9049605 Rev C WaterPro PS Systems User Manual
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco WaterPro PS Polishing Systems: General Chemistry (90005 series), HPLC (90006 series), UF Life Science (90007 series) and HPLC/UF Hybrid (9000722/9000732 series), in 115V/60Hz and 230V/50Hz versions, with dispensing valve or pistol/gun dispensing. Includes component identification, setup, operation, maintenance, troubleshooting, plumbing diagrams, component part numbers, dimensions and specifications.
Section-by-Section Technical Breakdown
Specifies site preparation: mounting surface must support at least 500 lb; electrical outlets rated 115 VAC/60 Hz/10 A or 230 VAC/50 Hz/10 A dedicated to the PS; feedwater must be prepurified to 100 µS or less at 0–100 psi and ≥2 L/min; drain required for HPLC, UF and Hybrid models; wall-mounted units need 6 in bottom clearance.
- Mounting surface must support at least 500 pounds; four ¼ x 2.0 in lag screws (1897432) supplied for wood frame studs.
- 115 V systems: detachable 15 A cord; outlet rated 115 VAC, 60 Hz, 10 A.
- 230 V systems: 10 A cord; outlet rated 230 VAC, 50 Hz, 10 A, dedicated to the PS only.
- Feedwater: prepurified (distilled, RO or DI) with conductivity 100 µS or less; pressure 0–100 psi; flow ≥2 L/min.
- Non-pressurized feed (storage tank): cabinet no more than 5 ft above tank; supply tube ≥1/4 in ID and ≤5 ft long.
- HPLC, UF and Hybrid models require a drain line routed to an open drain.
- Wall-mounted units require minimum 6 in clearance at bottom for cartridge/sump bowl removal.
- Optional electrical connection cord 1306000 connects a WaterPro RO to the PS; PS should be on the left side of the RO.
Describes the main functional components: inlet/drain ports, pressure regulator, recirculation pump, carbon/DI/organic adsorption cartridge bowls, moisture sensor, ultrafilter, TOC dump valve, UF flush valve, inlet valve, UV reactor, resistivity sensor, electronic control center, dispense valve/housing and check valve.
- Inlet port accepts 3/8 in OD rigid plastic tubing; feedwater 0–100 psi, ≥2 L/min, ≤100 µS.
- Pressure regulator is factory adjusted; call Labconco Product Service before adjusting.
- Recirculation pump is a rotary vane self-priming pump requiring water flow for lubrication; running without inlet water damages the pump.
- Moisture sensor closes the inlet valve if moisture is detected in the drip pan.
- Ultrafilter (UF models) removes particles, bacteria, microorganisms and pyrogens >0.01 micron (10,000 Dalton cut-off).
- UV reactor (HPLC, UF, Hybrid) irradiates at dual wavelengths 185 and 254 nm.
- TOC dump valve on HPLC and Hybrid systems directs a small portion of purified water to drain during automatic recirculation.
- UF flush valve on UF and Hybrid systems allows UF membrane flush during automatic recirculation.
Covers mounting with the supplied template, feedwater and drain connections, tubing installation with tube supports, cartridge installation sequence by model, and initial air purging. Sumps must be hand-tightened only.
- Feedwater tubing is 3/8 in OD rigid plastic; install an isolation valve in the supply line; line pressure must not exceed 100 psi (7 bar).
- Drain line must be weighted or permanently attached to prevent it coming out during flush.
- Flexible tubing requires a tube support insert; rigid tubing does not.
- Cartridge sequence (bowls 1–4 left to right): General Chemistry = Carbon, DI, DI, DI; HPLC/Hybrid = Carbon, DI, DI, Organic Adsorption; Life Science = Carbon, DI, DI, DI.
- Hand tighten sumps only — do not use wrench.
- After new cartridges, system is full of air; dispense to drain until air removed; on UF models turn unit off and press dispense, repeat.
- Optional hollow fiber final filter 9092900 threads onto dispense valve/gun after removing nozzle; hand tighten only.
- Initial timed dispense: set 1.5 hours and start unit; unit shuts off automatically.
Lists precautions on rinse-up time, feedwater quality, UV radiation, electrical safety, cartridge exhaustion behavior and UF membrane conditioning.
- After new cartridges, resistivity rises slowly; rinse-up may take up to 24 hours to reach 16–18 megohm-cm.
- After idle periods, allow five minutes of operation; UF models may require longer.
- Feedwater must be ≤100 µS; poorer quality exhausts cartridges faster and destroys the UF membrane.
- Always unplug before removing cover or servicing the UV lamp.
- As cartridges approach exhaustion, recirculation times lengthen and resistivity fluctuates, then drops rapidly without recovery.
- New or unused UF membranes: dispense to drain for two hours and recirculate overnight before use; membrane contains a preservative.
Describes the control center: ON/OFF, MODE, SET POINT and DISPENSE keys; MegOhm, Temp and Time Dispense modes; automatic intermittent recirculation; UF flush and TOC dump valve operation; start-up and shut-down procedures.
- When switch is OFF, pump automatically recirculates 4 minutes every 2 hours after being OFF for a full 2 hours.
- Modes: MegOhm (resistivity), Temp (°C), Time Dispense (minutes, adjustable with arrow keys).
- Set point adjustable up to 16 MegOhm; display flashes if resistivity drops below set point.
- DISPENSE key must be held down to continue dispensing; timed dispense available.
- UF flush valve opens 1 minute every 4 hours (alternate recirculation cycles).
- TOC dump valve on HPLC models opens 1 minute every 4 hours.
- Shut down: turn feed valve OFF, press dispense to relieve pressure, unplug unit.
Provides maintenance schedule, sanitization procedure with approved solutions, UV lamp annual replacement, cartridge replacement criteria, optional final filter replacement interval and ultrafilter replacement procedure.
- As needed: replace cartridges, clean dust, sanitize. Annually: replace UV lamp.
- Sanitization solutions: 0.3% bleach (1 L 5.25% household bleach in 15 L water), 2.0% formaldehyde, or 3.0% hydrogen peroxide; prepare ~16 liters.
- Recirculate sanitizer at least two hours, periodically opening dispense valve; discard without rinsing bowls; then dispense ~10 liters before use.
- Replace carbon cartridge at same time as deionization cartridges.
- Optional 0.2 µm hollow fiber final filter 9092900: replace every 2–3 weeks or when flow drops; lowers flow rate when installed.
- Ultrafilter 9104400 requires replacement if damaged by sanitization solution or biofouled/releasing pyrogens.
Problem/cause/corrective action table covering unit inoperative, no dispense, reduced flow, erratic water quality display, poor polished water quality, reduced cartridge life and odor.
- Unit inoperative causes: not plugged in, tripped breaker/blown fuse, power switch off, defective circuit board, inlet valve closed due to moisture in drip pan.
- No dispense with pump running: closed feed valve, restricted/no feedwater, clogged final filter, defective return line check valve, defective recirculation pump.
- Reduced flow: blocked optional final filter or restricted tubing.
- Erratic water quality display: new cartridges/air in system, air in resistivity cell, exhausted DI cartridges, resistivity cell disconnected or defective.
- Poor rinse-up: expired cartridges, wrong cartridge sequence, water bypassing cartridges (O-rings/posts).
- Reduced cartridge life: expired cartridges, changed feedwater quality.
- Odor: bacterial growth on filters or old DI resins breaking down.
Figures 8–11 show plumbing layouts and component part numbers for General Chemistry, HPLC Analytical Instrument, Ultrafilter Life Sciences and HPLC/Ultrafilter Hybrid models.
- General Chemistry models 9000520,-30,-40,-60,-70,-80 and pistol 9000521,-31,-41,-61,-71,-81.
- HPLC models 9000620,-30,-40,-50,-60,-70,-80 and pistol 9000621,-31,-41,-51,-61,-71,-81.
- UF and Hybrid diagrams include drain, UF, UV and TOC/UF flush valves.
- Common parts include inlet pump 91090 & -01, inlet valve 9106501, regulator 9047700 or 1359700, check valve 1365000, UV assembly 9114000.
Appendix A lists replacement part numbers; Appendix B gives dimensional drawings; Appendix C gives technologies, production rates and water quality.
- Pump assemblies: 9047500 (115 VAC), 9047501 (220 VAC); motor pump 1210101 (115 VAC).
- Resistivity cell 9105300; dispense valve 9106500 (115 V) / 9108900 (230 V); PCB 9103502.
- UV lamp 9113800 for 90006XX and 90007XX series; ultrafilter element 9104400 for 90007XX.
- Typical production: 1.8 L/min General Chemistry and HPLC; 1.2 L/min with 0.2 µm final filter; 1.1 L/min UF models; 1.0 L/min UF with final filter.
- Technologies: General Chemistry = carbon + DI; HPLC = carbon + DI + organic adsorption + UV 185/254 nm; UF = carbon + DI + ultrafiltration + UV; Hybrid = all.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Mounting surface must support at least 500 pounds. p.11 ↗ | Verify wall structure before wall mounting; inadequate support may damage surface or equipment. | All WaterPro PS models |
| 115 V models require an outlet rated 115 VAC, 60 Hz, 10 Amps; 230 V models require 230 VAC, 50 Hz, 10 Amps dedicated to the PS. p.11 ↗ | Provide a dedicated circuit with protective earthing; use only adequately rated detachable power cords. | 115 V and 230 V variants respectively |
| Feedwater must be prepurified to conductivity 100 µS or less, pressure 0–100 psi, flow ≥2 L/min. p.12 ↗ | Tap water or higher-conductivity feed rapidly exhausts DI cartridges and is not recommended. | All models |
| For non-pressurized feed (storage tank), cabinet must not be more than 5 feet above the tank; supply tube ≥1/4 in ID and ≤5 feet long. p.12 ↗ | Locate the PS within these limits when fed from a tank. | All models |
| Wall-mounted units require minimum 6 in clearance at the bottom for cartridge/sump bowl removal. p.12 ↗ | Plan mounting height to allow sump removal. | Wall-mounted units |
| Cartridge installation sequence: General Chemistry and Life Science = Carbon, DI, DI, DI; HPLC and Hybrid = Carbon, DI, DI, Organic Adsorption (bowls 1–4 left to right). p.20 ↗ | Wrong sequence causes poor water quality; install cartridges with O-rings pointing up. | 90005, 90006, 90007 series |
| Hand tighten sumps only — do not use wrench. p.20 ↗ | Prevents bowl/O-ring damage and leaks. | All models |
| After new cartridges, rinse-up may take up to 24 hours to reach 16–18 megohm-cm; after idle periods allow five minutes of operation. p.22 ↗ | Do not judge water quality immediately after cartridge change. | All models; UF models may require longer |
| Automatic recirculation runs 4 minutes every 2 hours when the ON/OFF switch is OFF for a full 2 hours. p.25 ↗ | Leaving the unit off still maintains water quality; leaving it on provides Type I water instantly during frequent use. | All models |
| UF flush valve opens 1 minute every 4 hours; TOC dump valve opens 1 minute every 4 hours. p.26 ↗ | These are preset and require no adjustment. | UF/Hybrid and HPLC models respectively |
| Set point adjustable up to 16 MegOhm; display flashes if resistivity drops below set point. p.25 ↗ | Use set point as an alarm for cartridge exhaustion. | All models |
| Sanitization solutions: 0.3% bleach (1 L 5.25% household bleach in 15 L water), 2.0% formaldehyde, or 3.0% hydrogen peroxide; prepare ~16 liters; recirculate at least 2 hours. p.30 ↗ | Follow approved concentrations; discard without rinsing bowls; dispense ~10 liters before use. | All models |
| UV lamp replacement is an annual maintenance item. p.29 ↗ | Schedule annual UV lamp replacement (part 9113800 for 90006XX/90007XX). | UV-equipped models |
| Optional 0.2 µm hollow fiber final filter 9092900 should be replaced every 2–3 weeks or when flow drops; it lowers flow rate. p.32 ↗ | Install only when required; remove/replace when flow becomes unacceptable. | All models with optional final filter |
| Typical production rates: 1.8 L/min General Chemistry and HPLC; 1.2 L/min with 0.2 µm final filter; 1.1 L/min UF models; 1.0 L/min UF with final filter. p.44 ↗ | Size dispense needs accordingly; final filter and UF reduce flow. | By model family |
| Ultrafilter removes particles, bacteria, microorganisms and pyrogens >0.01 micron (10,000 Dalton cut-off). p.17 ↗ | UF models suit bacteria/pyrogen-free applications. | UF and Hybrid models |
| UV reactor operates at dual wavelengths 185 and 254 nm. p.17 ↗ | Provides oxidation of organics and germicidal action. | HPLC, UF and Hybrid models |
| New or unused UF membranes: dispense to drain for two hours and recirculate overnight before use. p.22 ↗ | Prevents contamination and false megohm readings; membrane contains preservative. | UF and Hybrid models |
| Line pressure must not exceed 100 psi (7 bar). p.19 ↗ | Install a pressure regulator upstream if supply may exceed 100 psi. | All models |
| Optional electrical connection cord 1306000 connects a WaterPro RO to the PS; PS should be located on the left side of the RO. p.11 ↗ | Plan layout when pairing with a WaterPro RO System. | PS next to WaterPro RO |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Unit inoperative p.34 ↗ | Unit not plugged into outlet | Plug the Polishing Station into appropriate electrical service. | |
| Unit inoperative p.34 ↗ | Circuit breaker tripped/fuse blown | Reset circuit breaker/replace fuse. | |
| Unit inoperative p.34 ↗ | Power switch is in the off position | Press power button to switch system on. | |
| Unit inoperative p.34 ↗ | Defective circuit board | Replace circuit board. | |
| Unit inoperative p.34 ↗ | Inlet valve closed do to moisture in drip pan. | Check system for leaks. Dry drip pan. | |
| Recirculation pump operates but no water is dispensed p.34 ↗ | Customer supplied feed valve is closed | Open the valve. | |
| Recirculation pump operates but no water is dispensed p.34 ↗ | Feed-water line is restricted, or no supply | Inspect and adjust feedwater line as required. | |
| Recirculation pump operates but no water is dispensed p.34 ↗ | Final filter is clogged | Replace final filter. | |
| Recirculation pump operates but no water is dispensed (cont.) p.35 ↗ | Return line check valve is defective | Replace check valve assembly. | |
| Recirculation pump operates but no water is dispensed (cont.) p.35 ↗ | Recirculation pump is defective | Replace recirculation pump. | |
| Reduced flow at dispense valve/gun p.35 ↗ | Optional final filter is blocked | Replace final filter. | |
| Reduced flow at dispense valve/gun p.35 ↗ | Restricted tube in the feed line or polishing system | Inspect polishing loop tubing for any restrictions. | |
| Water quality display acting erratically p.35 ↗ | New cartridges installed in system | Purge all of the air out of the polishing system. | |
| Water quality display acting erratically p.35 ↗ | Air trapped in resistivity cell | Purge all of the air out of the polishing system. | |
| Water quality display acting erratically p.35 ↗ | Deionization cartridges are exhausted | Replace the deionization cartridges. | |
| Water quality display acting erratically p.35 ↗ | Polishing loop resistivity cell not connected to wiring harness | Reconnect cell to wiring harness. | |
| Water quality display acting erratically p.35 ↗ | Polishing loop resistivity cell is defective | Replace cell. | |
| Polished water will not rinse up to desired quality p.35 ↗ | Cartridges are past expiration date | The shelf life of unopened cartridges is two years from the date of manufacture. Replace the expired cartridges with new ones. Check the manufactured date. | |
| Polished water will not rinse up to desired quality p.35 ↗ | Cartridges installed in wrong sequence | Install the cartridges as described in the “Initial Operation and Cartridge Replacement” section. | |
| Polished water will not rinse up to desired quality p.35 ↗ | Water is bypassing the cartridges | Ensure that the O-rings are at the top of the deionization cartridges and the posts at the top of the bowls are inserted correctly. | |
| Reduced cartridge life p.36 ↗ | Cartridges are past expiration date | Check the manufactured date on the cartridge package. The shelf life of unopened cartridges is one year from the date of manufacture. Replace the expired cartridges with new ones. | |
| Reduced cartridge life p.36 ↗ | Feed-water quality has changed | If the source is a reverse osmosis system, ensure that that it is working properly. | |
| Reduced cartridge life p.36 ↗ | Feed-water quality has changed | If the source is a central water supply, verify that it is still working properly. Check the average tap water supply versus the deionization capacity chart on page 43. | |
| Odor from polished water p.36 ↗ | Bacterial growth on the filters | Remove old filters. Sanitize system. Install new filters. | |
| Odor from polished water p.36 ↗ | Old DI filters. DI resins breaking down | Remove old filters. Sanitize system. Install new filters. |