Rev j protector classmate laboratory fume hoods

Technical Reference Summary

Covered Models & Document Scope

Covers Labconco Protector ClassMate Laboratory Fume Hoods in 4', 5' and 6' widths, in Standard Single-Rising Sash and Combination Reduced Air Volume Sash models, with or without 2 fixtures & 1 GFCI duplex, chain drive or auto return, 115V and 230V variants, remote blower style, and VAV configurations. Catalog numbers 160404100, 160404200, 160404102, 160404202, 160504100, 160504200, 160504102, 160504202, 160604100, 160604200, 160604102, 160604202, 160405100, 160405102, 160505100, 160505102, 160605100, 160605102.

Section-by-Section Technical Breakdown

pp. 6
Location and Support Requirements

The fume hood must be located away from traffic patterns, doors, windows, fans, ventilation registers and other air-handling devices that could disrupt airflow; all room windows should be closed. Installations must be permanent and stationary on a base cabinet and chemically resistant work surface; carts, dollies and mobile benches are prohibited.

  • Locate away from traffic patterns, doors, windows, fans, ventilation registers and air-handling devices.
  • All windows in the room should be closed.
  • Do not locate on a cart, dolly or mobile bench; all installations must be permanent and stationary.
  • Supporting structure usually consists of a base cabinet and chemically resistant work surface.
pp. 6-7
Exhaust Requirements

The exhaust duct connection is designed for 12" nominal pipe (12.75" OD) to minimize static pressure loss at 60-100 fpm face velocity and to allow transport velocities of 1000-2500 fpm away from the hood. Two back-to-back hoods may be combined with a 12" x 12" x 12" dual exhaust adapter into one 12" (12.81" ID) outlet. Exhaust volume and static pressure loss are tabulated per catalog number.

  • 12" nominal pipe (12.75" OD) exhaust connection.
  • Transport velocity away from hood: 1000 to 2500 fpm.
  • Dual exhaust adapter 12" x 12" x 12" combines two 12" outlets into one 12" (12.81" ID) outlet.
  • Standard single-rising sash 48" hood at 60 fpm: 295 CFM at 0.07" H2O (18" sash opening) and 470 CFM at 0.16" H2O (28" full open).
  • Combination sash 48" hood at 60 fpm: 245 CFM at 0.04" H2O (50% open).
pp. 8, 16
Electrical Requirements

Hoods feature internal wiring for the LED light assembly and light switch; duplex outlet and switch wires are terminated at internal boxes for hook-up by a qualified electrician. All models have 115 VAC, 60 Hz, 15 Amp requirements. Remote blower style hoods are fully wired internally for 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A service.

  • All models: 115 VAC, 60 Hz, 15 Amp.
  • Remote blower style: 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A.
  • Each circuit recommended as a dedicated branch circuit.
  • Maximum combined load if wired together is sum of individual outlets plus unit rating (i.e. 2 Amps).
  • 115V wiring: Phase (Hot) black, Neutral white; 230V: Phase1 brown, Phase2 blue.
pp. 8
Service Line Requirements

All service lines should be 1/4" OD copper (brass for gas) with an easily accessible shut-off valve. Recommended operating pressure is 40 PSI with a maximum allowable pressure of 200 PSI; a pressure regulator is recommended to reduce line pressure to 40 PSI.

  • Service lines: 1/4" OD copper, brass for gas.
  • Recommended operating pressure: 40 PSI.
  • Maximum allowable pressure: 200 PSI.
  • Supply lines installed from above/under the hood and terminated at valves by a qualified contractor.
pp. 10
Unpacking and Shipping Skid Removal

Inspect the hood for transit damage and notify the carrier immediately; claims must be filed with the delivery carrier within 15 days. Leave the hood attached to its shipping skid until it is as close to its final location as possible, and move it with a floor jack or furniture dolly under the skid, never by tilting onto a hand truck.

  • Do not return goods without prior authorization from Labconco.
  • File transit damage claims directly with the freight carrier within 15 days of delivery.
  • Move hood using a suitable floor jack or furniture dolly under the skid.
  • Do not move the hood by tilting it onto a hand truck.
  • Remove side panels by unscrewing concealed screws at the top, then remove lag screws attaching hood to skid.
pp. 11
Sash Weight Release

Sash weights are secured to the back of the hood with screws for shipment. Remove the screws and verify sash cables or chains are on the pulleys or sprockets before operating the sash. Sash weights are individually matched to each hood and must not be exchanged with another unit.

  • Remove shipping screws securing sash weights.
  • Verify cables or chains are on pulleys or sprockets before operating sash.
  • Sash weights are individually matched and must not be exchanged between units.
pp. 11-12
Supporting Structure and Work Surface Installation

The hood weighs 400-700 lbs (182-318 kg); use at least six people if lifting manually. The work surface must be at least as wide as the unit and 25-7/8 inches deep, level, and supported by a 2 x 4 wood cross support at the rear after base cabinets and work surface are leveled.

  • Hood weight: 400-700 lbs (182-318 kg).
  • Manual lifting requires at least six people.
  • Work surface depth: 25-7/8 inches minimum.
  • Install 2 x 4 cross support at rear after leveling base cabinets and work surface.
  • Work surface should be chemically resistant epoxy resin, nonporous, with spillway.
pp. 13-14
Exhaust System Connection

The exhaust ductwork weight must be supported independently of the hood superstructure. The exhaust connection is 12" nominal (12.75" OD) and should be installed by a qualified HVAC contractor. A dual exhaust adapter (part 9595100) allows combining two back-to-back hoods.

  • Support ductwork weight independently of hood structure.
  • Exhaust connection: 12" nominal pipe (12.75" OD).
  • Dual exhaust adapter part number 9595100.
  • Exhaust connection to be installed by a qualified HVAC contractor.
pp. 15-16
Electrical Supply Connection

The building electrical supply should include overload protection with a switch or circuit breaker in close proximity to the equipment, marked as the disconnecting device. The hood must be grounded to the MAINS protective earthing ground using a ring terminal sized for a 10-24 machine screw on the grounding lug.

  • Provide overload protection and a marked disconnecting device near the equipment.
  • Ground to MAINS protective earthing using ring terminal for 10-24 machine screw.
  • Only MAINS ground conductors connect to the protective earthing lug.
  • 115V: black Phase, white Neutral; 230V: brown Phase1, blue Phase2.
pp. 18-20
Service Line Connection

Fixtured hoods are pre-plumbed for up to eight valves (four per side). Supply tubing must be 1/4" OD soft metal with deburred ends; push the tube approximately 3/4 inch into the fitting, tighten hand tight, then 3/4 to 1-1/4 turns more with a 7/16 inch wrench. Natural gas must only be used in the fixture pre-plumbed with brass tubing.

  • Up to eight valves (four per side) on fixtured models.
  • Supply tubing: 1/4" OD soft metal, deburred.
  • Tube insertion depth approximately 3/4 inch.
  • Tighten fitting hand tight plus at least 3/4 turn, typically 1-1/4 turns, using 7/16 inch wrench.
  • Natural gas only in brass-tubed fixture; do not use oxygen with supplied service fixtures.
pp. 20-21
Sealing and Certification

After the hood is set, ducted, wired and plumbed, seal the hood structure to the work surface with silicone sealant to prevent spilled materials from collecting under the walls. Certify the hood by measuring face velocity across the sash opening; Labconco recommends an average face velocity of 60 to 100 feet per minute. The hood is certified per ASHRAE 110-1995 with tracer gas containment less than 0.05 PPM.

  • Seal hood structure to work surface with silicone sealant.
  • Recommended average face velocity: 60 to 100 fpm.
  • Certified per ASHRAE 110-1995 with tracer gas containment < 0.05 PPM.
  • Face velocity profiles and smoke testing should be done periodically.
pp. 22-23
Performance Features

The Protector ClassMate is a by-pass high performance fume hood with clear tempered safety glass sides, back and baffle, a vertical-rising sash with 37.5" maximum visibility, 25" interior work depth and 48" interior height. LED T8 lighting provides 50,000 hour life at approximately half the power of fluorescent bulbs.

  • Vertical-rising sash operating heights: 18" or 28".
  • Interior work depth: 25"; interior height: 48".
  • Visibility up to 45" on sides and back; sash visibility 37.5" high.
  • LED T8 direct-wired bulbs: 50,000 hour life, approximately 1/2 power of fluorescent.
  • By-pass airflow design maintains containment as sash moves.
  • Factory prepared for up to 8 service fixtures.
pp. 34-36
Routine Maintenance

Chapter 6 provides a routine maintenance schedule and routine service operations for the ClassMate fume hood.

  • Routine maintenance schedule provided.
  • Routine service operations described.
pp. 37-41
Modifications and Optional Equipment

Chapter 7 describes installing work surfaces, additional service fixtures, exhaust adapter for back-to-back hoods, airflow monitors, distillation grids, ceiling enclosures, rear panels, and electrical receptacle outlets.

  • Installing work surfaces.
  • Installing additional service fixtures.
  • Installing exhaust adapter for back-to-back hoods.
  • Installing airflow monitors.
  • Installing distillation grids.
  • Installing ceiling enclosures and rear panels.
  • Installing electrical receptacle outlet.
pp. 42-44
Troubleshooting

Chapter 8 contains a table of problems, probable causes and corrective actions for the fume hood, including blower/lights failures, contamination, sash operation issues, electrical outlets and service valves.

  • Remote blower and lights won't turn on: check switch connections, control box connections, reset circuit breakers.
  • Remote blower won't turn on but lights work: inspect blower wiring and switch, replace defective blower motor.
  • Fume hood blower turns on but lights don't work: inspect lamp installation and wiring, replace lamp or ballasts, reset lamp circuit breaker.
  • Contamination outside hood: check technique, baffle slots, exhaust outlet, external airflow disruptors, re-certify at 60-100 fpm.
  • Vertical sash no longer operates smoothly: inspect cable, pulley bearing, cable/chain on pulleys, weight pulleys.
  • Combination sash no longer operates smoothly: re-install horizontal glass on tracks, re-align vertical sash frame.
  • Electrical duplex outlets no longer have power: check wire connection or replace duplex, reset circuit breakers.
  • Service valves no longer operate: inspect building supply below 40 PSI, replace valve, fix leaks.
pp. 49-54, 64
Specifications and Dimensions

Appendix B provides comprehensive dimensional diagrams for the 4, 5 and 6 foot models. Appendix C contains electrical requirements and wiring diagrams. Appendix E provides the serial number tag description and current rating code.

  • Dimensional diagrams for 4', 5' and 6' models in Appendix B.
  • Electrical requirements and wiring diagrams in Appendix C.
  • Serial number tag current rating code in Appendix E.
pp. 55-62
Accessories

Appendix D lists part numbers and descriptions for SpillStopper work surfaces, cup sinks, storage cabinets, remote blowers, ductwork and accessories, service fixtures, dual exhaust adapter, airflow monitors, distillation grids, electrical receptacle outlet kits, decorative rear panels, sash stop kits and ceiling enclosure kits.

  • SpillStopper work surfaces.
  • Cup sinks and storage cabinets.
  • Remote blowers.
  • Ductwork and accessories.
  • Service fixtures.
  • Dual exhaust adapter for back-to-back hoods (part 9595100).
  • Airflow monitors, distillation grids, electrical receptacle outlet kits.
  • Decorative rear panels, sash stop kits, ceiling enclosure kits.
pp. 73
VAV Open Area Guidelines

Appendix G provides guidelines to properly assess fume hood effective area whenever variable air volume (VAV) mechanical systems are installed, allowing the VAV system to be programmed at various sash positions.

  • Assess effective fume hood area for VAV systems.
  • Program VAV mechanical system at various sash positions.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Exhaust duct connection is 12" nominal pipe (12.75" OD). p.6 ↗ Ductwork and blower must be sized for 12" nominal; transport velocity away from hood should be 1000-2500 fpm. All models
Standard single-rising sash 48" hood at 60 fpm requires 295 CFM at 0.07" H2O (18" sash opening) and 470 CFM at 0.16" H2O (28" full open). p.7 ↗ Blower selection must account for exhaust volume and total system static pressure loss at the specified sash opening. 160404100, 160404102
Standard single-rising sash 60" hood at 60 fpm requires 375 CFM at 0.08" H2O (18" sash opening) and 595 CFM at 0.19" H2O (28" full open). p.7 ↗ Blower selection must account for exhaust volume and total system static pressure loss at the specified sash opening. 160504100, 160504102
Standard single-rising sash 72" hood at 60 fpm requires 465 CFM at 0.10" H2O (18" sash opening) and 735 CFM at 0.26" H2O (28" full open). p.7 ↗ Blower selection must account for exhaust volume and total system static pressure loss at the specified sash opening. 160604100, 160604102
Combination sash 48" hood at 60 fpm requires 245 CFM at 0.04" H2O (50% open). p.7 ↗ Combination sash models have lower exhaust volume requirements than standard models at the same face velocity. 160405100, 160405102
Combination sash 60" hood at 60 fpm requires 315 CFM at 0.05" H2O (50% open). p.7 ↗ Combination sash models have lower exhaust volume requirements than standard models at the same face velocity. 160505100, 160505102
Combination sash 72" hood at 60 fpm requires 385 CFM at 0.07" H2O (50% open). p.7 ↗ Combination sash models have lower exhaust volume requirements than standard models at the same face velocity. 160605100, 160605102
All models have 115 VAC, 60 Hz, 15 Amp electrical requirements. p.8 ↗ Provide a dedicated branch circuit rated for 15 A at 115 VAC, 60 Hz. All models
Remote blower style hoods are fully wired internally for 115 V, 50/60 Hz, 20 A or 230 V, 50/60 Hz, 10 A electrical service. p.16 ↗ Verify supply voltage and circuit rating match the hood configuration before connection. Remote blower style models
Service lines should be 1/4" OD copper (brass for gas) with recommended operating pressure of 40 PSI and maximum allowable pressure of 200 PSI. p.8 ↗ Install a pressure regulator to reduce line pressure to 40 PSI; use brass tubing for natural gas. Fixtured models
Hood weight is 400-700 lbs (182-318 kg); manual lifting requires at least six people. p.9 ↗ Use mechanical lifting equipment (floor jack, lift truck) or a minimum of six people for manual lifting. All models
Work surface must be at least as wide as the unit and 25-7/8 inches deep. p.11 ↗ Verify bench depth and width before installation; install 2 x 4 cross support at rear after leveling. All models
Recommended average face velocity through the fume hood opening is 60 to 100 feet per minute. p.20 ↗ Certify the hood by measuring face velocity across the sash opening; re-certify at 60-100 fpm if improper face velocity is suspected. All models
Hood is certified per ASHRAE 110-1995 with tracer gas containment readings less than 0.05 PPM. p.21 ↗ Field use ASHRAE testing should be performed by a certified on-site contractor. All models
Service tubing must be 1/4" OD soft metal with deburred ends; push tube approximately 3/4 inch into fitting, tighten hand tight plus at least 3/4 turn, typically 1-1/4 turns, using a 7/16 inch wrench. p.19 ↗ Follow this tightening procedure to ensure leak-free service line connections. Fixtured models
Natural gas must only be used in the service fixture pre-plumbed with brass tubing; sulfur content of gas could deteriorate standard copper supply lines. p.20 ↗ Verify gas fixture is brass-tubed before connecting natural gas. Fixtured models
Do not use oxygen with service fixtures as supplied with this hood. p.19 ↗ Contact Labconco for oxygen fixture information if oxygen service is required. Fixtured models
Dual exhaust adapter (part 9595100) combines two 12" nominal exhaust outlets into one 12" (12.81" ID) nominal exhaust outlet. p.14 ↗ Use adapter when combining air from two back-to-back hoods; verify blower sizing for combined exhaust volume. Back-to-back hood installations
LED T8 direct-wired bulbs provide 50,000 hour life and utilize approximately 1/2 the power of fluorescent bulbs. p.23 ↗ LED lighting reduces energy consumption and maintenance frequency. All models
Vertical-rising sash may be raised from closed to 18" or 28" operating height; maximum sash visibility is 37.5" high. p.22 ↗ Exhaust volume requirements differ between 18" (60% open) and 28" (100% full open) sash positions. Standard single-rising sash models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Remote blower and lights won’t turn on p.46 ↗ Wires not connected at junction boxes or switches. Check connection of switches.
Remote blower and lights won’t turn on p.46 ↗ Wires not connected at junction boxes or switches. Check connection to control box on top of unit.
Remote blower and lights won’t turn on p.46 ↗ Circuit breakers tripped in building electrical supply. Reset circuit breakers.
Remote Blower won’t turn on but lights work p.46 ↗ Blower wiring is disconnected. Inspect blower wiring, and switch.
Remote Blower won’t turn on but lights work p.46 ↗ Blower motor is defective. Replace blower motor.
Fume Hood blower turns on but lights don’t work p.46 ↗ Lamp not installed correctly. Inspect lamp installation.
Fume Hood blower turns on but lights don’t work p.46 ↗ Lamp is defective. Replace lamp.
Fume Hood blower turns on but lights don’t work p.46 ↗ Lamp circuit breaker in building is tripped. Reset the lamp circuit breaker.
Fume Hood blower turns on but lights don’t work p.46 ↗ Lamp wiring is disconnected. Inspect lamp wiring.
Fume Hood blower turns on but lights don’t work p.46 ↗ Defective lamp ballasts. Replace lamp ballasts.
Contamination outside of laboratory fume hood p.47 ↗ Improper technique or procedure for the fume hood. See “Certifying the Hood Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix F)
Contamination outside of laboratory fume hood p.47 ↗ Restriction of the baffle air slots or – blockage of the exhaust outlet. Ensure that all baffle air slots, and the exhaust outlet are unobstructed.
Contamination outside of laboratory fume hood p.47 ↗ External factors are disrupting the fume hood airflow patterns or acting as a source of contamination. See “Location Requirements” Chapter 2, “Certifying the Hood” Chapter 3, and “Safety Precaution” Chapter 4 sections of this manual. (Ref. Appendix F)
Contamination outside of laboratory fume hood p.47 ↗ Fume Hood has improper face velocity. Have fume hood re-certified at 60-100 fpm and check remote blower exhaust system.
Fume Hood Vertical Sash no longer operates smoothly p.47 ↗ Cable is frayed or plastic protection is damaged. Inspect cable and replace cable if worn or damaged; otherwise injury could result.
Fume Hood Vertical Sash no longer operates smoothly p.47 ↗ Pulley bearing is damaged. Replace pulley, bearing or add grease.
Fume Hood Vertical Sash no longer operates smoothly p.47 ↗ Cable or chain has slipped off the pulleys or sprockets. Re-install, cable or chain to be replaced if damaged.
Fume Hood Vertical Sash no longer operates smoothly p.47 ↗ Weight has broken pulleys or sprockets. Replace weight pulleys or sprockets.
Fume Hood Combination Sash no longer operates smoothly p.47 ↗ Horizontal glass panels have come off the tracks. Re-install horizontal glass on tracks and secure with clips.
Fume Hood Combination Sash no longer operates smoothly p.47 ↗ Vertical sash frame is crooked. Place horizontal glass symmetrically and pull sash down to air foil to re-align. Re-align the chain as needed.
Electrical duplex outlets no longer have power p.48 ↗ Wires not connected or faulty duplex. Check wire connection or replace duplex.
Electrical duplex outlets no longer have power p.48 ↗ Circuit breakers tripped in building electrical supply. Reset circuit breakers.
Service valves no longer operate p.48 ↗ Faulty building supply. Inspect building supply shut off valves and appropriate pressures below the recommended 40 PSI.
Service valves no longer operate p.48 ↗ Valve no longer operates. Replace valve and check for leaks.
Service valves no longer operate p.48 ↗ Supply line or outlet line has leaks. Inspect line for leaks and fix any leaking plumbing connections.