Technical Reference Summary

Covered Models & Document Scope

Labconco TXE – Tall XPert® Enclosures, Part #3892500 Rev. G. Covers 2 FT, 3 FT, 4 FT, 5 FT and 6 FT enclosure widths; 115V and 230V electrical models; integral motorized impeller models with recirculating HEPA-filtered exhaust or optional Canopy Exhaust Connection for outside exhaust via remote blower. 5 FT and 6 FT models use two HEPA filters.

Section-by-Section Technical Breakdown

pp. 4, 8, 12
Exhaust and Blower Requirements

The TXE uses an integral motorized impeller to draw room air past the operator and through a HEPA filter, discharging filtered air out the top. Exhaust volume, noise and static pressure data are given at 80 and 100 fpm face velocity. For outside exhaust, a dedicated blower or house exhaust system is required, and an adjustable damper must be installed when connected to a house exhaust system.

  • 2' TXE at 80 fpm: 115 CFM, 0.48"–0.57" w.g. pressure loss with HEPA filter, 50–56 dB(A), max equivalent resistance 200 ft, max external static pressure 0.18".
  • 2' TXE at 100 fpm: 145 CFM, 0.62"–0.72" w.g. pressure loss, 53–58 dB(A), max equivalent resistance 80 ft, max external static pressure 0.13".
  • Exhaust connection accepts 6" diameter ductwork and ships uninstalled.
  • Adjustable damper (or valve) must be installed when connected directly to a house exhaust system, including systems controlling multiple filtered enclosures.
pp. 10, 11, 14, 15
Electrical Requirements and Connection

115V models connect via the supplied power cord to the IEC inlet on the back of the enclosure. Remote blower wiring is made in the electrical junction box behind the front panel and is switched by an electrical relay via the blower switch. 230V models ship without a plug and require an appropriate plug per local codes.

  • Maximum circuit load for the electrical relay is 6 amps.
  • Only remote blower wiring is connected inside the electrical junction box; lights and airflow monitor are powered from a nearby standard duplex receptacle.
  • Grounding connection shall not be made to the terminal box cover.
  • Building supply for the remote blower must include overload protection (switch or circuit breaker) marked as the disconnecting device; consult NEC-2002.
  • At 50 Hz operation, blower maximum airflow is reduced by 17%.
  • Do not use any detachable power cord not adequately rated for the unit.
pp. 9, 10, 11, 12, 14
Installation, Supporting Structure and HEPA Filter Installation

Each TXE weighs approximately 213 lbs (79 kg) and may be top-heavy; it must be kept vertical and never lifted by the front doors. The enclosure is fastened to a supporting structure/stand with mounting brackets and screws. HEPA filters ship installed with the gasket on the downstream side and are leak-checked at Labconco; a second leak check is recommended before use and at least annually.

  • Secure glass doors with tape whenever moving or lifting the enclosure.
  • Do not move the enclosure by tilting it onto a hand truck; use a flat low dolly under the shipping skid or a floor jack.
  • Supporting structure must safely support the combined weight of the enclosure and related equipment.
  • Exhaust ductwork weight must be supported independently of the enclosure superstructure.
  • All seams upstream of the HEPA filter are jacketed by negative pressure.
pp. 12, 16
Setting the Face Velocity with the Speed Control

The speed control behind the front panel sets the face velocity, which should be adjusted between 80 and 100 fpm. Face velocity is measured with an anemometer as an average of several readings in the largest opening with the glass doors closed. Turning the speed control counterclockwise increases face velocity; clockwise decreases it.

  • Adjust face velocity to 80–100 fpm; containment is maximized within this range.
  • Lowest appropriate face velocity gives the quietest operation.
  • Use a small Phillips screwdriver to adjust the speed control.
  • Face velocity profiles and smoke testing should be performed frequently per the organization's quality system.
pp. 13, 15, 16, 17, 19, 20
High Performance Features

The TXE contains powders and particulates using a zoned rear perforated baffle that produces horizontal laminar airflow, with most contaminants removed on the first pass. The plenum and HEPA filter are jacketed by negative pressure so leaks are recaptured and refiltered. A true bag-in/bag-out HEPA filter disposal system protects the worker during filter changes.

  • HEPA filter rated to remove 99.99% of particles 0.3 micron in size.
  • Zoned rear perforated baffle has increasingly more open area at the bottom to form laminar airflow.
  • Vibration-isolated motorized impeller eliminates transfer of vibration to the work surface.
  • Filter system pressure gauge indicates total system pressure across the HEPA filter and pressurized plenum to help predict filter loading.
  • Optional Guardian Airflow Monitor or Guardian 1000 Digital Airflow Monitor provides audio/visual low-airflow alarm; Guardian 1000 also displays face velocity, provides RS232 output, night setback mode and auxiliary relay ports.
pp. 21, 22, 16
Routine Maintenance Schedule

Maintenance consists primarily of monitoring airflow and changing filters. Daily/weekly tasks include cleaning interior and exterior surfaces and checking airflow with visible smoke. Monthly tasks include measuring actual face velocity with a calibrated thermal anemometer; annually, replace fluorescent lamps and have the enclosure validated by a qualified certification technician.

  • Daily/weekly: wipe interior with disinfectant or neutralizing cleaner; clean exterior with damp cloth; operate exhaust and note airflow with visible smoke.
  • Monthly or more often: determine actual face velocity through sash openings with calibrated thermal anemometer or approved apparatus.
  • Annually: replace fluorescent lamps and have enclosure validated by a qualified certification technician.
  • Recertify whenever the enclosure is serviced or relocated, and at least annually.
pp. 22, 23, 24
HEPA Filter Replacement and Bag-In/Bag-Out Procedure

HEPA filters typically last two to five years and are replaced when face velocity in the largest opening drops below 80–100 fpm with the speed control at full speed, or when the filter fails the leak test. The true bag-in/bag-out procedure uses gloves and sealed bags to remove the contaminated filter safely. Filter clamp bolts are tightened until the HEPA filter gasket is compressed 50%.

  • Loosen/tighten the two filter clamp bolts with a 9/16" socket.
  • Tighten filter clamp bolts uniformly until the HEPA filter gasket is compressed 50%.
  • Create two seals (heat sealed or tie wraps) between the access port and old filter, then cut the bag between the seals.
  • Do not remove the black trim seal on the leading edge of the access port flange.
  • Do not contact the blower wheel while still in motion.
  • Wear appropriate personal protective equipment during HEPA filter changes.
pp. 25, 26
HEPA Filter Leak Test

After installing a new HEPA filter, a qualified technician performs a leak test of the filter, housing and mounting frames using an aerosol photometer and Laskin nozzle generator. The filter passes at 0.01% penetration or better. Scanning is done on the downstream exhaust side with the sampling port not more than 1 inch from the filter media at a traverse rate not exceeding 2 inches per second.

  • Aerosol photometer ATI model 2D, 2E, 2G or equivalent; Laskin nozzle generator with compressed air at 10 ± 1 psig.
  • Photometer reference values at 90 fpm: 52 (2'), 34 (3'), 26 (4'), 31 (5'), 26 (6').
  • Airflows at 90 fpm: 2' = 130 CFM, 3' = 200 CFM, 4' = 265 CFM, 5' = 215 CFM, 6' = 260 CFM; 5' and 6' use two filters.
  • Allow photometer to operate minimum 5 minutes and enclosure minimum 5 minutes before testing.
  • Aerosol penetration shall not exceed 0.01 percent measured by the photometer.
  • Reference IES-RP-CC001.3 for leak testing and photometer scanning.
pp. 19, 30, 41
Guardian Digital 1000 Airflow Monitor

The Guardian 1000 Digital Airflow Monitor continuously monitors airflow and provides an audio/visual alarm for low airflow. It displays face velocity, provides RS232 output, night setback mode and auxiliary relay ports. Programming and calibration of the alarm are described in Chapter 6 and the supplied Digital 1000 Monitor Manual.

  • Alarm can be temporarily silenced using the SILENCE ALARM or enter buttons.
  • Audible disable requires a continuously present alarm condition; fluctuating flow near the set point automatically resets it.
  • Airway to the airflow monitor sensor must be kept free of insects, dust and debris.
  • Calibration steps are given in Chapter 6 or the supplied Digital 1000 Monitor Manual.
pp. 15, 19, 34, 40
Motorized Impeller and Speed Control Replacement

The motorized impeller and speed control are serviceable components described in Chapter 6. The impeller is vibration-isolated to prevent transfer of vibration to the work surface. Replacement is required when the impeller or blower is defective or when face velocity cannot be maintained.

  • Replace motorized impeller or blower if defective (see Chapter 6).
  • Speed control is located behind the front panel and is used by the certifier to validate and adjust inflow velocity.
  • Do not contact the blower wheel while still in motion.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
2' TXE at 80 fpm face velocity: exhaust volume 115 CFM, pressure loss with HEPA filter 0.48"–0.57" w.g., noise 50–56 dB(A), max equivalent resistance 200 ft, max external static pressure 0.18". p.8 ↗ Use these values to size a remote blower and duct system for outside exhaust at 80 fpm. 2 FT TXE
2' TXE at 100 fpm face velocity: exhaust volume 145 CFM, pressure loss with HEPA filter 0.62"–0.72" w.g., noise 53–58 dB(A), max equivalent resistance 80 ft, max external static pressure 0.13". p.8 ↗ Higher face velocity increases required blower static pressure and reduces allowable equivalent duct resistance. 2 FT TXE
Maximum circuit load for the electrical relay is 6 amps. p.14 ↗ Remote blower connected through the TXE relay must not draw more than 6 A. All models with remote blower wiring
At 50 Hz operation, blower maximum airflow is reduced by 17%. p.15 ↗ Account for reduced airflow when selecting blower and setting face velocity on 230V/50 Hz installations. 230V models
Face velocity should be adjusted between 80 and 100 fpm; containment is maximized within this range. p.16 ↗ Set speed control to achieve 80–100 fpm average in the largest opening with glass doors closed. All TXE models
For weighing procedures the enclosure should have an average face velocity of 60–100 fpm; for biohazardous procedures 75–105 fpm. p.44 ↗ Select and verify face velocity according to the application type. All TXE models
Each TXE model weighs approximately 213 lbs (79 kg). p.9 ↗ Verify supporting structure and lifting equipment can handle the weight; keep enclosure vertical and never lift by front doors. All TXE models
HEPA filter is rated at least 99.99% efficient for 0.3-micron particles. p.5 ↗ Suitable for containment of toxic particulates and powders, not for gases or vapors without carbon filtration. All TXE models
HEPA filter leak test acceptance: aerosol penetration shall not exceed 0.01 percent. p.26 ↗ Filter, housing and mounting frames must pass this test after installation and at least annually. All TXE models
Leak test airflows at 90 fpm: 2' = 130 CFM, 3' = 200 CFM, 4' = 265 CFM, 5' = 215 CFM, 6' = 260 CFM; 5' and 6' use two filters. p.26 ↗ Set enclosure speed control to these airflows during leak testing. 2 FT, 3 FT, 4 FT, 5 FT, 6 FT models
Photometer reference values at 90 fpm: 52 (2'), 34 (3'), 26 (4'), 31 (5'), 26 (6'). p.25 ↗ Enter the correct value for the enclosure width when setting up the ATI 2G photometer. All TXE models
Aerosol generator compressed air supply should be adjusted to 10 ± 1 psig. p.25 ↗ Maintain correct generator pressure for valid leak test aerosol concentration. All TXE models
Sampling nozzle traverse rate shall not exceed 2 inches per second, with sampling port not more than 1 inch from the filter media. p.26 ↗ Follow this scan rate and distance for valid leak test results. All TXE models
HEPA filter clamp bolts are tightened until the HEPA filter gasket is compressed 50%. p.23 ↗ Uniformly tighten the two filter clamp bolts with a 9/16" socket to this compression. All TXE models
HEPA filters typically last two to five years before requiring replacement. p.22 ↗ Plan filter replacement based on face velocity drop below 80–100 fpm at full speed or leak test failure. All TXE models
Exhaust connection accepts 6" diameter ductwork. p.12 ↗ Duct system must be sized for 6" connection and supported independently of the enclosure. All TXE models with outside exhaust
The TXE motorized impeller cannot be exposed to acids; acid applications require a separate ventilated enclosure or vented fume hood with remote blower ducted outside. p.24 ↗ Do not use the TXE for acid applications; impeller life will be limited. All TXE models
Leave the blower on for at least one minute after work; allow two to three minutes undisturbed for final purging before shutdown. p.21 ↗ Purge airborne contaminants from the work area before turning off the blower. All TXE models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Contaminants outside of enclosure. p.44 ↗ Improper user techniques for the enclosure. See "Certifying the Enclosure" Chapter 3 and "Safety Precautions" Chapter 4 sections in the manual. (Ref. Appendix D). Chapter 8
Contaminants outside of enclosure. p.44 ↗ Restriction of the baffle air slots or blockage of the exhaust outlet. Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. Chapter 8
Contaminants outside of enclosure. p.44 ↗ External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. See "Location Requirements" Chapter 2, "Certifying the Enclosure" Chapter 3, and "Safety Precautions" Chapter 4 sections of this manual. (Ref. Appendix D). Chapter 8
Contaminants outside of enclosure. p.44 ↗ Enclosure has improper face velocity. Have enclosure certified and check exhaust system. Check HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average face velocity of 60-100 fpm for weighing procedures and 75-105 fpm for biohazardous procedures. Chapter 8
Blower won't operate. p.44 ↗ Unit not plugged into outlet. Plug the enclosure into appropriate electrical service. Chapter 8
Blower won't operate. p.44 ↗ Circuit breaker(s) or Ground Fault Interrupter. Reset circuit breaker. Chapter 8
Blower won't operate. p.44 ↗ Blower wiring is disconnected. Inspect blower wiring. Chapter 8
Blower won't operate. p.44 ↗ Blower switch is defective. Replace switch. Chapter 8
Blower won't operate. p.44 ↗ Motorized impeller or blower is defective. Replace motorized impeller or blower. See Chapter 6. Chapter 8
Low face velocity or poor containment of contaminants. p.44 ↗ Enclosure sash not closed. Close sash to the lowest position. Chapter 8
Low face velocity or poor containment of contaminants. p.44 ↗ HEPA filter clogged. Replace HEPA filter or increase speed. Chapter 8
Low face velocity or poor containment of contaminants. p.44 ↗ Carbon filter loaded with chemicals. Replace carbon filter. Chapter 8
Blower and lights won't operate. p.44 ↗ Unit not plugged into outlet. Plug enclosure into appropriate electrical service. Chapter 8
Blower and lights won't operate. p.44 ↗ Circuit breaker(s) tripped. Reset or replace circuit breaker. Chapter 8
Lights do not work. p.45 ↗ Lamp not installed properly. Inspect lamp installation. Chapter 8
Lights do not work. p.45 ↗ Lamp wiring disconnected. Inspect lamp wiring. Chapter 8
Lights do not work. p.45 ↗ Defective lamp. Replace lamp. Chapter 8
Lights do not work. p.45 ↗ Light switch is defective. Replace light switch. Chapter 8
Lights do not work. p.45 ↗ Defective electronic ballast. Replace ballast. Chapter 8
Airflow monitor malfunction. No power. No lights. No display. p.45 ↗ Power cable to circuit board is disconnected and needs to be connected. Sensor cable needs to be connected. Power supply is not plugged into proper voltage; plug in power supply. Verify that all airflow monitor interface cables are connected. Check fuses on enclosure. Chapter 8
No audible alarm. p.45 ↗ Alarm has been temporarily silenced using "SILENCE ALARM" or "enter" buttons. Wrong alarm set point. Airflow monitor was not properly adjusted. Repeat calibration steps outlined in this manual in Chapter 6 or in the supplied Digital 1000 Monitor Manual. Chapter 8
Constant audible alarm. p.45 ↗ Airflow monitor was not properly adjusted. Check airflow and calibration of airflow monitor. See Chapter 6. Chapter 8
Continuous alarm. p.45 ↗ Airflow of the system may have changed; HEPA filter may have become loaded. Check the face velocity of the enclosure. If incorrect, adjust the speed control to increase face velocity. If face velocity is correct, calibrate the airflow monitor outlined in this manual in Chapter 6. Chapter 8
Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. p.45 ↗ Airway to airflow monitor sensor is blocked by insects, dust or debris. Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. Chapter 8
Audible disable will not stay operational. p.45 ↗ An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. Chapter 8