Rev g paramount ductless enclosures

Technical Reference Summary

Covered Models & Document Scope

Covers the Labconco Paramount® Ductless Enclosure series (ductless fume hood), manual part #6968600 Rev. G. Models: 2', 3', 4', 5', 6' widths in 100-115V and 208-230V variants, including standard 22.7"/9.4" sash, 32"/12" sash, 32" x 360° sash, and 35"/12" sash configurations (e.g. 6963200, 6963220, 6963300-6963302, 6963320-6963322, 6963400-6963403, 6963420-6963423, 6963501/6963503/6963521/6963523, 6963601/6963603/6963621/6963623). Ships without filters; seven filter types (HEPA and carbon) ordered separately.

Section-by-Section Technical Breakdown

pp. 9-12
Site Prerequisites: Support, Location, Exhaust and Electrical

Defines the supporting structure, location/air-current avoidance, exhaust/blower arrangement, electrical supply and space requirements before installation. The enclosure uses a 95% energy-efficient ECM impeller and normally recirculates filtered air into the room; outside exhaust requires the accessory Canopy Exhaust Connection and remote blower.

  • Supporting structure: base cabinet or stand plus chemically resistant, thick rigid work surface; floor-mounted bench or mobile bench with locking casters acceptable.
  • Avoid high-traffic areas, HVAC diffusers, fans, radiators, doorways and operable windows.
  • All models shipped to operate at 75-80 fpm; can be reset by Safety Officer/certification technician via password-protected menu.
  • If connected to a house exhaust system, an adjustable damper or valve must be installed to control airflow.
  • Standard duplex receptacles should be nearby; iris pass-through ports allow cords through the back.
  • No provision for remote blower control; a wall-mounted switch receptacle may be wired by a qualified electrician.
pp. 11
Model Airflow, Exhaust Volume and Noise Data

Chart of face velocity (60, 80, 100 fpm) versus exhaust volume (CFM) and noise pressure (dbA) for every model, with enclosure width, internal working height, sash opening, exterior depth and height.

  • 2' models: 85/115/145 CFM at 60/80/100 fpm; 50/54/59 dbA.
  • 3' models (22.7" sash): 130/175/220 CFM; 46/52/56 dbA.
  • 3' models (32" sash): 170/230/285 CFM; 52/57/62 dbA.
  • 4' models (22.7" sash): 175/235/295 CFM; 53/57/62 dbA.
  • 4' models (32" sash): 230/305/380 CFM; 57/62/67 dbA.
  • 5' models: 220/295/370 CFM; 57/62/67 dbA.
  • 6' models: 350/465/580 CFM; 57/61/66 dbA.
pp. 13-15
Unpacking, Work Surface and Enclosure Mounting

Describes moving the enclosure on its skid, inspecting for transit damage, and securing it to a level work surface using four mounting screws. The front of the enclosure must align within 0.3" of the work surface front.

  • Enclosure weight 125-350 lbs (55-152 kg); use mechanical lift truck or floor jack; do not lift by front air foil.
  • Do not move by tilting onto a hand truck.
  • Work surface must be at least as wide as the enclosure, smooth, rigid, non-porous and chemical resistant.
  • Insert four mounting screws leaving minimum 1/8" clearance under the head; stainless surfaces use #10-24 x 3/8" hex washer head screw (Labconco P/N 1885806).
  • Slide rear flange and front air foil flanges under screw heads and tighten.
  • Optional RTV silicone bead seals enclosure to work surface.
pp. 16-18
Carbon and HEPA Filter Installation

Lists the seven filter types and part numbers, organic vapor capacities, stacking rules and the number of filters per model size. Filters are always loaded gasket side up and stacked two high per compartment.

  • HEPA 6938100: 15" x 18" x 1.5", 99.99% efficient on 0.3 micron.
  • Carbon filters: 6938200 organic vapor 7.5 lbs; 6938201 acid-sulfur 10 lbs; 6938202 ammonia-amine 11 lbs; 6938203 formaldehyde-formalin 10 lbs; 6938204 mixed bed 10 lbs; 6938205 radioisotope 10 lbs.
  • Impregnated carbon filters have 25% organic vapor capacity; mixed bed has 50%.
  • Any combination of two filter types may be used; in clean rooms place HEPA above carbon.
  • 2' = 1 compartment/2 filters/15 lbs; 3' = 2/4/30 lbs; 4' = 3/6/45 lbs; 5' = 3/6/45 lbs; 6' = 4/8/60 lbs.
  • Push stacked filters to the rear stop and close gas-spring cylinders to compress gaskets.
pp. 18-19
Electrical Connection, Filter Timers and Vapor Sensor Warm-Up

Covers connecting the 115V power cord to the IEC inlet, installing a plug on 230V units, setting filter life/check timers, and the initial 10-30 minute Safety-First Vapor Sensor warm-up.

  • 115V units use supplied cord to IEC inlet on rear; 230V units ship without a plug—install per local codes.
  • Set Filter Life Timers in Filter Setup menus using the Paramount Chemical Guide, Appendix E and/or Labconco Specialist.
  • Vapor sensor has a 5V, 0.28W integrated heater; warm up 10-30 minutes with blower running.
  • With new filters, Organic Sensor should read 1300-1500 mV after warm-up.
  • Change all filters when Organic Sensor rises 400-500 mV and display shows 'Replace Filters.'
pp. 19-20
Face Velocity Calibration and Validation

Explains factory preset 75-80 fpm, recalibration via the password-protected Calibration menu, alarm setpoint adjustment, and the measurement method for average face velocity.

  • Factory preset face velocity 75-80 fpm; 100 fpm available for special requirements.
  • Airflow alarm preset at 60 fpm; adjust to 10-20 fpm below actual face velocity.
  • Measure with calibrated thermal anemometer at 5" directly below sash handle; readings 6" from ends and every 12".
  • Recalibrate at altitudes above or below 900 feet.
  • Smart-Flow automatically compensates motor speed for filter loading, barometric pressure and temperature.
  • ASHRAE 110-1995 containment rating <0.05 ppm at 4 lpm.
pp. 20
Connecting to an Exhaust System

Describes optional outside exhaust using the Canopy Exhaust Connection accepting 6" ductwork, with independent support of ductwork and HVAC contractor installation.

  • Normal operation recirculates filtered air into the room; canopy not required.
  • Canopy Exhaust Connection accepts 6" diameter ductwork.
  • Exhaust ductwork weight must be supported independently of the enclosure superstructure.
  • Installation by qualified HVAC contractor; consult Labconco for blower sizing.
  • Duct material must match enclosure, procedures and chemical applications.
pp. 21-26
Performance Features and Safety Systems

Describes the containment air foils, Safety-First Vapor Sensor detection range and chemical sensitivity table, Smart-Flow airflow monitor, ECM impeller, alarms and LCD display.

  • Safety-First Vapor Sensor optimal detection range 1-10 ppm; primary and secondary sensors.
  • Detects organic solvents, smoke particulates, ammonia, formaldehyde and hydrogen sulfide; does NOT detect mineral acids (HCl, nitric, sulfuric).
  • Early warning at 5-10% remaining filter life: intermittent beep; final warning: constant alarm.
  • ECM impeller 95% energy efficient, located after filters; all Paramounts ≤62 dbA at 80 fpm, most ≤59 dbA.
  • LCD 3" x 1.5" displays filter status, face velocity and alarms.
  • Smart-Flow uses pitot tube and pressure sensor on control board for constant velocity.
pp. 39-50
Routine Maintenance and Filter Replacement

Provides the routine maintenance schedule, criteria for replacing carbon and HEPA filters, HEPA leak test, face velocity change, and replacement procedures for lamp, impeller, main control board and organic sensor board.

  • Carbon filter replacement determined by vapor sensor alarm, elapsed timer, or chemical guide calculation.
  • HEPA filter replaced when clogged/low airflow or per schedule.
  • HEPA filter leak test procedure included.
  • Fluorescent light, motorized impeller, main control board and organic sensor board replacement procedures.
  • Re-certification and initial certification guidance.
pp. 51-59
Accessories and Ordering

Lists work surfaces, stands, storage cabinets, exhaust transitions, remote blowers, exhaust dampers, utility shelf kits, syringe pump kit, detector tubes and sash closures.

  • Work surfaces with deck-mounted valves, gooseneck and cupsink.
  • Base stands, accessory shelves, seismic supports, hydraulic lift base stands.
  • Storage cabinets, exhaust transitions, remote blowers, exhaust dampers.
  • Utility shelf kits and holders, syringe pump kit and detector tubes, sash closures.
pp. 60-62
Troubleshooting

Table of problems, causes and corrective actions for filter alarms, airflow alarms, blower/lamp failures and airflow monitor malfunctions.

  • Covers 'Replace Filters' continuous and intermittent alarms.
  • Covers 'Check Filters' intermittent alarm.
  • Covers blower will not operate, low face velocity, lights do not work.
  • Covers airflow monitor malfunction (no audible alarm, constant alarm, low airflow).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
All models are shipped to operate at 75-80 fpm face velocity; can be reset via password-protected menu. p.10 ↗ Default commissioning setpoint; verify with thermal anemometer before use. whole series
Airflow alarm preset at 60 fpm; adjust to 10-20 fpm below actual face velocity. p.19 ↗ Set alarm after calibration to avoid nuisance alarms while maintaining protection. whole series
Face velocity measurement: calibrated thermal anemometer, 5" directly below sash handle, readings 6" from ends and every 12". p.19 ↗ Standardized validation method for certification. whole series
Recalibration required at altitudes above or below 900 feet. p.19 ↗ Site altitude affects pressure sensor and face velocity; recalibrate after installation. whole series
Organic Sensor should read 1300-1500 mV with new filters after 10-30 minute warm-up with blower running. p.18 ↗ Baseline for sensor health; replace filters when reading rises 400-500 mV and 'Replace Filters' appears. whole series
Safety-First Vapor Sensor optimal detection range 1-10 ppm; does not detect mineral acids (HCl, nitric, sulfuric). p.24 ↗ Use alternative detection (mineral acid sensor or sampling tubes) for acid applications. whole series
Impregnated carbon filters have 25% organic vapor capacity; mixed bed has 50%. p.17 ↗ For solvent+acid service, use one organic vapor filter plus one impregnated acid filter. filter selection
Filter counts: 2'=2 filters/15 lbs; 3'=4/30 lbs; 4'=6/45 lbs; 5'=6/45 lbs; 6'=8/60 lbs total organic vapor weight. p.17 ↗ Determines spare filter inventory and filter life calculation. model sizes
HEPA filter 6938100 is 99.99% efficient on 0.3 micron; dimensions 15" x 18" x 1.5". p.17 ↗ Particulate containment specification for powder work. HEPA option
Enclosure weight 125-350 lbs (55-152 kg); do not lift by front air foil. p.13 ↗ Rigging and installation planning; use mechanical lift or floor jack. whole series
Front of enclosure must align within 0.3" of front of work surface; mounting screws leave minimum 1/8" clearance under head. p.14 ↗ Ensures proper air foil function and secure mounting. whole series
Stainless work surfaces use #10-24 x 3/8" hex washer head screw, Labconco P/N 1885806. p.15 ↗ Correct fastener for mounting enclosure to stainless surface. stainless work surfaces
Canopy Exhaust Connection accepts 6" diameter ductwork; ductwork weight must be supported independently. p.20 ↗ Outside exhaust installation requires separate duct supports and HVAC contractor. exhaust option
ECM impeller is 95% energy efficient and located after filters; all Paramounts ≤62 dbA at 80 fpm, most ≤59 dbA. p.25 ↗ Noise planning and intrinsically safe design (contaminated air filtered before impeller). whole series
ASHRAE 110-1995 containment rating <0.05 ppm at 4 lpm. p.20 ↗ Validated containment performance for operator protection. whole series
Vapor sensor warm-up: 5V, 0.28W integrated heater; 10-30 minutes with blower activated. p.18 ↗ Allow warm-up before relying on sensor readings after shipment or filter change. whole series
Filter loading: always gasket side up; stack two high per compartment; in clean rooms HEPA above carbon. p.17 ↗ Correct orientation prevents bypass and contamination. filter installation
230V units ship without a plug; install appropriate plug per local codes. p.18 ↗ Electrical contractor must supply plug for 208-230V models. 208-230V models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
"Replace Filters" continuous alarm occurs p.66 ↗ Improper final time programmed and display reads "Replace Filters-t", (t = time) See Chapters 3 and 5 to properly set the Filter Final Life Timer. Chapter 8
Organic sensor alarms and display reads "Replace Filters-s", (s = organic sensor) p.66 ↗ Organic sensor has detected filter breakthrough. Replace filters immediately. A spare set of filters should always be on hand. Consult Chapters 3, 4, and 5, the Labconco Specialist, and the Chemical Guide for estimated filter life based on specific chemical use. Always limit chemical use, cover up chemicals when not in use and turn off blower to preserve filter life. Chapter 8
"Mute" will not stay operational p.66 ↗ The "Mute" alarm can only be off for 5 minutes and must be pushed every 5 minutes until the filters have been replaced. See Chapters 3, 5 & 6. Chapter 8
"Check Filters" intermittent alarm occurs frequently p.66 ↗ Improper check time programmed and display reads "Check Filters-t", (t = time) See Chapters 3 and 5 to properly set the Filter Check Life Timer. Chapter 8
Replace filters intermittent alarm occurs p.67 ↗ Organic sensor alarms intermittently and display reads "Replace Filters-s", s = organic sensor; organic sensor has detected filter breakthrough and only 5-10% of filter life remains. Replace filters with a spare set of filters stored. Consult Chapters 3, 4, and 5, Filter Specialist, and Chemical Guide for filter life based on specific chemical use. Always limit chemical use, cover up chemicals and turn off blower when not in use to preserve life. Chapter 8
Replace filters intermittent alarm occurs p.67 ↗ Carbon filter loaded with chemicals. Replace carbon filter. Chapter 8
Contaminants outside of enclosure p.67 ↗ Improper user techniques for the enclosure. See "Validate the Enclosure" Chapter 3 and "Safety Precautions" Chapter 4 sections in the manual. Chapter 8
Contaminants outside of enclosure p.67 ↗ Restriction of the baffle air slots or blockage of the exhaust outlet. Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. Chapter 8
Contaminants outside of enclosure p.67 ↗ External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. See "Location Requirements" Chapter 2, "Validate the Enclosure" Chapter 3, and "Safety Precautions" Chapter 4 sections of this manual. Chapter 8
Contaminants outside of enclosure p.67 ↗ Enclosure has improper face velocity. Have enclosure certified per Chapter 6 and check exhaust system. Check carbon filters and replace if necessary. Enclosure should have an average face velocity of 60-100 fpm. Chapter 8
Blower will not operate p.67 ↗ Unit not plugged into outlet. Plug the enclosure into appropriate electrical service. Chapter 8
Blower will not operate p.67 ↗ Circuit breaker(s) or Ground Fault Interrupter. Reset circuit breaker. Chapter 8
Blower will not operate p.67 ↗ Blower wiring is disconnected. Inspect blower wiring. Chapter 8
Blower will not operate p.67 ↗ Blower switch is defective. Replace membrane switch on control panel. Chapter 8
Blower will not operate p.67 ↗ Motorized impeller is defective. Replace motorized impeller. See Chapter 6. Chapter 8
Low face velocity p.67 ↗ Enclosure sash not closed. Close sash to the lowest position. Chapter 8
Low face velocity p.67 ↗ HEPA filter clogged. Replace HEPA filter. Chapter 8
Blower and lights will not operate p.67 ↗ Unit not plugged into outlet. Plug enclosure into appropriate electrical service. Chapter 8
Blower and lights will not operate p.67 ↗ Circuit breaker(s) tripped. Reset or replace circuit breaker. Chapter 8
Lights do not work p.68 ↗ Lamp not installed properly. Inspect lamp installation. Chapter 8
Lights do not work p.68 ↗ Lamp wiring disconnected. Inspect lamp wiring. Chapter 8
Lights do not work p.68 ↗ Defective lamp. Replace lamp. Chapter 8
Lights do not work p.68 ↗ Light switch is defective. Replace membrane switch on control panel. Chapter 8
Lights do not work p.68 ↗ Defective electronic ballast. Replace ballast. Chapter 8
Airflow monitor malfunction - No audible alarm. p.68 ↗ Alarm has been temporarily silenced using "mute" button. Wrong alarm set point. Airflow monitor alarm is preset at 60 fpm. Repeat calibration steps outlined in this manual in Chapter 3, Chapter 5 and Chapter 6. Chapter 8
Airflow monitor malfunction - Constant audible alarm. p.68 ↗ Check airflow and calibration of airflow monitor. See Chapter 5 & 6. Chapter 8
Airflow monitor malfunction - Continuous alarm. p.68 ↗ Check the face velocity of the enclosure as the airflow of the system may have changed. If incorrect, follow the calibration steps in Chapters 3, 5 and 6. The carbon filter pre-filter or HEPA filter may have become loaded. Check the face velocity of the enclosure as the airflow of the system may have changed. If incorrect, follow the calibration steps in Chapters 3, 5 and 6. The carbon filter pre-filter or HEPA filter may have become loaded. Chapter 8
Airflow monitor malfunction - Low airflow. p.68 ↗ Check airflow and reset airflow monitor. See Chapter 5 & 6. Look for filter blockage or paper debris. Chapter 8