Rev E WaterPro RO System User Manual

Technical Reference Summary

Covered Models & Document Scope

WaterPro RO Systems, 90750 Series. Covers 115V model 90750-20 and 230V models 90750-30, -40, -60, -70, -80 (wall-mounted). Options/accessories referenced: optional dispensing gun kit (9113100), support stand 90774-00, wall mounting panel 9078800, RO/PS mobile stand 9113200, RO/PS connecting cord 13060-00/1306000, 14-gal (9203001) and 20-gal (9203000) bladder tanks, and 70-L storage tanks (9100000/9100001, 9101000/9101001).

Section-by-Section Technical Breakdown

pp. 9-10
Location and Mounting

Site selection and wall-mounting requirements. The mounting surface must support at least 500 lb; four 1/4 x 2.0 in tapping screws (1897432) are supplied for wood-frame studs. Do not install over or near electrical equipment because of spillage risk.

  • Mounting surface must support at least 500 pounds.
  • Four 1/4 x 2.0 in tapping screws (1897432) provided for typical wood-frame wall studs.
  • Do not install directly over or near equipment using electrical service.
  • Optional support stand 90774-00 converts wall unit to bench-top.
pp. 10
Electrical Requirements

Electrical supply and cord requirements for 115V and 230V models. The 115V model requires a 115 VAC, 60 Hz, 15 A outlet; 230V models require a dedicated 230 VAC, 50 Hz, 10 A outlet. A three-conductor earthed cord and a nearby disconnect are required.

  • 90750-20: 115 VAC, 60 Hz, 15 A outlet; detachable 15 A power cord.
  • 90750-30/-40/-60/-70/-80: 230 VAC, 50 Hz, 10 A dedicated outlet; 10 A cord.
  • Use three-conductor power cord with protective earthing ground.
  • Provide switchable circuit breaker or switch near unit for MAINS disconnect.
pp. 10-11
Feedwater and Drain Requirements

Feedwater must be at least 30 psi at 1.6 gal/min with negative LSI; line pressure must not exceed 100 psi. Two drain lines (tank overflow and RO reject) must handle at least 2 gal/min. Chlorine above 0.5 ppm irreversibly damages the RO membrane.

  • Minimum feed pressure 30 psi; minimum flow 1.6 gal/min (6 L/min).
  • Maximum line pressure 100 psi (7 bar).
  • Drain must handle minimum 2 gal/min.
  • Feedwater chlorine must not exceed 0.5 ppm.
  • Positive LSI requires water softener pretreatment.
pp. 11
Space Requirements

Wall-mounted units need at least 6 inches clearance at the bottom for cartridge and membrane installation/removal. Dimensional drawings are in Appendix B.

  • Minimum 6 in (15.2 cm) bottom clearance for cartridge and membrane service.
pp. 14-15
Component Identification

Identifies the main components and their functions: control panel/PC board, ON/OFF switch, pressure gauge preset to 150 psi, air vent filter, inlet valve, conductivity sensor, pressure switch (~40 psi), rotary vane pump, low-pressure switch, flush valve, dispensing valve, check valve, dispensing outlet port, inlet water fitting, 17-L storage tank, 1/4 HP pump motor, float switch, tank cover, back-pressure relief valve, inline strainer, and inlet pressure gauge (30-100 psi).

  • Pressure gauge indicates RO back pressure, preset to 150 psi; do not adjust.
  • Pressure switch shuts unit down at approximately 40 psi.
  • Low-pressure switch shuts off pump on low inlet pressure.
  • Flush valve opens after 12 hours of inactivity.
  • Float switch turns unit off when 17-L tank is full.
  • Inlet pressure should be between 30 psi and 100 psi (2-7 bar).
pp. 16-17
Setting Up and Tubing Installation

Mounting with the supplied template, feed-water and drain connection, and tubing installation. Feedwater uses 3/8 in OD rigid plastic tubing; an isolation valve should be installed in the supply line. Flexible tubing requires a tube support insert.

  • Feedwater tubing: 3/8 in OD (0.95 cm) rigid plastic.
  • Install isolation valve in feed line.
  • Line pressure must not exceed 100 psi (7 bar).
  • Feed line must supply 1.6 gal/min at minimum 30 psi or pump cycling occurs.
  • Cut tube square, remove burrs, insert into inlet fitting.
  • Flexible tubing requires tube support; rigid tubing does not.
pp. 18-19
RO Membrane Installation

Step-by-step installation of prefilter 9092700, carbon/prefilter 9059400, and the RO membrane in the third bowl. Includes orientation of spacer disk and adapters, bowl tightening, drain connection, and initial 1.5-hour timed dispense.

  • Prefilter 9092700 in first bowl (right side), O-rings up.
  • Carbon/prefilter 9059400 in second bowl from right, O-rings up.
  • Lower spacer disk slots upward; install lower and upper adapters until seated.
  • Do not over-tighten bowl.
  • Set timed dispense for 1.5 hours and start unit for initial run.
pp. 20
Optional Dispense Gun Installation

Field installation procedure for the optional dispense gun, including depressurization, cover removal, tank removal, bracket mounting, connection to the fitting marked 'GUN', and leak check.

  • Close feed valve and relieve pressure before disconnecting power.
  • Connect gun tubing to fitting marked 'GUN' next to pressure switch.
  • Set gun trigger to constant ON and direct to drain during fill/leak check.
  • Place gun in bracket with nozzle through grommet.
pp. 21
WaterPro PS and Glassware Washer Feed

Connecting the RO system to a WaterPro PS Polishing System and/or Labconco Glassware Washer via the storage tank outlet. A tee allows both connections. Notes on 17-L tank capacity and need for optional storage tanks for multiple pure rinse cycles.

  • Empty storage tank, remove gray plug at bottom, insert tubing from PS or washer.
  • Both PS and washer can be connected using a tee.
  • 14-gal (9203001) and 20-gal (9203000) tanks include connection tubing.
  • Without optional tank, washer requires 17-L tank full at start of pure rinse.
  • Multiple pure rinse cycles require an optional storage tank.
pp. 22
Operating Precautions

Operational precautions: pretreat feedwater per Water Profile test, change prefilters frequently, follow electrical codes, disconnect power before servicing, preserve membrane if unplugged over a week, and establish feedwater quality before operation.

  • Pretreat feedwater if indicated by Labconco Water Profile test.
  • Change prefilters frequently to prevent chlorine damage to membrane.
  • Disconnect RO system from outlet before removing or servicing electrical components.
  • Preserve RO membrane if unplugged more than one week.
pp. 24-25
Display and Operation

Display functions and operating modes: conductivity in µS, temperature in °C, timed dispense 0-99.9 minutes, tank full light, and dispense key. Includes initial start-up purge procedure and automatic flush valve behavior.

  • Modes: µS (normal), Temp (°C), Time Dispense (0-99.9 min).
  • Tank Full light illuminates when storage is full.
  • Dispense key must be held down to continue dispensing unless timed mode selected.
  • Initial start-up: set timed dispense to 99 minutes to purge preservative to drain.
  • Automatic flush every 12 hours of inactivity; keep unit plugged in.
pp. 27
Maintenance Schedule

Maintenance intervals: change prefilter and carbon/prefilter after 500 gallons of feedwater or every two months; clean dust; preserve unit if unplugged more than one week.

  • Change prefilter and carbon/prefilter after 500 gallons feedwater processed.
  • Change prefilter and carbon/prefilter once every two months.
  • Clean unit of dust as needed.
  • Preserve unit if unplugged for more than one week.
pp. 29-32
Sanitization and Membrane Cleaning

Sanitization with 0.1% formaldehyde for biological fouling, and cleaning procedures for inorganic fouling (0.5% HCl, pH 2) and organic fouling (NaOH, SDS, TSP, etc.). Includes fouling indicators and preservation procedure.

  • Sanitize with 0.1% (wt/wt) formaldehyde if biologically fouled.
  • Inorganic fouling cleaned with 0.5% HCl, final pH 2.
  • Organic fouling cleaners: 0.5% NaOH, 0.1% SDS (pH 11.5), 1% TSP, 1% sodium tri-polyphosphate, 1% tetra sodium EDTA.
  • Fouling indicated by 20% drop in normalized product flow or noticeable permeate conductivity rise.
  • Preserve membrane with 0.1% formaldehyde if unused over a week.
  • Use only chlorine-free water for cleaning solutions.
pp. 32
Service Tips

Warning that the unit automatically starts and stops during nonuse; always unplug before servicing to prevent water spillage or electrical shock.

  • Always unplug unit prior to servicing.
pp. 35-37
Troubleshooting

Problem/cause/corrective action table covering unit inoperative, pump not operating, reduced flow, erratic water quality display, display reading 000, and conductivity changes.

  • Covers no display, pump not operating, reduced flow, erratic display, 000 reading, and µS increase.
  • Includes causes such as feed valve closed, restricted line, plugged filters, air in conductivity cell, membrane scaling/fouling.
pp. 42-43
Specifications

RO specifications: polyamide thin-film composite membrane, minimum rejection 94%, max pressure drop 6 psig, production up to 1 L/min at 25°C, 8.7 L/min gravity flow from storage tank, electrical 115V/60Hz/12A or 230V/50Hz/6A, sound <70 dB(A). Inlet water limits: 5-40°C, pH 4-10, 30-100 psi, 1.6 gal/min, SDI 5, turbidity <1 NTU, iron/manganese 0.1 ppm, free chlorine 0.5 ppm.

  • Minimum rejection rate 94%.
  • Maximum pressure drop per element 6 psig.
  • RO water production up to 1 L/min at 25°C; 8.7 L/min typical gravity flow from storage tank.
  • Electrical: 115V, 60 Hz, 12 A or 230V, 50 Hz, 6 A.
  • Inlet water temperature 5-40°C (41-104°F); pH 4-10.
  • Maximum silt density index 5 SDI; turbidity <1 NTU.
  • Maximum iron 0.1 ppm, manganese 0.1 ppm, free chlorine 0.5 ppm.
  • Environmental: indoor use, altitude up to 6562 ft (2000 m), RH <80% decreasing to 50% at 104°F.
pp. 39, 46-47
Components and Accessories

Replacement parts list (Appendix A) and expendables/accessories (Appendix D) with part numbers for valves, sensors, pump assemblies, PC board, tank, filters, membrane, gun kit, stands, and storage tanks.

  • Pump assembly 115V 9047503; 230V 9047502.
  • Motor pump 115V 1210101; 230V 1210102.
  • PC board assembly 9112300; conductivity sensor 9105300.
  • RO membrane and adapter package 9078700.
  • Prefilter/carbon filter kit 9067201 includes 3 prefilters and 3 carbon filters.
  • Gun kit 9113100; wall mounting panel 9078800; support stand 9077400; mobile stand 9113200.
  • RO/PS connecting cord 1306000 (125-230 V, 10 A, within 15 inches).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Mounting surface must support at least 500 pounds. p.10 ↗ Verify wall structure before installation; inadequate support may damage surface or equipment. Whole series
Four 1/4 x 2.0 in tapping screws (1897432) are provided for securing to typical wood-frame wall studs. p.10 ↗ Use supplied fasteners for wood stud mounting; other surfaces require suitable anchors. Whole series
115V model 90750-20 requires a 115 VAC, 60 Hz, 15 A outlet; 230V models 90750-30/-40/-60/-70/-80 require a dedicated 230 VAC, 50 Hz, 10 A outlet. p.10 ↗ Provide a dedicated earthed circuit with a nearby disconnect for 230V models. 90750-20 vs 90750-30/-40/-60/-70/-80
Feedwater minimum pressure 30 psi at minimum flow 1.6 gal/min (6 L/min); maximum line pressure 100 psi (7 bar). p.10 ↗ Insufficient pressure/flow causes pump cycling and reduced performance; overpressure may damage system. Whole series
Feedwater chlorine must not exceed 0.5 ppm; higher levels cause irreversible damage to the RO membrane. p.11 ↗ Test feedwater and pretreat if necessary; change prefilters frequently. Whole series
Drain must handle a minimum of 2 gal/min water flow; two drain lines (tank overflow and RO reject). p.11 ↗ Route both drain lines to a drain with adequate capacity. Whole series
Wall-mounted units require a minimum of 6 in (15.2 cm) clearance at the bottom for cartridge and membrane service. p.11 ↗ Plan installation height to allow filter and membrane removal. Wall-mounted models
RO back pressure is preset to 150 psi; do not increase or decrease. p.6 ↗ Preset provides optimal membrane conversion rate and life; field adjustment is not permitted. Whole series
Pressure switch shuts down the unit when system pressure reaches approximately 40 psi. p.14 ↗ Explains automatic stop/start behavior during tank filling. Whole series
Inlet pressure should be between 30 psi and 100 psi (2-7 bar). p.14 ↗ Monitor inlet pressure gauge; adjust supply accordingly. Whole series
Feedwater tubing is 3/8 in OD (0.95 cm) rigid plastic; flexible tubing requires a tube support insert. p.16 ↗ Use correct tubing and insert supports to prevent leaks. Whole series
Initial start-up requires timed dispense set to 99 minutes to purge preservative to drain. p.25 ↗ Do not use water from initial purge; allow tank to fill and unit to shut off automatically. New installations
Automatic flush valve opens every 12 hours of inactivity to flush membrane and storage tank for 3 minutes. p.6 ↗ Keep unit plugged in at all times for flush function to operate. Whole series
Change prefilter and carbon/prefilter after 500 gallons of feedwater or every two months. p.27 ↗ Maintain schedule to prevent chlorine breakthrough and membrane damage. Whole series
Preserve RO membrane with 0.1% formaldehyde if unit will be unplugged for more than one week. p.32 ↗ Prevents biological fouling during extended shutdown. Whole series
Inorganic fouling cleaned with 0.5% HCl (final pH 2); organic fouling with 0.5% NaOH, 0.1% SDS (pH 11.5), 1% TSP, 1% sodium tri-polyphosphate, or 1% tetra sodium EDTA. p.30 ↗ Use only chlorine-free water for cleaning solutions; follow soak and flush procedure. Whole series
Minimum rejection rate 94%; maximum pressure drop per element 6 psig. p.42 ↗ Performance baseline for evaluating membrane condition. Whole series
RO water production up to 1 L/min at 25°C; typical gravity flow from storage tank 8.7 L/min. p.42 ↗ Size storage and downstream equipment based on these flow rates. Whole series
Electrical ratings: 115V, 60 Hz, 12 A or 230V, 50 Hz, 6 A. p.42 ↗ Confirm supply capacity; 230V models require dedicated circuit. Whole series
Inlet water limits: temperature 5-40°C, pH 4-10, SDI max 5, turbidity <1 NTU, iron 0.1 ppm, manganese 0.1 ppm, free chlorine 0.5 ppm. p.42 ↗ Feedwater outside these limits requires pretreatment or risks membrane damage. Whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Unit inoperative no display p.35 ↗ Unit not plugged into outlet Plug WaterPro RO into appropriate electrical service Chapter 6
Unit inoperative no display p.35 ↗ Circuit breaker tripped/fuse blown Reset circuit breaker/replace fuse Chapter 6
Unit inoperative no display p.35 ↗ Power Switch is in the OFF position Press power button to switch system on Chapter 6
Unit inoperative no display p.35 ↗ Defective circuit board Replace circuit board Chapter 6
Pump does not operate and no water is dispensed p.35 ↗ Customer supplied Feed valve is closed Open the valve Chapter 6
Pump does not operate and no water is dispensed p.35 ↗ Feed-water line is restricted, or no supply Inspect and adjust feed-water line as required; Check inline strainer for obstruction Chapter 6
Pump does not operate and no water is dispensed p.35 ↗ Pump is defective Replace pump Chapter 6
Reduced flow at dispense valve/gun and storage tank slow to fill p.36 ↗ Restricted tube in the feed line or polishing system Inspect polishing loop tubing for any restrictions Chapter 6
Reduced flow at dispense valve/gun and storage tank slow to fill p.36 ↗ Insufficient feed water pressure and flow rate Pressure must be a minimum of 30 psi at a flow rate of at least 6 liters per minute; Check inline strainer for obstruction Chapter 6
Reduced flow at dispense valve/gun and storage tank slow to fill p.36 ↗ Partial open feed valve Open feed valve fully Chapter 6
Reduced flow at dispense valve/gun and storage tank slow to fill p.36 ↗ Plugged prefilter and/or carbon filter Replace filters Chapter 6
Reduced flow at dispense valve/gun and storage tank slow to fill p.36 ↗ Water temperature less than 25°C Install water temperature mixing valve Chapter 6
Reduced flow at dispense valve/gun and storage tank slow to fill p.36 ↗ Kinked or restricted tubing in system Trace restriction and remove or straighten tube Chapter 6
Reduced flow at dispense valve/gun and storage tank slow to fill p.36 ↗ Decrease in RO back pressure Increase back pressure relief valve to 150 psi Chapter 6
Water quality display acting erratically p.36 ↗ New cartridges installed in system Purge all of the air out of the polishing system Chapter 6
Water quality display acting erratically p.36 ↗ Air trapped in conductivity cell Purge all of the air out of the RO system Chapter 6
Water quality display acting erratically p.36 ↗ Polishing loop conductivity cell not connected to wiring harness Reconnect cell to wiring harness Chapter 6
Water quality display acting erratically p.36 ↗ Conductivity cell is defective Replace cell Chapter 6
Display reads 000 p.36 ↗ Conductivity sensor disconnected from harness Connect sensor to harness from circuit board Chapter 6
Display reads 000 p.36 ↗ Air in conductivity sensor Purge unit until air is expelled by opening the dispensing valve Chapter 6
Increase in µS (microsiemen) reading p.37 ↗ High inlet TDS This is normal. With increased inlet TDS, RO µS increases Chapter 6
Increase in µS (microsiemen) reading p.37 ↗ RO membrane scaling or other deposits Clean membrane per instructions in RO cleaning section Chapter 6
Excessively high µS p.37 ↗ RO membrane fouled; Biofilm formed Clean RO membrane per instructions in RO cleaning section Chapter 6
Excessively high µS p.37 ↗ RO poisoned; Chlorine damage Replace RO membrane and prefilters per installation instructions Chapter 6