Rev a waterpro bt with remote manual
Technical Reference Summary
Covered Models & Document Scope
WaterPro® BT Water Purification System with Remote Dispense, including non-UV and UV variants, for 100 VAC, 120 VAC and 230 VAC. Covers installation, operation, calibration, maintenance and troubleshooting.
Section-by-Section Technical Breakdown
Defines the RO water flowrate, dispensing flowrate, reject water flowrate and product water quality for the WaterPro BT with Remote Dispenser. Resistivity, conductivity, TOC and micro-organism limits are stated.
- RO water flowrate: 0.8 GPH (3 LPH) ±15% @ 25°C
- Dispensing flowrate: up to 11.1 GPH (42 LPH)
- Reject water flowrate: 3.9 GPH (15 LPH) to 6.6 GPH (25 LPH) for 5°C < T < 35°C
- Resistivity: 18.2 MΩ·cm @ 25°C; Conductivity: 0.056 µS/cm @ 25°C
- TOC: <5 µg/L (ppb) with UV; <10 µg/L (ppb) without UV
- Micro-organisms: <0.1 CFU/ml with 0.22 µm Final Filter
Lists the main hydraulic and electrical components and describes the purification train from feedwater through pretreatment, RO, tank storage, UV, ion exchange and final filtration. Consumable devices are identified.
- Components include Booster Pump, Inlet Solenoid Valve, Filter Pack (pretreatment + RO cartridge), RO Reject Solenoid Valve, RO Reject Capillary, Check Valve, RO Permeate Conductivity Cell, Tank, Distribution Pump, UV Lamp 185 nm (UV System), Ion Exchange Polisher Cartridge, Product Resistivity Cell, Remote POU Solenoid Valve and Final Filter
- Filter Pack is an all-in-one twin-cartridge: first cartridge combines pretreatment media and RO membrane; second contains ion exchange resin
- RO membrane produces permeate (stored in tank) and reject (sent to drain)
- During dispensing, Distribution Pump sends RO water through UV Lamp (UV System) then Ion Exchange Polisher and Final Filter
- UV Lamp emits at 185 nm and 254 nm to reduce organic molecules
- Filter Pack, UV Lamp and Final Filter are consumables requiring periodic replacement
Provides dimensions, weights, noise level and electrical requirements for the WaterPro BT with Remote Dispenser, including UV and non-UV variants. Electrical source must be earth grounded and within 8.2 ft (2.5 m).
- System Shipping Box: H 22.5 in (57 cm) × W 16.5 in (42 cm) × D 25.2 in (64 cm)
- Remote Shipping Box: H 22.5 in (57 cm) × W 16.4 in (41.5 cm) × D 10.4 in (26.5 cm)
- Operating weight: 44.7 lbs (20.3 kg) non-UV; 46.1 lbs (14.3 kg) UV
- Dry weight: 22.5 lbs (10.2 kg) non-UV; 23.6 lbs (9.2 kg) UV
- Shipping weight: 35.1 lbs (15.9 kg) non-UV; 36.2 lbs (16.4 kg) UV
- Maximum noise level: 50 dB at 39.4 in (1 m)
- 100 VAC ±10%, 50/60 Hz, 0.56 A source, 2 A T fuse, 80 VA
- 120 VAC ±10%, 50/60 Hz, 0.56 A source, 2 A T fuse, 80 VA
- 230 VAC ±10%, 50/60 Hz, 0.31 A source, 2 A T fuse, 80 VA
- Power source must be within 8.2 ft (2.5 m) and earth grounded
Specifies feedwater quality and pressure limits, connection type, drain capacity and environmental operating conditions. These limits must be met for proper operation.
- Feedwater type: potable; flowrate ≥10.7 GPH (40 LPH)
- Feedwater pressure: ≥7.3 psi (0.5 bar) and ≤87.0 psi (6 bar)
- Feedwater conductivity <2000 µS/cm; temperature 5–35°C; pH 4–10
- Fouling index <10; iron <0.1 ppm as CaCO3; aluminum <0.05 ppm; manganese <0.05 ppm
- Free chlorine <1 ppm; Langelier Saturation Index <+0.2; TOC <2000 ppb
- Feedwater connection: 1/2 inch Male GAZ, NPTM or BSPM
- Drain capacity: 7.9 GPH (30 LPH)
- Indoor use only; storage and operating temperature 5–40°C
- Relative humidity ≤80% below 31°C; 50–80% between 31°C and 40°C
- Altitude <9843 ft (3000 m); Installation Category II; Pollution Degree 2
Describes routing and connection of feedwater, reject, overflow and pure permeate tubing, and connection of the remote dispenser tubing and electrical connectors.
- Feedwater tubing has 1/2 inch Female GAZ fitting with screen filter; apply white tape to male thread of feedwater source
- Reject tubing and overflow tubing must be routed and secured to drain
- Tank Outlet Valve and Pure Permeate Tubing are installed from Accessories Bag; open Tank Outlet Valve to allow tank emptying during Filter Pack flush
- Remote dispenser tubing: black to black, transparent to transparent
- Connect 2 electrical connectors (Molex + Picoflex) to PC Board
- Verify system operates correctly before installing cover panel
Covers opening the front cover to allow STANDBY mode, connecting the power cord, opening feedwater supply and observing start-up displays.
- Open front cover (may require removal of Final Filter) to allow system to go into STANDBY mode once powered
- Plug Power Cord into system and appropriate electrical source; open Feedwater Supply Valve
- System displays information before going into STANDBY mode
Details installation of the Filter Pack cartridge, the automatic 15-minute FLUSH mode, and the required rinsing period in FILLING TANK mode before use.
- Open Tank Outlet Valve before installing new Filter Pack to prevent tank filling until rinsed
- Do not touch UV Lamp when replacing Filter Pack
- Wet O-rings on ports with water, preferably ultrapure water
- Install Filter Pack fully seated into system ports and close front cover
- System enters FLUSH mode for 15 minutes to remove air and hydrate media
- After FLUSH, system enters FILLING TANK mode automatically
- Rinse in FILLING TANK mode for minimum 2 hours with Tank Outlet Valve open to drain; overnight recommended for sensitive applications
- Tank fill takes approximately 1.5 to 2 hours after closing Tank Outlet Valve
Describes installation of the Vent Filter and temporary Clear Tubing with Barbed Fitting on the POU Dispenser for initial flushing and purging.
- Insert Vent Filter firmly into port
- Screw Barbed Fitting onto bottom end of POU Dispenser; do not use white tape (O-ring provides seal)
- Push Clear Tubing onto Barbed Fitting and place other end in sink
Covers dispensing water to purge air from the Filter Pack and hydrating the system overnight before Final Filter installation.
- Verify full tank via Tank Level display
- Press Dispensing Button once to enter DISPENSING mode
- Dispense to 60% tank level (approximately 3.6 L); system enters FILLING TANK mode below 60%
- Leave system overnight or several hours in FILLING TANK or PRE OPERATE mode; do not leave in STANDBY
- Dispense about 1 L before removing Clear Tubing and Barbed Fitting
Details installation of the Final Filter, venting trapped air, and rinsing before use. Air trapped in the Final Filter reduces product water flow.
- Screw Final Filter onto POU Dispenser until hand tight; do not over tighten
- Do not use white tape on Final Filter threads; O-ring provides seal
- Open air vent valve on top of Final Filter slowly to vent air
- Air will not pass through membrane; trapped air reduces Product Water flowrate
- Dispense about 1 L to purge all air, then leave system in FILLING TANK mode
- Do not put Clear Tubing or other tubing on Final Filter outlet (bacteria growth risk)
Procedure to calibrate the Product Water flowrate using a 1 L graduated cylinder and the F02 menu option. Calibration dispenses water for 60 seconds.
- Enter menu by pressing Main and '-' buttons together; Display shows F01
- Press Main once to show F02 and 60-second timer
- Press Dispensing Button once; system dispenses for 60 seconds
- Calibration cancels if Dispensing Button pressed again or tank level reaches 10% before 60 seconds
- Measure volume with 1 L graduated cylinder and adjust display with '+' or '-' buttons
- Exit menu by holding Main Button for 2 seconds
Menu option C01 selects temperature-compensated or non-temperature-compensated resistivity or conductivity units.
- Enter menu with Main and '+' buttons; Display shows C01 and current units
- Selectable units: MΩ·cm @25°C, MΩ·cm, µS/cm @25°C, µS/cm
- Exit menu by holding Main Button for 2 seconds
Sets the resistivity threshold below which the Resistivity display flashes and the red Pack Alarm blinks. Factory default is 15 MΩ·cm @25°C.
- Factory default resistivity setpoint: 15 MΩ·cm @25°C
- Adjustable from 1.0 to 18.0 MΩ·cm @25°C
- If conductivity units selected, setpoint adjustable from 0.999 to 0.055 µS/cm @25°C
- Conductivity setpoint display must be multiplied by 0.001 to get real value (e.g., '055' = 0.055 µS/cm)
Explains display layout and how to dispense water using the Dispensing Button, including automatic stop at 10% tank level.
- Display shows Operating Modes, Operating Parameters, Maintenance/Alarms and Tank Level (100% full, 10% minimum)
- Press and release or press and hold Dispensing Button to dispense
- Dispensing stops automatically at 10% Tank Level
- System designed for permanent connection to feed water and power; do not unplug except for maintenance
Describes STANDBY, FLUSH, FILLING TANK, PRE OPERATE, DISPENSING and AUTO-DISPENSING modes and their behavior.
- STANDBY: front cover removed; system depressurizes with STANDBY blinking for 10 seconds; all operations disabled
- FLUSH: 15 minutes after new Filter Pack installed and cover closed; Tank Outlet Valve must be open
- FLUSH resumes from remaining time if cover removed; new FLUSH cycle starts if power cycled
- 2-minute FAST FLUSH every 24 hours of PRE OPERATE mode
- DISPENSING: Distribution Pump on; resistivity and temperature displayed and remain up to 10 seconds after stop
- AUTO-DISPENSING (F01): dispenses pre-set volumes; resistivity remains displayed up to 10 seconds after stop
Explains how to view last product resistivity/temperature and RO permeate conductivity, and how to manually recirculate water before dispensing.
- Product resistivity and temperature displayed for 5 seconds when requested in FILLING TANK or PRE OPERATE mode
- RO permeate conductivity displayed in µS/cm only; value is last memorized, not real-time
- Manual recirculation runs Distribution Pump for up to 3 minutes
- Auto-recirculation occurs for 3 minutes every 2 hours
- Pressing Dispensing Button during recirculation stops recirculation and starts dispensing
Describes the red Pack Alarm and UV Lamp Alarm icons, including blinking versus steady states and their meanings.
- Blinking red Pack Alarm: Filter Pack lifetime exhausted or product resistivity below Resistivity Setpoint (C02)
- Steady red Pack Alarm: Filter Pack not installed correctly or not installed at all
- Blinking red UV Lamp Alarm: UV Timer displays 0 days; replace UV Lamp and reset UV Timer (C05)
- Steady red UV Lamp Alarm: UV Lamp not installed correctly or not installed at all
Lists maintenance tasks and their triggers, including Filter Pack, Final Filter, Vent Filter, flow calibration, tank level calibration, UV Lamp replacement, screen filter cleaning and sanitization.
- Replace Filter Pack when Pack Alarm blinks, resistivity display blinks, or after system/tank sanitization
- Replace Final Filter when Filter Pack is replaced or when Product Water flowrate drops
- Replace Vent Filter when Filter Pack is replaced
- Calibrate flow when Final Filter is replaced or periodically
- Calibrate tank level periodically (C04)
- Replace UV Lamp and reset UV Timer when UV Lamp Alarm blinks (C05)
- Clean screen filter 2 times a year or as necessary
- Sanitize system once a year; Filter Pack must be replaced after system sanitization
- Sanitize tank once a year
Step-by-step procedure to remove and install the Filter Pack cartridge, including flush, rinse, vent filter replacement, purging air, hydrating and Final Filter replacement.
- Open front cover to enter STANDBY; wait for 10-second depressurization
- Remove Final Filter before removing Filter Pack
- Press thumbs on system and swing pack toward you to remove
- Wet O-rings with water, preferably ultrapure water, before installing new pack
- After installation, system enters 15-minute FLUSH mode with Tank Outlet Valve open
- Rinse in FILLING TANK mode for minimum 2 hours with Tank Outlet Valve open to drain
- Replace Vent Filter whenever Filter Pack is replaced
- Hydrate system and replace Final Filter before use
Procedure to replace a clogged Final Filter, vent air and recalibrate product water flowrate.
- Final Filter normally replaced with Filter Pack or earlier if clogged
- Ensure Filter Pack has been hydrated overnight
- Screw new Final Filter hand tight; do not over tighten
- Do not use white tape on threads; O-ring provides seal
- Open air vent valve slowly to vent trapped air
- Dispense about 1 L to purge air, then leave in FILLING TANK mode
- Recalibrate Product Water flowrate after replacement (F02)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| RO water flowrate is 0.8 GPH (3 LPH) ±15% at 25°C. p.8 ↗ | Sizing of tank and downstream demand must account for this permeate generation rate. | WaterPro BT with Remote Dispenser |
| Dispensing flowrate is up to 11.1 GPH (42 LPH). p.8 ↗ | Point-of-use draw can be much higher than RO generation; tank buffer is required. | WaterPro BT with Remote Dispenser |
| Reject water flowrate is 3.9 GPH (15 LPH) to 6.6 GPH (25 LPH) for feedwater temperature 5°C < T < 35°C. p.8 ↗ | Drain must handle at least 7.9 GPH (30 LPH) capacity per installation requirements. | WaterPro BT with Remote Dispenser |
| Product water resistivity 18.2 MΩ·cm @25°C; conductivity 0.056 µS/cm @25°C. p.8 ↗ | Meets Type I ultrapure water quality when system is properly maintained. | WaterPro BT with Remote Dispenser |
| TOC <5 µg/L (ppb) with UV and <10 µg/L (ppb) without UV. p.8 ↗ | UV variant is required for lower TOC applications. | UV and non-UV variants |
| Feedwater pressure must be ≥7.3 psi (0.5 bar) and ≤87.0 psi (6 bar). p.12 ↗ | Install pressure regulation if supply exceeds 87 psi; booster pump may be needed below 7.3 psi. | Whole series |
| Feedwater conductivity must be <2000 µS/cm; free chlorine <1 ppm; fouling index <10. p.12 ↗ | Feedwater outside these limits will prematurely foul or damage the RO membrane and Filter Pack. | Whole series |
| Feedwater connection is 1/2 inch Male GAZ, NPTM or BSPM; drain capacity 7.9 GPH (30 LPH). p.12 ↗ | Plumbing must provide matching fitting and adequate drain. | Whole series |
| Electrical: 100 VAC ±10% 0.56 A; 120 VAC ±10% 0.56 A; 230 VAC ±10% 0.31 A; all 50/60 Hz, 2 A T fuse, 80 VA. p.10 ↗ | Select model matching site voltage; circuit must be earth grounded and within 8.2 ft (2.5 m). | Whole series |
| Maximum noise level 50 dB at 39.4 in (1 m). p.10 ↗ | Suitable for laboratory environments without acoustic enclosure. | Whole series |
| FLUSH mode runs for 15 minutes after new Filter Pack installation and front cover closure. p.22 ↗ | Tank Outlet Valve must remain open during FLUSH to prevent tank filling with unrinsed water. | Whole series |
| Rinsing in FILLING TANK mode requires minimum 2 hours with Tank Outlet Valve open to drain; overnight recommended for sensitive applications. p.23 ↗ | Commissioning time must allow for rinse; incomplete rinse may exhaust polishing media prematurely. | Whole series |
| Tank fill time is approximately 1.5 to 2 hours after closing Tank Outlet Valve. p.23 ↗ | Plan for this delay before full-capacity dispensing is available. | Whole series |
| Dispensing automatically stops at 10% Tank Level. p.34 ↗ | Minimum tank reserve is 10%; low-level alarms or refill logic should account for this. | Whole series |
| Resistivity Setpoint factory default is 15 MΩ·cm @25°C; adjustable 1.0 to 18.0 MΩ·cm @25°C. p.32 ↗ | Setpoint determines when Pack Alarm blinks; adjust to application water quality requirement. | Whole series |
| Conductivity setpoint display must be multiplied by 0.001 to get real value (e.g., '055' = 0.055 µS/cm). p.32 ↗ | Avoid misreading conductivity setpoint by factor of 1000. | Whole series |
| Manual recirculation runs up to 3 minutes; auto-recirculation occurs for 3 minutes every 2 hours. p.40 ↗ | Recirculation maintains water quality in the distribution loop; manual recirculation recommended before dispensing after standby. | Whole series |
| 2-minute FAST FLUSH cycle occurs every 24 hours of PRE OPERATE mode. p.37 ↗ | Periodic flush helps maintain water quality during low-use periods. | Whole series |
| Screen filter cleaning required 2 times a year or as necessary; sanitization of system and tank once a year. p.45 ↗ | Preventive maintenance schedule must include these tasks; Filter Pack must be replaced after system sanitization. | Whole series |
| Final Filter must be hand tight only; do not use white tape on threads (O-ring provides seal). p.27 ↗ | Over-tightening or tape can damage seal and cause leaks or bypass. | Whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| The Display screen is blank. p.70 ↗ | The power cord is not plugged in. No source of electrical power. Main Power Fuse is blown. | Check that the power cord is plugged in. Check the source of electrical power. Contact your technical support. | |
| In FILLING TANK mode, the Tank Level display is not rising. (The Tank Level display should be rising steadily.) p.70 ↗ | The Tank Outlet Valve is open. The water in the tank is diverted into the drain. No water is kept in the tank. The tank level is not calibrated properly. The RO membrane is dirty or clogged. | Close the Tank Outlet Valve. Check the Feedwater source. See Section How to Calibrate the Tank Level (C04). Replace the Filter Pack. | |
| Low water flow or no water flow when the Dispensing Button is pressed. p.70 ↗ | The water level is less than 10%. No dispensing is allowed. The Tank Outlet Valve is open. The tank is not filling up with water. Final Filter is air locked. Final Filter is clogged. | Allow the system to go into FILLING TANK mode until there is enough water available in the tank, viewed by the Tank Level display. Close the Tank Outlet Valve. Vent all air from the Final Filter. Replace the Final Filter, see Section How to Replace the Final Filter. | |
| DISPENSING mode continuously runs even when the Tank Level display is below 10%. p.70 ↗ | The water level is not calibrated. The system does not detect the minimum level to disable DISPENSING mode. | Calibrate the tank level. See Section How to Calibrate the Tank Level (C04). | |
| AUTO-DISPENSING mode is not accurate. p.70 ↗ | The Product Water flowrate is not calibrated. The Final Filter is clogged or airlocked. | Calibrate the Product Water flowrate. See Section How to Calibrate the Flowrate from the WaterPro® BT System (F02). Replace the Final Filter. | |
| The last RO Permeate conductivity value is not displayed when you press the “+” button. p.70 ↗ | The measurement was not made while dispensing water during FILLING TANK mode. The value is out of measurement range. | Dispense water during FILLING TANK mode to start a RO Permeate conductivity reading again. | |
| The last Product resistivity value is not displayed when you press the “-“ button in FILLING TANK mode or in PRE OPERATE mode. p.71 ↗ | A measurement was not made during DISPENSING mode or during recirculation. The value is out of measurement range. | Dispense or recirculate water manually to start a Product resistivity reading again. | |
| The Tank Outlet Valve is not open. The system detects that there is water in the tank. The system will not resume FLUSH mode until the tank is emptied of water. p.71 ↗ | The Tank Outlet Valve is not open. The system detects that there is water in the tank. | Open the Tank Outlet Valve to drain the water from the tank. Check that the Tank Outlet Valve and Tubing are directed downwards into the drain. | |
| The Filter Pack lifetime is exhausted. The Product resistivity is less than the Resistivity Setpoint set in menu option C02. p.71 ↗ | The Filter Pack lifetime is exhausted. The Product resistivity is less than the Resistivity Setpoint set in menu option C02. | Replace the Filter Pack. See Section How to Replace the Filter Pack Cartridge. | |
| The Filter Pack is not installed correctly or not installed at all. p.71 ↗ | The Filter Pack is not installed correctly or not installed at all. | Reinstall the Filter Pack. If the red Pack Alarm is still displayed, then contact your technical support. | |
| The system may have been in STANDBY mode for a while. The Product resistivity is below the Resistivity Setpoint set in menu option C02. p.71 ↗ | The system may have been in STANDBY mode for a while. The Product resistivity is below the Resistivity Setpoint set in menu option C02. | Recirculate the water in the system. See Section How to recirculate water manually before dispensing. Replace the Filter Pack. See Section How to Replace the Filter Pack Cartridge. | |
| UV System only p.72 ↗ | The UV Timer is exhausted. | Replace the UV Lamp. See Section How to Replace the UV Lamp (UV System only). After a new UV Lamp has been installed, reset the UV Timer. See Section How to View or Reset the UV Lamp Timer (C05). | |
| UV System only p.72 ↗ | The UV Lamp is not installed correctly or not installed at all. | Power OFF the system and reinstall the UV Lamp. If the red UV Lamp Alarm is still displayed, then contact your technical support. |