the intended high-temperature process at all.

Source
“Operating temperature range 80...425 °C”
Installation

The power supply must be 200–240 V single- or two-phase at 50/60 Hz with a 16 A maximum current and 16 A fuse.

Determines site electrical supply, breaker sizing and cabling before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 16 A Fuse 16 A”
Installation

Heating power depends on the actual supply voltage, dropping from 3 kW at 230 V to 2.2 kW at 200 V.

Low-voltage sites will not achieve the nominal heating performance, affecting heat-up time calculations.

Source
“Heating power at 230V 3 kW”
Installation

The cooling water connection is a G1/2 male thread, which must be provided for on site.

Needed to plan the cooling water piping and fittings before commissioning.

Source
“Cooling water connection G1/2 male”
Installation

The unit requires at least 3 l filling capacity and has a 2.9 l expansion volume, which sets the minimum fluid charge.

Used to size the thermal fluid charge and expansion vessel behaviour.

Source
“min. filling capacity 3 l Volume of expansion 2,9 l”
Installation

The controller is connected to the machine by a cable of approximately 5 m, which constrains where the controller can be placed.

Affects layout of the operating position relative to the thermostat.

Source
“Length of cable between machine and controller approx. 5m.”
Safety

The expansion tank lid can be fixed and has connections for a nitrogen blanket limited to 0.2 bar maximum pressure.

Important for inerting high-temperature fluids and for respecting the pressure limit of the expansion vessel.

Source
“The lid has connections for a nitrogen blanket (maximum pressure 0,2 bar)”
Safety

Additional heaters with over-temperature protection are fitted for the expansion vessel and internal pipework.

Confirms built-in protection against overheating in the expansion and piping sections.

Source
“Additional heaters with over-temperature protection for expansion vessel and internal pipework”
Installation

Voltage and frequency tolerances are ±10 % only if they do not run in opposite directions, e.g. -10 % voltage with +3 % frequency is not allowed.

Site power quality must be checked against this combined tolerance rule before connection.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Compliance

The equipment is EMC class A group 1, so it is intended for industrial environments and may cause radio interference in residential areas.

Affects installation location and EMC planning for the surrounding equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

External branch circuit protection according to UL 489 is required.

The panel builder must provide a UL 489 listed breaker in the supply circuit.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

The correct plug type must be stated on the order to avoid rework at site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The unit is hydraulically closed, so no oil vapour escapes into the laboratory atmosphere.

Relevant for lab air quality and for choosing this unit for high-temperature oil operation.

Source
“hydraulically closed: No oil vapour.”
Performance

The circulation pump delivers up to 31 l/min at a maximum delivery pressure of 1 bar.

Used to check whether the external circuit pressure drop can be overcome.

Source
“max. delivery 31 l/min max. delivery pressure 1 bar”