Induztr
Engineering Digest Document Viewer Related Models

Rev e protector downdraft powder stations user manual

Technical Reference Summary

Covered Models & Document Scope

Covers Labconco Protector Downdraft Powder Stations, models 3410000, 3410001, 3410020, 3410021, 3410022, 3410023, 3410024, 3410025. Electrical variants: 3410000/-01 at 115 VAC, 60 Hz, 1 Amp; 3410020 through -25 at 230 VAC, 50/60 Hz, 0.5 Amp. Available in 2-foot (24.0 in) and 3-foot (36.0 in) wide units. Accessory: Transition Kit #3413300 to tie two stations together.

Section-by-Section Technical Breakdown

pp. 6-7
Space, Location and Exhaust Requirements

The station must be installed on a flat, level, solid bench or work surface with a nearby electrical source. A minimum clearance of 6 inches (150 mm) above the station is required for unobstructed discharge of filtered air. The exhaust discharge is on the upper left rear side and must never be obstructed; the back may be pushed against a wall but no obstruction is allowed within 6 inches of the exhaust.

  • Minimum top clearance: 6 inches (150 mm) for filtered-air discharge.
  • Exhaust discharge located on upper left side of the back of the station.
  • Back surface may be pushed against a wall; no obstructions within 6 inches of the exhaust discharge.
  • Station must sit on a flat, level bench or work surface.
pp. 7, 25
Electrical Requirements

Model 3410000-01 requires 115 VAC, 60 Hz, 1 Amp; models 3410020-25 require 230 VAC, 50/60 Hz, 0.5 Amp. A dedicated outlet with a 10 A circuit breaker (115 V models) or 5 A breaker (230 V models) should be located near the right rear side of the station. Supplied power cords are 3 meters (10 feet) long.

  • 3410000-01: 115 VAC, 60 Hz, 1 Amp.
  • 3410020-25: 230 VAC, 50/60 Hz, 0.5 Amp.
  • Circuit breaker rating: 10 A for 115 V models, 5 A for 230 V models.
  • Power cord length: 3 m (10 ft).
  • Outlet should be as close as possible to the right rear side of the station.
pp. 8-10
Unpacking and Station Assembly

The station weighs up to 75 lbs (34 kg) and requires safe lifting by at least two persons. The work surface and work surface pan ship unassembled from the filter module. Assembly involves engaging the pan notches with the lower pins on the filter module, hooking the work surface rear flange behind the front panel, installing both side shields, and plugging the power cord into the socket at the top right rear corner.

  • Station weight: up to 75 lbs (34 kg); use two persons for lifting.
  • Tools required: flat-blade screwdriver, scissors or box knife.
  • Work surface pan engages lower pins on the filter module.
  • Work surface rear flange must go behind the front panel.
  • Side shields engage notches between front panel and filter module; side panel pins go into oversized holes on work surface edges.
  • Power cord socket is at the top right rear corner of the station.
pp. 11-13
Directional Airflow, Prefilter, HEPA Filter and Blower

Air is drawn through the perforated work surface and filter cover, then through the prefilter and HEPA filter, and exhausted out the top left rear of the filter module. The prefilter is a disposable pleated paper filter with MERV 8 rating (70-85% removal of 3.0-10 micron particles). The HEPA filter is at least 99.99% efficient at 0.3 micron. A motorized backward-curved blower pulls air through the system.

  • Prefilter: disposable pleated paper, MERV 8, removes 70-85% of particles 3.0-10 microns.
  • HEPA filter: at least 99.99% efficient for 0.3 micron particles.
  • HEPA filters only remove particulate matter; gases pass through.
  • Blower: motorized backward-curved fan for quiet, energy-efficient operation.
  • Perforated work surface and exhaust outlet must remain unobstructed for proper operation.
pp. 16-17
Using the Station

The fan switch is on the upper right rear; turning it to 'Fan' starts the blower. The Filter System Light illuminates when airflow across the filters is insufficient; first inspect or replace the prefilter, and if still lit with a clean prefilter, replace the HEPA filter. Dusting procedures include PPE, wipe-down with 70% ethanol, positioning evidence for airflow, and shutdown cleaning.

  • Fan switch located on upper right rear of station.
  • Filter System Light indicates insufficient airflow across filters.
  • If Filter System Light illuminates during use, finish dusting, then correct the problem before reuse.
  • Wipe down perforated work surface with 70% ethanol or suitable cleaner; do not use bleach on stainless steel.
  • Do not obstruct perforated work surface or exhaust air discharge.
  • After shutdown, remove perforated work surface and clean powder from pan.
pp. 18
Routine Maintenance Schedule

Weekly maintenance includes cleaning the perforated work surface and pan with 70% ethanol and operating the blower to check the filter system light. Monthly tasks add exterior cleaning with a damp cloth or LabSolutions Glass & Surface Wipes (part #1570000) and prefilter inspection/replacement. Annually, replace the HEPA filter.

  • Weekly: clean perforated work surface and pan with 70% ethanol; operate blower and note filter system light.
  • Monthly: clean exterior surfaces, especially perforated work surface; check prefilter and replace if needed.
  • Annually: replace HEPA filter.
  • LabSolutions Glass & Surface Wipes part #1570000 for exterior cleaning.
pp. 19-21
Service Operations: Prefilter, HEPA Filter and Circuit Breaker

Always disconnect power before service. Prefilter replacement: remove filter cover by pulling two latches, pull prefilter straight off the three clamps, install new prefilter with airflow arrow toward the HEPA filter. HEPA replacement: remove prefilter, remove HEPA clamps with nut driver or wrench, install replacement with gasket facing the filter frame, reinstall clamps, prefilter and cover. Reset circuit breaker by depressing the white button on the top rear until it sets.

  • Disconnect power cord before any service operation.
  • Prefilter held by three clamps; airflow arrow must point toward HEPA filter.
  • HEPA filter gasket must face the filter frame.
  • Do not touch HEPA filter media; it is very fragile.
  • Circuit breaker reset: depress white button on top rear until it sets.
  • Wear PPE: eye, breathing and skin protection.
pp. 23
Replacement Parts

Appendix A lists replacement parts: prefilter #3768903, HEPA filter #6938100, pressure switch #3897300, rocker switch #1307000, power cords for various regions, and blowers for 115 VAC and 230 VAC.

  • Prefilter: part #3768903.
  • HEPA filter: part #6938100.
  • Pressure switch: part #3897300.
  • Rocker switch: part #1307000.
  • Power cord 115 VAC (N America): #1331300; 230 VAC (UK): #1332600; 230 VAC (Schuko): #1336100; 115 VAC (China, Australia): #1332700.
  • Blower 115 VAC: #3932600; blower 230 VAC: #3932601.
pp. 24
Dimensions

Appendix B provides dimensional data: 2-foot wide units have dimension A = 24.0 in, 3-foot wide units have dimension A = 36.0 in. Overall depth is 25.8 in (655 mm), height 18.8 in (478 mm), side depth 15.1 in (384 mm), and accessory shelf height 4.4 in (112 mm).

  • Dimension A: 24.0 in for 2-foot wide units; 36.0 in for 3-foot wide units.
  • Overall depth: 25.8 in (655 mm).
  • Height: 18.8 in (478 mm).
  • Side depth: 15.1 in (384 mm).
  • Accessory shelf height: 4.4 in (112 mm).
pp. 25
Specifications

Electrical data: 3410000/-01 at 115 VAC, 60 Hz, 1 Amp, 100 W; 3410020 through -25 at 230 VAC, 50/60 Hz, 0.5 Amp, 100 W. Environmental conditions: indoor use only, maximum altitude 6562 ft (2000 m), ambient temperature 41° to 104°F (5° to 40°C), maximum relative humidity 80% up to 88°F (31°C) decreasing linearly to 50% at 104°F (40°C), supply voltage fluctuation ±10%, transient overvoltages per Installation Category II (1500 V for 115 V models, 2500 V for 230 V models), Pollution Degree 2.

  • Energy use: 100 W for all models.
  • Indoor use only; maximum altitude 6562 ft (2000 m).
  • Ambient temperature range: 41° to 104°F (5° to 40°C).
  • Maximum relative humidity: 80% up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C).
  • Main supply voltage fluctuations not to exceed ±10% of nominal.
  • Transient overvoltages per Installation Category II: 1500 V for 115 V models, 2500 V for 230 V models.
  • Pollution Degree 2 environment.
pp. 26
Accessories

Appendix D lists the Transition Kit (#3413300) that allows two separate stations to be tied together. Weight: 1 lb (0.5 kg).

  • Transition Kit part #3413300 connects two stations.
  • Weight: 1 lb (0.5 kg).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Minimum clearance of 6 inches (150 mm) above the station is required for unobstructed discharge of filtered air. p.6 ↗ Ensure no shelves, cabinets or obstructions are within 6 inches above the station. All models
Exhaust discharge is on the upper left rear side; no obstructions allowed within 6 inches of the exhaust discharge. p.7 ↗ Back may be against a wall, but top clearance must be maintained. All models
Model 3410000-01 requires 115 VAC, 60 Hz, 1 Amp; models 3410020-25 require 230 VAC, 50/60 Hz, 0.5 Amp. p.7 ↗ Select model variant according to available supply voltage. 3410000-01; 3410020-25
Circuit breaker rating: 10 A for 115 V models, 5 A for 230 V models; outlet should be near the right rear side. p.7 ↗ Provide a dedicated outlet with the specified breaker rating close to the station. 115 V models; 230 V models
Power cords supplied are 3 meters (10 feet) long. p.7 ↗ Plan outlet location within 10 ft of the station's right rear side. All models
Station weight can be up to 75 lbs (34 kg). p.8 ↗ Use at least two persons and safe lifting guidelines when moving or unpacking. All models
Prefilter is MERV 8, removing 70-85% of particles 3.0-10 microns. p.12 ↗ Prefilter protects HEPA filter from larger particulates; replace regularly. All models
HEPA filter is at least 99.99% efficient in removing particles 0.3 micron. p.13 ↗ Provides high-efficiency particulate containment for nuisance powders. All models
HEPA filters only remove particulate matter; gases pass through. p.13 ↗ Do not use with hazardous chemicals, biohazards, flammable gases or solvents. All models
Prefilter airflow arrow must point toward the HEPA filter during installation. p.19 ↗ Incorrect orientation reduces filtration performance. All models
HEPA filter must be installed with its gasket facing the filter frame. p.20 ↗ Proper gasket orientation ensures seal and filtration efficiency. All models
Circuit breaker reset: depress the white button on the top rear until it sets. p.21 ↗ Locate breaker on top rear for reset after trip. All models
Weekly maintenance: clean perforated work surface and pan with 70% ethanol; operate blower and note filter system light. p.18 ↗ Regular cleaning and function check maintain performance. All models
Monthly maintenance: clean exterior with damp cloth or LabSolutions Glass & Surface Wipes (part #1570000); check prefilter and replace if needed. p.18 ↗ Prevent dust accumulation and prefilter loading. All models
Annually replace HEPA filter. p.18 ↗ Scheduled HEPA replacement ensures continued efficiency. All models
Environmental limits: indoor use, max altitude 6562 ft (2000 m), ambient 41° to 104°F (5° to 40°C), max RH 80% up to 88°F (31°C) decreasing linearly to 50% at 104°F (40°C). p.25 ↗ Install within these conditions to ensure safe and reliable operation. All models
Supply voltage fluctuations must not exceed ±10% of nominal; transient overvoltages per Installation Category II: 1500 V for 115 V models, 2500 V for 230 V models. p.25 ↗ Verify supply quality and overvoltage protection. 115 V models; 230 V models
Energy use: 100 W for all models. p.25 ↗ Low power consumption; standard circuit capacity sufficient. All models
Dimensions: 2-foot wide units have dimension A = 24.0 in; 3-foot wide units have dimension A = 36.0 in; overall depth 25.8 in (655 mm), height 18.8 in (478 mm). p.24 ↗ Plan bench space accordingly. 2-foot and 3-foot wide units
Transition Kit #3413300 allows two separate stations to be tied together; weight 1 lb (0.5 kg). p.26 ↗ Use for connecting two stations in series. All models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Station blower won’t turn on p.22 ↗ Unit not plugged into outlet Plug the station into appropriate electrical service.
Station blower won’t turn on p.22 ↗ Check connection to the top of station. Check connection to the top of station.
Station blower won’t turn on p.22 ↗ Circuit breaker tripped Reset circuit breaker.
Filter System Light turns on p.22 ↗ Blockage of the prefilter Check the prefilter; replace if needed. Ensure that there is no obstruction of the exhaust on top of the station.
Filter System Light turns on p.22 ↗ HEPA filter is blocked If replacing prefilter does not help, replace the HEPA filter.