Technical Reference Summary

Covered Models & Document Scope

Protector Demonstration Hoods, 3 ft wide, in two heights: 22.75 in with 8 in sash opening and 32 in with 10 in sash opening. Covers the Mobile Protector Demonstration Hood system (3945001) including FilterMate Portable Exhauster, 5 in hose, mobile cart and work surface. Accessory airflow monitors: Guardian 500 (Kit 3944700 or 3944701) and Guardian Digital 1000 (Kit 3908800 or 3908801). Accessories include work surfaces, fluorescent lights, exhaust transition adapters, remote blowers, exhaust dampers, FilterMate exhausters/filters, storage cabinets, mobile cart 8026000, and hose kits.

Section-by-Section Technical Breakdown

pp. 7-9
Prerequisites: Support, Location, Exhaust and Electrical Requirements

Site preparation requirements for the Protector Demonstration Hood. The supporting structure must be level and solid, typically a base cabinet or stand with chemically resistant work surface; a mobile cart is allowable. The hood must be located away from high-traffic areas, HVAC diffusers, fans, radiators, doorways and windows that may be opened. Exhaust connection accepts a 2 in x 10 in nominal collar with top or bottom connection; only one is required. Electrical receptacles must be nearby, and the airflow monitor must be powered from the FilterMate switched auxiliary outlet (8 A) or the same circuit as the exhaust blower.

  • Support structure must be level and of solid construction; base cabinet or base stand with chemically resistant work surface.
  • Avoid high-traffic areas, overhead/wall HVAC diffusers, fans, radiators, and doorways/windows that may be opened.
  • Exhaust connection accepts 2 in x 10 in (5.1 cm x 25.4 cm) nominal collar; top or bottom connection, only one required.
  • Bottom exhaust is used on the Mobile Protector Demonstration Hood to connect to the FilterMate Portable Exhauster.
  • FilterMate auxiliary switched outlet is rated for 8 amps; fluorescent light kit auxiliary outlet can supply any accessory under 8 amps.
  • Airflow monitor must be connected to the FilterMate switched auxiliary outlet so it is ON when the blower is ON.
pp. 8
Exhaust Volume and Static Pressure Loss by Model

Exhaust volume and enclosure static pressure loss are tabulated for each model at face velocities of 60, 80 and 100 fpm. The 3 ft wide, 22.75 in high hood with 8 in sash opening requires 110/145/185 CFM at 0.04/0.06/0.10 in w.g. The 3 ft wide, 32 in high hood with 10 in sash opening requires 140/185/230 CFM at 0.06/0.10/0.15 in w.g. Proper blower selection is determined from these exhaust requirements and total system static pressure loss.

  • 3 ft x 22.75 in hood with 8 in sash: 60 fpm = 110 CFM at 0.04 in w.g.; 80 fpm = 145 CFM at 0.06 in w.g.; 100 fpm = 185 CFM at 0.10 in w.g.
  • 3 ft x 32 in hood with 10 in sash: 60 fpm = 140 CFM at 0.06 in w.g.; 80 fpm = 185 CFM at 0.10 in w.g.; 100 fpm = 230 CFM at 0.15 in w.g.
  • Enclosure must be connected to a dedicated blower, house exhaust system, or dedicated filtered exhauster such as FilterMate Portable Exhauster.
  • Labconco offers three accessory remote blowers listed in Chapter 7.
pp. 10-12
Installing the Hood on Supporting Structure, Work Surface and Exhaust System

Installation procedure for mounting the hood on a work surface and connecting to the exhaust system. The work surface must be smooth, rigid, non-porous, chemically resistant, and at least as wide as the hood, with a dished recessed area for spills. The front of the hood must be aligned within 1/8 in of the front of the work surface. Mounting screws are inserted with a minimum 1/8 in clearance under the head, and the hood flanges slide under the screw heads. Exhaust ductwork weight must be supported independently of the hood superstructure.

  • Work surface should be smooth, rigid, durable, non-porous, chemically resistant (e.g., epoxy resin) and contain a dished recessed area.
  • Work surface must be at least as wide as the hood; front of hood aligned within 1/8 in of front of work surface.
  • Insert four mounting screws with minimum 1/8 in clearance under the head; slide rear flange and front air foil flanges under screw heads and tighten.
  • Exhaust ductwork weight must be supported independently of the superstructure or damage may occur.
  • Exhaust system should be installed by a qualified HVAC contractor.
  • Hoods weigh either 100 or 120 lbs (45 to 54 kg); do not lift by the front air foil.
pp. 18
Connecting Electrical Supply and Sealing to Work Surface

A standard duplex receptacle should be nearby for the FilterMate Portable Exhauster, airflow monitor and accessories. All building duplex outlet wiring should be performed by a licensed electrician and conform to local codes. After the hood is set, ducted and wired, it may be sealed to the work surface with a bead of silicone sealant to prevent spilled materials from collecting under the walls.

  • Standard duplex receptacle required nearby for FilterMate, airflow monitor and accessories.
  • Building duplex outlet wiring should be performed by a licensed electrician and conform to all local codes.
  • Seal hood to work surface with a bead of silicone sealant after installation.
pp. 18
Validating the Hood: Face Velocity Measurement

Validation procedure for determining average face velocity at the sash opening. Readings are taken in two rows across the enclosure, 6 in from the ends and evenly spaced every 12 in; the first row is 3 in down from the upper sash foil and the second row is 3 in up from the work surface. Labconco recommends an average face velocity of 60 to 100 fpm. The hood is factory-tested per ASHRAE 110-1995 with an as-manufactured rating of less than 0.05 ppm at 4 lpm.

  • Take airflow velocity readings in two rows: 6 in from ends, evenly spaced every 12 in.
  • First row 3 in down from upper sash foil; second row 3 in up from work surface.
  • Recommended average face velocity at sash opening: 60 to 100 fpm.
  • Factory-tested per ASHRAE 110-1995; as-manufactured rating AM <0.05 ppm at 4 lpm.
  • Face velocity profiles and smoke testing should be performed frequently per organization's quality system.
pp. 19-23
High Performance Features

Description of the aerodynamic and containment features of the Protector Demonstration Hood. Features include the Clean-Sweep air foil, containment-enhancing upper sash foil, upper dilution air supply (5-7% of required air volume), visible clear rear baffle, side entry air foils, ergonomic slope, 23 in internal depth, two heights (22.75 in or 32 in), flush sash with spring-loaded latch, electrical pass-through iris, and accessory exhaust connections. Accessory airflow monitors, fluorescent light kits, roof-mounted blower, FilterMate Portable Exhauster, and mobile cart are described.

  • Upper dilution air supply introduces 5 to 7 percent of required air volume to ensure maximum containment.
  • Internal depth of 23 in supports equipment without extending outside the hood or resting on the lower air foil.
  • Two standard heights: 22.75 in or 32 in; taller hoods typically used for pipet operations, titration or taller auxiliary equipment.
  • Bottom exhaust incorporates a tissue screen to prevent debris from clogging the exhaust blower.
  • Exhaust transition connectors available for either 6 in OD outside exhaust or 5 in ID hose.
  • FilterMate Portable Exhauster provides up to 280 cfm for HEPA or carbon filtration, or up to 220 cfm for combination HEPA/carbon filtration.
pp. 25-26
Using Your Protector Demonstration Hood

Operating procedure covering planning, start-up, loading, work techniques, final purging, unloading and shutdown. Turn on exhaust system and accessory light; only raise sash for loading; check baffle air slots for obstructions; allow enclosure to operate unobstructed for 1 minute. After work, allow the hood to operate for two to three minutes undisturbed to purge airborne contaminants before shutting down the blower. Decontaminate objects before removal and cover open containers.

  • Allow enclosure to operate unobstructed for 1 minute after start-up before beginning work.
  • Only raise the sash for loading; keep sash down during use.
  • After work, allow hood to operate for two to three minutes undisturbed before shutting down blower.
  • Surface decontaminate objects in contact with contaminated material before removal; cover open trays and containers.
  • Keep all materials inside the lower air foil and perform contaminated operations as far to the rear as possible.
pp. 27-28
Routine Maintenance Schedule

Suggested maintenance schedule for the hood. Weekly: clean interior and operate exhaust system noting airflow velocity with visible smoke. Monthly (or more often): determine actual face velocity with calibrated thermal anemometer, clean exterior surfaces, check rear baffle for blockage. Annually: have hood validated by a qualified certification technician. Only trained and experienced certification technicians should perform certain service operations after proper decontamination.

  • Weekly: clean hood interior; operate exhaust system and note airflow velocity using visible smoke.
  • Monthly: determine actual face velocity with calibrated thermal anemometer or approved apparatus; clean exterior; check rear baffle for blockage.
  • Annually: have hood validated by a qualified certification technician.
  • Only trained and experienced certification technicians should perform service operations requiring the wrench icon after proper decontamination.
pp. 28-31
Guardian 500 Airflow Monitor: Installation, Calibration and Alarm

The Guardian 500 Airflow Monitor (Kit 3944700 or 3944701) continuously monitors airflow with visual and audible alarms. Green light indicates normal flow; flashing red light and audible alarm activate when flow is below set point. Low flow set points are typically 10-20 fpm below normal flow. Installation involves removing a 1.19 in hole plug, installing elbow/locknut/washer, cutting 1 in hose to 15.5 in, and securing monitor with double-stick tape. Calibration requires 10-minute warm-up and adjustment of potentiometer in small increments about 30 seconds apart. Alarm activates approximately 10-30 seconds after alarm condition.

  • Guardian 500 Kit No. 3944700 or 3944701; built-in sensor, 3 relay inputs and 1 relay output.
  • Low flow set points are 10-20 fpm below normal flow; alarm activates approximately 10-30 seconds after set point reached.
  • Installation: remove 1.19 in dia. gray hole plug; cut 1 in hose to 15.5 in approximate length.
  • Calibration: allow 10 minutes warm-up; adjust potentiometer in small increments about 30 seconds apart.
  • Typical alarm set points: 100 fpm normal -> 80-90 fpm alarm; 80 fpm -> 60-70 fpm; 60 fpm -> 40-50 fpm.
  • If airflow cannot be adjusted, 1/4 to 1/3 turn counterclockwise sets alarm at 10-25% below normal operating levels.
pp. 33
Guardian Digital 1000 Airflow Monitor: Operation and Installation

The Guardian Digital 1000 Airflow Monitor (Kit 3908800 or 3908801) consists of airflow sensor, alarm unit and 15 VDC power supply. It displays face velocity, provides RS232 communications, can be configured for external inputs such as night setback or external alarm, and provides up to three configurable output relays. LEDs indicate SAFE (green), CAUTION (yellow), and LOW (red) with audible alarm. Installation is similar to Guardian 500 but the 1 in hose is cut to 10.5 in approximate length. For 115V operation the alarm unit is powered from the factory-prepared digital airflow monitor socket; for 230V from a building outlet.

  • Guardian Digital 1000 Kit No. 3908800 or 3908801; includes airflow sensor, alarm unit and 15 VDC power supply.
  • Displays face velocity value; provides RS232 communications port and up to three configurable output relays.
  • LEDs: green SAFE, yellow CAUTION, red LOW with audible alarm; audible alarm can be permanently muted.
  • Installation: remove 1.19 in dia. hole plug; cut 1 in hose to 10.5 in approximate length.
  • For 115V operation, plug power supply into factory-prepared digital airflow monitor socket; for 230V, plug into building outlet.
pp. 41-45
Accessories: Work Surfaces, Remote Blowers, Dampers, Storage Cabinets

Accessory ordering information. Optional dished work surfaces are available in black or gray, 48 in x 26.66 in x 1 in (Catalog 3908402 and 3908405). Remote blowers include 115V 60Hz 4.4A (4863500), 115/230V 50Hz 5.6/2.8A (4863501), and explosion-proof 115V 60Hz 4.4A (7053501); performance table gives CFM at static pressures from 0.0 to 0.87 in H2O. Exhaust dampers include 6 in epoxy-coated steel (3924000) and 6 in PVC (4724200). Storage cabinets are listed by size and type (solvent, acid, standard base, vacuum pump) with catalog numbers.

  • Dished work surfaces: 48 in x 26.66 in x 1 in; black 3908402, gray 3908405.
  • Remote Blower 4863500: 115V, 60Hz, 4.4A; 4863501: 115/230V, 50Hz, 5.6/2.8A; 7053501: explosion-proof, 115V, 60Hz, 4.4A.
  • Remote blower performance: 595 CFM at 0.0 in S.P., 560 at 0.125 in, 515 at 0.25 in, 420 at 0.50 in, 300 at 0.75 in, 167 at 0.87 in.
  • Exhaust dampers: 6 in epoxy-coated steel in-line adjustable (3924000); 6 in PVC in-line adjustable (4724200).
  • Storage cabinets available in solvent, acid, standard base, and vacuum pump types with various widths and door configurations.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Recommended average face velocity at sash opening is 60 to 100 fpm. p.18 ↗ Blower/damper must be sized and adjusted to maintain face velocity within this range for proper containment. All Protector Demonstration Hoods
3 ft wide, 22.75 in high hood with 8 in sash opening requires 110 CFM at 60 fpm, 145 CFM at 80 fpm, and 185 CFM at 100 fpm; static pressure loss 0.04, 0.06, and 0.10 in w.g. respectively. p.8 ↗ Use these values to select a blower or exhaust system that can deliver the required volume at the system static pressure. 3 ft x 22.75 in model with 8 in sash opening
3 ft wide, 32 in high hood with 10 in sash opening requires 140 CFM at 60 fpm, 185 CFM at 80 fpm, and 230 CFM at 100 fpm; static pressure loss 0.06, 0.10, and 0.15 in w.g. respectively. p.8 ↗ Taller hood requires higher exhaust volume and has higher static pressure loss; blower selection must account for this. 3 ft x 32 in model with 10 in sash opening
Exhaust connection accepts a 2 in x 10 in (5.1 cm x 25.4 cm) nominal collar; only one exhaust connection (top or bottom) is required. p.8 ↗ Transition adapters are available for 5 in hose or 6 in duct; plan ductwork accordingly. All Protector Demonstration Hoods
FilterMate auxiliary switched outlet is rated for 8 amps; fluorescent light kit auxiliary outlet can supply any accessory under 8 amps. p.9 ↗ Do not exceed 8 A total load on these auxiliary outlets; airflow monitor should be connected to the FilterMate switched outlet so it is ON when blower is ON. All Protector Demonstration Hoods with FilterMate or fluorescent light accessory
Hoods weigh either 100 or 120 lbs (45 to 54 kg); do not lift by the front air foil. p.10 ↗ Use mechanical lift truck or floor jack; if manual lifting, follow safe-lifting guidelines and avoid lifting by air foil. All Protector Demonstration Hoods
Front of hood should be aligned within 1/8 in of the front of the work surface; mounting screws require minimum 1/8 in clearance under the head. p.11 ↗ Ensures proper seating and secure mounting of hood on work surface. All Protector Demonstration Hoods
Upper dilution air supply introduces 5 to 7 percent of the required air volume. p.22 ↗ This bypass air helps contain powders and chemical fumes near the sash plane; blower sizing must account for total volume including dilution air. All Protector Demonstration Hoods
Internal depth of 23 in provides space to support equipment without extending outside the hood or resting on the lower air foil. p.22 ↗ Equipment deeper than 23 in may protrude or obstruct airflow; plan apparatus placement accordingly. All Protector Demonstration Hoods
Large bulky objects should be elevated 3/4 in on level, secured blocks to allow airflow under the object and into rear baffle exhaust slots. p.24 ↗ Prevents airflow blockage and turbulence; ensures proper containment around equipment. All Protector Demonstration Hoods
Guardian 500 low flow set points are typically 10-20 fpm below normal flow; alarm activates approximately 10-30 seconds after set point reached. p.28 ↗ Calibrate alarm set point considering normal airflow fluctuations and cross drafts. Guardian 500 Airflow Monitor
Guardian 500 calibration requires 10-minute warm-up; adjustment screw should be turned in small increments about 30 seconds apart. p.31 ↗ Allows for delayed reaction of alarm; ensures accurate set point. Guardian 500 Airflow Monitor
Guardian 500 installation requires cutting the supplied 1 in hose to 15.5 in approximate length; Guardian Digital 1000 requires 10.5 in approximate length. p.29 ↗ Hose length differs between monitor models; use correct length for proper airflow sensing. Guardian 500 and Guardian Digital 1000 Airflow Monitors
Remote Blower 4863500: 115V, 60Hz, 4.4A; 4863501: 115/230V, 50Hz, 5.6/2.8A; 7053501: explosion-proof, 115V, 60Hz, 4.4A. Performance: 595 CFM at 0.0 in S.P. down to 167 CFM at 0.87 in S.P. p.43 ↗ Select blower based on required CFM and system static pressure; explosion-proof model for hazardous environments. Accessory remote blowers
FilterMate Portable Exhauster provides up to 280 cfm for HEPA or carbon filtration, or up to 220 cfm for combination HEPA/carbon filtration. p.23 ↗ Verify that FilterMate capacity meets hood exhaust volume requirements, especially for taller hood at 100 fpm (230 CFM). Mobile Protector Demonstration Hood and FilterMate accessory
Factory-tested per ASHRAE 110-1995 with as-manufactured rating AM <0.05 ppm at 4 lpm. p.18 ↗ Provides baseline containment performance; field certification should be performed to confirm. All Protector Demonstration Hoods
Hood should be validated before initial use, whenever relocated or serviced, and at least annually thereafter. p.24 ↗ Regular validation ensures continued safe operation and containment. All Protector Demonstration Hoods
After work, allow hood to operate for two to three minutes undisturbed to purge airborne contaminants before shutting down blower. p.26 ↗ Ensures contaminants are exhausted before blower shutdown, reducing exposure risk. All Protector Demonstration Hoods

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Contaminants outside of enclosure p.47 ↗ Improper user techniques for the enclosure. See “Certifying the Hood” Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix D) Chapter 8 Troubleshooting
Contaminants outside of enclosure p.47 ↗ Restriction of the baffle air slots or blockage of the exhaust outlet. Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. Chapter 8 Troubleshooting
Contaminants outside of enclosure p.47 ↗ External factors are disrupting the hood airflow patterns or acting as a source of contamination. See “Location Requirements” Chapter 2, “Certifying the Hood” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix D) Chapter 8 Troubleshooting
Contaminants outside of enclosure p.47 ↗ Hood has improper face velocity. Have hood certified and check exhaust system. Check FilterMate filters for loading. Adjust FilterMate speed control. Hood should have average face velocity of 60-100 fpm. Chapter 8 Troubleshooting
Airflow monitor malfunction. No power. No lights. No display. p.48 ↗ Power supply is not plugged into proper voltage; plug in power supply. Verify that airflow monitor interface cables are connected. Check fuses on FilterMate or accessory light. Plug in power supply to proper voltage; verify interface cables; check fuses on FilterMate or accessory light. Chapter 8 Troubleshooting
No audible alarm. p.48 ↗ Alarm has been temporarily silenced using “test/reset” or “enter” buttons. Press test/reset or enter button to re-enable audible alarm, or wait for alarm condition to reset. Chapter 8 Troubleshooting
Wrong alarm set point. p.48 ↗ Airflow monitor was not properly adjusted. Repeat calibration steps outlined in the airflow monitor manual. Chapter 8 Troubleshooting
Constant audible alarm. p.48 ↗ Check airflow and calibration of airflow monitor. Check airflow and calibration of airflow monitor. Chapter 8 Troubleshooting
Continuous alarm. p.48 ↗ Check the face velocity of the enclosure as the airflow may have changed. If face velocity is correct, calibrate the airflow monitor as outlined in the User’s Manual. Check face velocity; if correct, calibrate airflow monitor per User’s Manual. Chapter 8 Troubleshooting
Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. p.48 ↗ Airway to airflow monitor sensor is blocked by insects, dust or debris. Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. Chapter 8 Troubleshooting
Audible disable will not stay ON p.48 ↗ An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset itself. Action should be taken to bring the hood airflow into proper operating parameters or adjust the alarm set point lower. Chapter 8 Troubleshooting