Basic Laboratory Hoods Manual
Technical Reference Summary
Covered Models & Document Scope
Protector Basic Laboratory Fume Hoods, 47-inch (4-foot) and 70-inch (6-foot) models, assembled or unassembled, with or without built-in blower, with vapor-proof incandescent light or explosion-proof blower/light. Catalog numbers 2247300–2247501 (Basic 47), 2246300–2246501 (Basic 70), 2246600–2246606 (unassembled Basic 47), 2246700–2246706 (unassembled Basic 70). Electrical variants: 115 V/60 Hz, 220 V/50 Hz, and explosion-proof 115/230 V 60 Hz or 220 V 50 Hz.
Section-by-Section Technical Breakdown
Lists the catalog numbers and descriptions of all Basic 47 and Basic 70 fume hood variants, including assembled/unassembled, with/without blower, and explosion-proof configurations with their electrical ratings.
- Basic 47 models: 2247300/2247301 with blower, 2247400/2247401 without blower, 2247500/2247501 explosion-proof.
- Basic 70 models: 2246300/2246301 with blower, 2246400/2246401 without blower, 2246500/2246501 explosion-proof.
- Unassembled Basic 47: 2246600–2246606; unassembled Basic 70: 2246700–2246706.
- Electrical variants: 115 V single-phase 60 Hz, 220 V single-phase 50 Hz, explosion-proof 115/230 V 60 Hz or 220 V 50 Hz.
Describes the low-profile 25-inch depth epoxy-coated steel construction and the by-pass airflow design that maintains containment as the sash is moved. The hood is available with or without blower and assembled or unassembled.
- Low overall profile and 25-inch depth fits most work surfaces and low-ceiling areas.
- Epoxy-coated steel construction resists corrosion, impact, cracking and peeling.
- By-pass airflow design diverts air behind the front panel and under the sash foil to minimize face velocity increases as sash is lowered.
- Single wall design maximizes work area.
Identifies key components: sash foil, adjustable baffle, tempered safety glass sash, light fixture, exhaust connection, and blower assembly. Includes details on the 100-watt light bulb, 10-inch duct connection, and blower mounting options.
- Sash foil is part of the air by-pass system and directs smooth airflow into the hood.
- Adjustable baffle has an adjustment slot controlled by loosening baffle retaining knobs.
- Tempered safety glass sash can be positioned at any suitable height.
- Light fixture accepts a 100-watt bulb; explosion-proof models have explosion-proof incandescent light.
- Exhaust outlet ring accepts built-in motor/blower or 10-inch diameter plastic duct.
- Fractional horsepower motor/blower can be mounted in the hood top, in the duct run, or on the roof.
Covers preparation, location, installation factors, electrical connection, sash weight release, placing and sealing, exhaust connection, baffle adjustment, and service line requirements. Includes specific duct sizes, electrical service requirements, and warnings about lifting and duct support.
- Assembled hood ships in one carton; unassembled hood ships structure and blower in separate cartons.
- Locate hood out of main traffic and away from doors, windows, air conditioning, ventilation supply or return outlets.
- Provide supporting structure with chemical-resistant work surface.
- Electrical service: 115 V single-phase 60 Hz or 220 V 50 Hz.
- 10-inch diameter exhaust duct for 4-foot models; 12-inch for 6-foot models.
- Lift hood only by sides and back panel; do not lift by sash foil.
- Explosion-proof models are not internally wired; wiring must be by licensed electrician with conduit shielding.
- Sash weight secured with two plastic wire ties; cut ties, remove packing, slide weight up and attach sash cables.
- Place lower airfoil flush with front of work surface or up to one inch in front; never behind front edge.
- Seal hood to work surface with silicone sealant.
- Exhaust duct under positive pressure; all joints must be sealed.
- Service lines: 1/4-inch OD copper (brass for gas), recommended 40 PSI, maximum 200 PSI.
Provides safety precautions and operating procedures for working in the fume hood, including planning, start-up, loading, work techniques, final purging, unloading, and shutdown. Includes specific clearances and purge times.
- Do not use perchloric acid in this hood.
- Radioisotope materials are not recommended.
- Flammable or explosive materials only with explosion-proof components.
- Do not store containers against the baffle.
- Start-up: turn on light, slowly raise sash, verify airflow, check baffle slots, allow two minutes unobstructed operation.
- Keep materials at least 6 inches inside the sash and at least 2 inches from side walls.
- After loading, wait one minute to purge airborne contaminants.
- After work, allow two minutes undisturbed operation before closing sash.
- Shutdown: turn off light and close sash.
Provides environmental conditions and electrical ratings for all models, plus wiring diagrams for 115 VAC and 230 VAC models.
- Indoor use only; maximum altitude 10,000 feet (3,048 m).
- Ambient temperature range 41° to 104°F (5° to 40°C).
- Maximum relative humidity 80% up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C).
- Main supply voltage fluctuations not to exceed ±10% of nominal.
- Transient over-voltages per Installation Category II: up to 1500 V for 115 V models, 2500 V for 230 V models.
- Pollution degree 2 environment.
- Electrical ratings: Basic 47 and Basic 70 with/without blower: 115 V 60 Hz 13 A; 230 V 50 Hz 7 A; explosion-proof blower: 115/230 V 60/50 Hz 8/4 A (Basic 47) or 13/7 A (Basic 70).
Covers lubrication of the exhaust blower motor and belt tension adjustment for the motor/blower assembly.
- Exhaust blower motor factory lubricated for two years; after that, 20 to 30 drops of SAE 20 automotive oil annually.
- Check belt tightness periodically.
- Belt tension: approximately 5 pounds pressure to deflect belt 1/2 inch from normal position.
- Do not contact blower wheel while in motion.
Explains that face velocity can be set based on work performed by regulating airflow across the sash opening. Actual face velocity is determined by taking airflow velocity readings across the sash opening in accordance with Industrial Ventilation.
- Face velocity is adjustable via combination of hood, exhaust duct, and exhaust blower.
- Airflow velocity readings should be taken across the sash opening.
- Testing should follow the section on laboratory hoods in Industrial Ventilation.
Lists replacement parts for Basic Hoods and blower assemblies with catalog numbers for 47-inch and 70-inch models.
- Hood parts include wall plate blank 9818300, switch 1302300, sash assembly 4847900 (47")/4847901 (70"), globe 1272900, knob 1879400, vibration mount 4847800, sash pulley 1861400, outlet gasket 4850200, explosion-proof light fixture 1270301.
- Blower assembly parts include motors, variable sheave 1860000/1858200, V-belts, fixed sheaves, pillow block 1863000, cover 4849000, housing assembly 2264301, inlet adapter 4850800, wheel assembly 4897300, transition 4852300.
- Switch and globe are not used on explosion-proof models.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| 10-inch diameter exhaust ductwork is required for 4-foot (Basic 47) models; 12-inch diameter for 6-foot (Basic 70) models. p.12 ↗ | Duct sizing must match hood width to maintain proper face velocity and transport velocity. | Basic 47 and Basic 70 models |
| Electrical service required: 115 V single-phase 60 Hz or 220 V 50 Hz. p.12 ↗ | Verify building electrical supply matches model nameplate before installation. | All non-explosion-proof models |
| Explosion-proof models are not internally wired; all wiring must be done by a licensed electrician with conduit shielding and grounding arranged so terminal box cover can be removed without disturbing connections. p.13 ↗ | Additional installation labor and conduit materials required for explosion-proof variants. | Explosion-proof models (2247500, 2247501, 2246500, 2246501, 2246602, 2246605, 2246606, 2246705, 2246706) |
| Service lines should be 1/4-inch OD copper (brass for gas) with easily accessible shut-off valve; recommended operating pressure 40 PSI, maximum allowable 200 PSI. p.15 ↗ | Use pressure regulator if line pressure exceeds 40 PSI; install shut-off valves for maintenance. | All models with service fixtures |
| Hood lower airfoil should be flush with front of work surface, or up to one inch in front; must not be placed behind front edge of work surface. p.14 ↗ | Proper positioning ensures containment and prevents spilled materials from collecting under hood walls. | All models |
| Exhaust duct is under positive pressure during operation; all duct joints must be sealed properly. p.14 ↗ | Leaking joints could release exhaust contaminants into laboratory. | All models |
| Weight of exhaust ductwork must be supported independently of hood structure; do not allow weight to be supported by blower scroll and housing or hood structure. p.14 ↗ | Provide separate duct hangers/supports to avoid damage to hood. | All models |
| Lift hood only by sides and back panel; do not lift by sash foil as it is not directly attached. p.12 ↗ | Improper lifting can damage sash foil or hood structure. | All models |
| Sash weight is secured with two plastic wire ties during shipment; cut ties, remove cardboard packing, slide weight up and attach sash cables into slots. p.13 ↗ | Must release sash weight before operation for proper sash balance. | All models |
| Light fixture accepts a 100-watt bulb; explosion-proof models have explosion-proof incandescent light fixture. p.9 ↗ | Use only 100-watt bulb; explosion-proof models require explosion-proof rated bulb. | All models |
| Environmental conditions: indoor use only, max altitude 10,000 ft (3,048 m), ambient 41° to 104°F (5° to 40°C), max RH 80% up to 88°F decreasing linearly to 50% at 104°F, supply voltage fluctuation ±10%, transient over-voltages up to 1500 V (115 V models) or 2500 V (230 V models), pollution degree 2. p.21 ↗ | Ensure installation environment meets these conditions for safe operation. | All models |
| Electrical ratings: Basic 47 and Basic 70 with/without blower: 115 V 60 Hz 13 A; 230 V 50 Hz 7 A. Explosion-proof blower: Basic 47 115/230 V 60/50 Hz 8/4 A; Basic 70 115/230 V 60/50 Hz 13/7 A. p.21 ↗ | Size branch circuit and overcurrent protection accordingly; ratings exclude receptacle loads. | All models |
| Exhaust blower motor factory lubricated for two years; after that, 20 to 30 drops of SAE 20 automotive oil annually. p.24 ↗ | Schedule annual lubrication after first two years to extend motor life. | Models with blower |
| Belt tension: approximately 5 pounds pressure to deflect belt 1/2 inch from normal position. p.24 ↗ | Check belt tension periodically and adjust to this specification. | Models with blower |
| Keep materials at least 6 inches inside the sash and at least 2 inches from side walls; perform contaminated operations as far to rear as possible. p.19 ↗ | Proper work positioning minimizes contamination escape and airflow disruption. | All models |
| Allow hood to operate unobstructed for two minutes after start-up and two minutes undisturbed after work before closing sash. p.19 ↗ | Purging periods ensure airborne contaminants are removed. | All models |
| After loading hood, wait one minute to purge airborne contaminants from work area. p.19 ↗ | Prevents exposure during loading. | All models |
| Perchloric acid should not be used in this hood. p.18 ↗ | Use a dedicated perchloric acid hood for such work. | All models |
| Radioisotope materials are not recommended for use in this hood. p.18 ↗ | Select a hood designed for radioisotope work if needed. | All models |
| Flammable or explosive materials should only be used if hood is equipped with explosion-proof components. p.18 ↗ | Standard models are not suitable for flammable/explosive materials. | All models |