PuriCare Open Access Stations (4' and 5' nominal widths; catalog numbers 3820001, 3820011, 3820023, 3820024, 3820000, 3820010, 3820033, 3820034) and PuriCare Procedure Stations (3', 4', 5', 6' nominal widths; ordering code 3/4/5/6 + electrical configuration). Accessories and base stands are also covered.

Technical Data

Parameter Value Condition
Nominal width 4' p.8 ↗ Open Access Station
Nominal width 5' p.8 ↗ Open Access Station
Exterior width A 48.4" (123.0 cm) p.8 ↗ Open Access Station, 4'
Exterior width A 58.4" (148.3 cm) p.8 ↗ Open Access Station, 5'
Interior width B 44.4" (123.0 cm) p.8 ↗ Open Access Station, 4'
Interior width B 48.5" (123.2 cm) p.8 ↗ Open Access Station, 5'
Electrical requirements 115 volts, 60 Hz, 12 amps p.8 ↗ Open Access Station, 3820001/3820000
Electrical requirements 100 volts, 50/60 Hz, 12 amps p.8 ↗ Open Access Station, 3820011/3820010
Electrical requirements 230 volts, 50/60 Hz, 6 amps p.8 ↗ Open Access Station, 3820024/3820034
Plug type 115 volts, 15 amps p.8 ↗ Open Access Station, 3820001/3820000
Plug type 100 volts, 15 amps p.8 ↗ Open Access Station, 3820011/3820010
Plug type Schuko p.8 ↗ Open Access Station, 3820023/3820033
Plug type British (UK) p.8 ↗ Open Access Station, 3820024/3820034
Nominal width 3' p.11 ↗ Procedure Station
Nominal width 4' p.11 ↗ Procedure Station
Nominal width 5' p.11 ↗ Procedure Station
Nominal width 6' p.11 ↗ Procedure Station
Energy requirements 12A p.11 ↗ Procedure Station, 3', 100V or 115V
Energy requirements 6A p.11 ↗ Procedure Station, 3', 230V
Energy requirements 16A p.11 ↗ Procedure Station, 4'/5'/6', 100V or 115V
Energy requirements 8A p.11 ↗ Procedure Station, 4'/5'/6', 230V
Watts 195 W p.11 ↗ Procedure Station, 3'
Watts 240 W p.11 ↗ Procedure Station, 4'
Watts 340 W p.11 ↗ Procedure Station, 5'
Watts 490 W p.11 ↗ Procedure Station, 6'

Operating Limits

  • HEPA filter efficiency: 99.99+% at 0.3 microns p.10 ↗
  • Sash opening height: 12" p.10 ↗
  • Sash opening height (Open Access): 14" (36 cm) p.6 ↗
  • Work surface liquid capacity: 1 liter p.6 ↗

Type Code Decoder

Position Token Meaning
field 1 p.11 ↗ 3 / 4 / 5 / 6 Nominal width of cabinet: 3 = 3', 4 = 4', 5 = 5', 6 = 6'
field 2 p.11 ↗ 115V, 15A Electrical configuration and plug: 115V, 15A for North America
field 2 p.11 ↗ 115V, 20A Electrical configuration and plug: 115V, 20A for North America
field 2 p.11 ↗ 230V Electrical configuration and plug: 230V for North America
field 2 p.11 ↗ 230V Schuko Electrical configuration and plug: 230V Schuko
field 2 p.11 ↗ 230V China/Australia Electrical configuration and plug: 230V for China/Australia
field 2 p.11 ↗ 230V Great Britain (UK) Electrical configuration and plug: 230V for Great Britain (UK)
field 2 p.11 ↗ 230V India Electrical configuration and plug: 230V for India

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
3820001 p.8 ↗ 3820001 Open Access Station, 4', 115 volts, 60 Hz, 12 amps
3820011 p.8 ↗ 3820011 Open Access Station, 4', 100 volts, 50/60 Hz, 12 amps
3820023 p.8 ↗ 3820023 Open Access Station, 4', Schuko
3820024 p.8 ↗ 3820024 Open Access Station, 4', 230 volts, 50/60 Hz, 6 amps, British (UK)
3820000 p.8 ↗ 3820000 Open Access Station, 5', 115 volts, 60 Hz, 12 amps
3820010 p.8 ↗ 3820010 Open Access Station, 5', 100 volts, 50/60 Hz, 12 amps
3820033 p.8 ↗ 3820033 Open Access Station, 5', Schuko
3820034 p.8 ↗ 3820034 Open Access Station, 5', 230 volts, 50/60 Hz, 6 amps, British (UK)